- Research
- Open access
- Published:
Quantification of both the area-at-risk and acute myocardial infarct size in ST-segment elevation myocardial infarction using T1-mapping
Journal of Cardiovascular Magnetic Resonance volume 19, Article number: 57 (2017)
Abstract
Background
A comprehensive cardiovascular magnetic resonance (CMR) in reperfused ST-segment myocardial infarction (STEMI) patients can be challenging to perform and can be time-consuming. We aimed to investigate whether native T1-mapping can accurately delineate the edema-based area-at-risk (AAR) and post-contrast T1-mapping and synthetic late gadolinium (LGE) images can quantify MI size at 1.5 T. Conventional LGE imaging and T2-mapping could then be omitted, thereby shortening the scan duration.
Methods
Twenty-eight STEMI patients underwent a CMR scan at 1.5 T, 3 ± 1 days following primary percutaneous coronary intervention. The AAR was quantified using both native T1 and T2-mapping. MI size was quantified using conventional LGE, post-contrast T1-mapping and synthetic magnitude-reconstructed inversion recovery (MagIR) LGE and synthetic phase-sensitive inversion recovery (PSIR) LGE, derived from the post-contrast T1 maps.
Results
Native T1-mapping performed as well as T2-mapping in delineating the AAR (41.6 ± 11.9% of the left ventricle [% LV] versus 41.7 ± 12.2% LV, P = 0.72; R2 0.97; ICC 0.986 (0.969–0.993); bias −0.1 ± 4.2% LV). There were excellent correlation and inter-method agreement with no bias, between MI size by conventional LGE, synthetic MagIR LGE (bias 0.2 ± 2.2%LV, P = 0.35), synthetic PSIR LGE (bias 0.4 ± 2.2% LV, P = 0.060) and post-contrast T1-mapping (bias 0.3 ± 1.8% LV, P = 0.10). The mean scan duration was 58 ± 4 min. Not performing T2 mapping (6 ± 1 min) and conventional LGE (10 ± 1 min) would shorten the CMR study by 15–20 min.
Conclusions
T1-mapping can accurately quantify both the edema-based AAR (using native T1 maps) and acute MI size (using post-contrast T1 maps) in STEMI patients without major cardiovascular risk factors. This approach would shorten the duration of a comprehensive CMR study without significantly compromising on data acquisition and would obviate the need to perform T2 maps and LGE imaging.
Background
Cardiovascular magnetic resonance (CMR) is considered the gold standard imaging modality for quantifying myocardial infarct (MI) size [1]. It can also provide information on the edema-based area-at-risk (AAR) [2, 3], which can be used to calculate myocardial salvage (by subtracting MI size from the AAR) - a more sensitive measure for assessing the efficacy of novel cardioprotection therapies for reducing MI size in reperfused ST-segment elevation myocardial infarction (STEMI) patients [4].
T2-weighted imaging in the first week following primary percutaneous coronary intervention (PPCI) has been used to delineate the AAR in STEMI patients [5], but has recently attracted controversy [6, 7]. T2-mapping [8] has emerged as a more robust technique but has not yet been established as the gold standard for the AAR (therefore referred to as the edema-based AAR throughout the article) and is a topic of ongoing research. Native T1-mapping has been shown to perform well against T2-mapping in a canine model of MI [3], and in clinical patients at 3.0 T [2]. So far, no study has directly compared the performance of T1-mapping against T2-mapping at 1.5 T in these patients.
Conventionally, MI can be identified by late gadolinium enhancement (LGE) with inversion recovery (IR) T1-weighted sequences and manual adjustment of the inversion times (TI) to null the remote normal myocardium. Post-contrast T1-mapping has also shown promise to detect acute MI in a small cohort of predominantly non-reperfused acute MI patients and chronic MI [9, 10]. Refinements of the T1 Modified Look-Locker Inversion recovery (MOLLI) prototype using IR images at different TIs initially designed to improve motion correction [11], can also create “synthetic” IR LGE images, namely magnitude-reconstructed IR (MagIR), phase-sensitive IR (PSIR) outputs. These synthetic read-outs have recently been shown to accurately detect chronic MI size in a small cohort of 17 patients [12]. To date, no studies have assessed the performance of post-contrast T1 mapping and synthetic LGE in quantifying MI size with full left ventricular (LV) coverage in a cohort of reperfused STEMI patients.
A comprehensive CMR study in STEMI can be time-consuming. Therefore, the aims of this study were to investigate, firstly, whether native T1-mapping can accurately delineate the edema-based AAR at 1.5 T (using T2-mapping as the reference); and secondly, whether post-contrast T1-mapping and synthetic IR LGE images can quantify MI size (using conventional LGE as the reference). This may allow one to omit conventional LGE imaging and T2-mapping to quantify MI size and the edema-based AAR, respectively, thereby shortening the duration of the scan.
Methods
Study population
Twenty-eight acute STEMI patients reperfused by PPCI were chosen from a cohort reported in previous studies [13,14,15,16,17] based on the availability of full LV coverage of post-contrast T1 mapping and the availability of the synthetic IR output dataset (recruitment period for the whole cohort was August 2013 to July 2014, the option to generate synthetic LGE images from the T1 maps were available from January 2014). The management of STEMI was as per current guidelines. Study exclusion criteria were known previous MI and standard recognized contraindications to CMR. The UK National Research Ethics Service approved this study and all patients provided informed consent. All 28 patients underwent CMR at 3 ± 1 days post-PPCI.
Imaging acquisition
All CMR scans were performed on a 1.5 Tesla scanner (Magnetom Avanto, Siemens Healthcare, Erlangen, Germany) using a 32-channel phased-array cardiac coil. The protocol included cine CMR images, T2 maps, native T1 maps, T2* maps, early gadolinium enhancement (EGE), conventional LGE and post-contrast T1 maps (All short axis T1/T2 maps, EGE and LGE images were aligned with the short axis cine MR images). Full LV short axis coverage was acquired for all sequences except for T2* maps (basal, mid and apical short axis). The T2* maps and EGE were not analyzed for this specific study but the time taken for each imaging sequence was recorded to the nearest minute to obtain representative duration of a comprehensive CMR study in these patients.
T2-mapping (work in progress #448B, Siemens healthcare)
T2 maps were acquired using 3 single-shot images at different T2 preparation times (0 ms, 24 ms, and 55 ms) as previously described [8]. The imaging parameters were: pixel bandwidth 930 Hz/pixel; echo time = 1.1 ms; repetition time = 3xR-R interval; flip angle = 70°; acquisition matrix = 116 × 192; slice thickness = 6 mm; slice gap = 4 mm; voxel size = 2.0 × 2.0 × 6.0 mm; temporal resolution = 190 ms; acceleration factor = 2). Fitting was performed inline to estimate T2 relaxation times and a colored T2 map consisting of pixel-wise T2 values was generated inline following motion correction.
Native T1-mapping (work in progress #448B, Siemens healthcare)
T1 maps were acquired using a steady state free precession-based MOLLI prototype. A 5 s(3 s)3 s modified MOLLI sampling protocol was used as previously described [11]. The acquisition parameters were: pixel bandwidth 977 Hz/pixel; echo time = 1.1 ms; flip angle = 35°; matrix = 256 × 144; slice thickness = 6 mm; slice gap = 4 mm; voxel size = 1.5 × 1.5 × 6.0 mm; temporal resolution = 227 ms; acceleration factor = 2). Motion correction and a non-linear least-square curve fitting were performed with the set of images acquired at different inversion times to generate a pixel-wise colored T1 map inline.
Late gadolinium enhancement
Conventional Mag-IR LGE imaging was acquired using predominantly a respiratory motion-corrected, free-breathing single shot steady state free precession averaged inversion recovery (MOCO-FB) LGE sequence in 22/28 patients (typical imaging parameters were: bandwidth = 977 Hz/pixel; echo time = 1.48 ms; repetition time = 700-900 ms; flip angle = 50°; acquisition matrix = 144 × 256; slice thickness = 8 mm; slice gap: 2 mm; voxel size = 1.6 × 1.6 × 8.0 mm; temporal resolution = 202 ms; acceleration factor 2) or a standard segmented ‘fast low-angle shot’ (FLASH) two-dimensional IR gradient echo sequence in 6/28 patients (imaging parameters were: bandwidth = 140 Hz/pixel; echo time = 3.17 ms; repetition time = 700-900 ms; flip angle = 21°; acquisition matrix = 125 × 256; slice thickness = 8 mm; slice gap: 2 mm; voxel size = 2.1 × 2.1 × 8.0 mm; temporal resolution = 140 ms; acceleration factor: 2) at 10 min after the injection of 0.1 mmol/kg of Gadoterate meglumine (Gd-DOTA marketed as Dotarem, Guerbet S.A., Paris, France). For both LGE sequences, the inversion times were progressively optimized to null the normal remote myocardium (typical values 360 to 440 ms).
Post-contrast T1-mapping (work in progress #448B, Siemens healthcare)
Post-contrast MOLLI T1 maps were obtained in-line using the 4 s(1 s)3 s(1 s)2 s sampling protocol (to improve the accuracy of T1 s in the 200-600 ms range as previously described [11]) 15 min after contrast injection (0.1 mmol/kg of Dotarem) using similar acquisition parameters as for native T1 maps.
Both the native and post-contrast MOLLI T1-mapping automatically generated 2 sets of IR images (MagIR and PSIR) in-line by the scanner for the TI range between 200 to 975 ms (25 ms increments) as illustrated in Fig. 1.
Imaging analysis
All imaging analysis was performed using CVI42 software (Version 5.1.2, Calgary, Canada).
Quantification of MI size
The endocardial and epicardial borders were manually delineated on all the conventional LGE short-axis LV images and were copied to all corresponding images for analysis with minimal manual adjustments. All sequences were subsequently analyzed independent of each other. A reference region of interest (ROI) was drawn in the remote normal myocardium using the automated option available on CVI42 (Circle Cardiovascular Imaging, Calgary, Canada) to minimize variability (with minimal manual adjustment when needed) and this was done for each imaging technique independently. MI size was quantified using a signal intensity threshold of 5 standard deviations (SD) above the normal remote myocardium [12, 18] and expressed as the percentage of the whole LV (%LV). Transmural extent of infarct (TEI) was also quantified and expressed in groups of “no LGE”, “1–25%TEI”, “26–50%TEI”, “51–75% TEI” and “76–100%TEI” as per the 16-segment American Heart Association model.
For the synthetic MagIR LGE, the image with the optimal TI (remote normal myocardium nulled as black) was manually selected and the corresponding matching PSIR image was selected as the representative synthetic PSIR LGE image (Fig. 1). A threshold of 5SD was also used to quantify synthetic MI size.
For the post-contrast T1 maps, we used 2 semi-automated techniques, namely a threshold of 2-SD below the mean remote myocardial post-contrast T1 and the Otsu technique (which identifies the intensity threshold from the signal intensity histogram using the value with minimal intra-class variance between low and high intensities) [19]. These 2 techniques were applied on a slice-by-slice basis. MI size was expressed as: by conventional LGE (reference standard) as MIConv; MagIR synthetic LGE as MISynthMagIR; PSIR synthetic LGE as MISynthPSIR; post-contrast T1 MI size by 2-SD as MIT1Post2SD; and by Otsu as MIT1PostOtsu.
Quantification of the edema-based AAR
The endocardial and epicardial borders were manually traced on the T2 maps and copied to the T1 maps with minimal manual adjustments. In the absence of established quantification techniques, we used manual contouring as reference (%LV) and compared against a threshold of 2-SD above the remote myocardium and the Otsu technique (expressed as AART2Man, AART22SD, AART2Otsu and AART1Man, AART12SD, AART1Otsu respectively). The Otsu technique was not applied on maps that were visually normal based on the colored look-up table and with normal wall motion on cine MR images were manually selected as normal.
Areas of hypo-intense core of microvascular obstruction (MVO) were included as part of the MI zone and edema-based AAR. Manual ROIs were drawn in MI zone and the remote myocardium to obtain representative native and post-contrast T1 and T2 values.
Statistical analysis
SPSS version 22 (International Business Machines Corporation, Chicago, Illinois, US) was used for all statistical analysis. Shapiro-Wilk Test was used to assess for normality and our data was normal distributed. Continuous data were expressed as mean ± SD. Categorical data were reported as frequencies and percentages. Paired student t test was used for edema-based AAR and MI size comparison by different thresholding techniques and by different sequences. T1 and T2 values in the MI zone and remote myocardium were compared with unpaired student t-test. Inter-method correlation was performed using the squared Pearson’s correlation coefficient (R2) and inter-method agreement was assessed using intra-class correlation coefficient (ICC) with 95% confidence intervals (95%CI) and Bland-Altman analysis (expressed as bias ±2SD for limits of agreement). All statistical tests were two-tailed, and P < 0.05 was considered statistically significant.
Results
The CMR was performed at 3 ± 1 days post-PPCI. Baseline characteristics are listed in Table 1. The mean age of the STEMI patients was 57 ± 13 years old and 22/28 (79%) were male. The mean MI size (MIConv) in the STEMI patients was 25.1 ± 14.3%LV and the mean edema-based AAR by T2 (AART2Man) was 41.5 ± 12.0%LV. The myocardial salvage index in this cohort was 0.41 ± 0.28. Out of 448 AHA segments the TEI were as follows: no LGE: 262 (59%); 1–25%TEI: 32 (7%); 26–50%TEI: 60 (13%); 51–75% TEI: 45 (10%); and 76–100%TEI: 49 (11%). 5 patients had TIMI flow grade 3 and 3 patients had TMI flow grade 2 pre-PPCI. The MI size in these 8 patients was significantly smaller than those with TIMI flow 0/1 pre-PPCI (14.0(2.3–25.7)%LV versus 25.0(22.3–34.8)%LV, P = 0.04). All patients had areas of hyperenhancement on the LGE areas of high signal on the T1 and T2 maps.
Figure 2 illustrates examples from 3 patients for the edema-based AAR by T2-mapping and native T1-mapping and the detection of the acute MI size by conventional LGE, synthetic MagIR LGE and synthetic PSIR LGE, and post-contrast T1 maps.
T2 and T1 values in MI zone and remote myocardium
T2 in the MI zone was significantly higher than in the remote myocardium (67 ± 7 ms versus 50 ± 3 ms, P < 0.0001). Native T1 in the MI zone was also significantly higher than in the remote myocardium (1258 ± 57 ms versus 1041 ± 57 ms, P < 0.001). Post-contrast T1 was significantly lower in the MI zone compared to the remote myocardium (441 ± 48 ms versus 625 ± 22 ms, P < 0.001).
Edema-based AAR quantification
The Otsu technique failed in 8% of the maps with extensive MVO and transmural LGE and the edema-based AAR was manually delineated based on the transmural extension of the MI for these slices. There was an excellent correlation (R2 0.99) and agreement (ICC 0.983 and 0.996 respectively) between AART2Man and the 2 semi-automated techniques. AART2Otsu was similar to AART2Man (41.7 ± 12.2%LV versus 41.5 ± 12.1%LV, P = 0.38) with no bias (0.2 ± 2.1%). However, AART22SD was significantly larger than AART2Man (42.2 ± 12.1%LV versus 41.5 ± 12.1%LV P = 0.011) with a bias of 1.0 ± 2.6%. (Further details in Table 2).
The same pattern was observed for T1-mapping. The AART1Otsu was similar to AART1Manual (41.6 ± 11.9%LV versus 41.0 ± 12.0%LV, P = 0.17) with no bias (−0.6 ± 4.2%), and AART12SD was significantly higher than AART1Manual (42.1 ± 12.2%LV versus 41.0 ± 12.0%LV, P = 0.005) with a small bias of 1.0 ± 3.9%. (Further details in Table 2).
Native T1-mapping performed as well as T2-mapping in delineating the edema-based AAR (AART1Otsu 41.6 ± 11.9%LV versus AART2Otsu 41.7 ± 12.2%LV, P = 0.72; R2 0.97; ICC 0.986(0.969–0.993); bias −0.1 ± 4.2%LV). Table 2 and Fig. 3 provide further details of the comparison between T2 and native T1-mapping for the quantification of the edema-based AAR using the different techniques. Figure 4 illustrates the comparison between AART2Otsu and AART1Otsu.
Acute MI size quantification
There was an excellent correlation (R2 0.96 to 0.99) and inter-method agreement (ICC 0.964 to 0.996) between conventional LGE and the other techniques. There was minimal bias between MIConv and MISynthMagIR (bias: 0.2 ± 2.2%LV, P = 0.35), MIConv and MISynthPSIR (bias: 0.4 ± 2.2%LV, P = 0.060) and MIConv and MIT1Post2SD (bias: 0.3 ± 1.8%LV, P = 0.10), with the latter having the narrowest limits of agreement. However, MIT1PostOtsu underestimated MI size when compared to MIConv (bias −2.2 ± 6.0%LV, P = 0.001). Figure 3 shows the comparison of the various techniques used for MI size. Figure 5 illustrates the correlations and Bland-Altman analysis of MIConv against the 4 other techniques and further details are summarized in Table 2.
CMR scan duration
The average time for the complete STEMI CMR scan was 58 ± 4 min (including natural breaks between sequences). The average scanning time for the T2 and T1 maps were 6 ± 1 min each and for LGE it was 10 ± 1 min. Figure 6 illustrates the hypothetical 15–20 min reduction in scanning time obtained from potentially omitting T2-mapping and LGE images to determine the edema-based AAR and MI size, respectively.
Discussion
The main findings from our study are as follows: firstly, the edema-based AAR can be accurately quantified using native T1-mapping when compared to T2-mapping at 1.5 T; secondly, the acute MI size can be accurately quantified from the post-contrast T1 maps (2-SD) and the synthetic LGE images when compared to conventional LGE imaging; and finally, using T1-mapping to quantify the edema-based AAR and MI size may allow one to omit T2 mapping and LGE image, thereby shortening the scanning time by 15–20 min.
We have previously shown that native T1-mapping can accurately quantify the edema-based AAR at 3 T (analyzed by the Otsu technique) when compared to T2-mapping [2]. Other groups [20, 21] have shown that the Otsu technique also performed well on T2-weighted STIR images to detect the edema-based AAR. However, when using the Otsu technique, manual input is required when no edema is present (using visual assessment of the colored maps and cine MR images as reference), and when extensive MVO is present on the maps. Langhans et al. [22] also demonstrated that native T1-mapping and T2-mapping quantified the AAR when compared to SPECT at 1.5 T, although a direct comparison of T1 against T2 was not performed in that study. We have now shown that native T1-mapping can also accurately delineate the edema-based AAR when compared to T2-mapping at 1.5 T in the clinical setting and that the Otsu technique performed best against manual contouring. The limits of agreement between T2 and T1-mapping were ±4.2%LV, but narrower than our previous study at 3 T (±5.1%LV). T2 mapping is sensitive to the water content and therefore performs well in the context of edema during acute MI. Inversion recovery-based T1 mapping is influenced by T2 due to the use of a steady-state free precession readout leading to a T2 dependent error in the estimation of T1 [11]. Furthermore, a T2-dependent error also occurs due to the imperfect inversion efficiency of the adiabatic radiofrequency preparation. These factors result in the T1 maps being T2-weighting and sensitive to edema. Both T1 and T2 mapping are also equally affected by MVO and IMH [17]. Therefore, it is not surprising that they performed equally well to quantify the AAR. The edema-based AAR by T2 mapping has been shown to be dynamic within the first week of a STEMI [7, 23]. Whether the AAR by T1-mapping will follow the same pattern or would be more stable within the first week of a STEMI is not known and warrants further investigations.
Messroghli et al. [9] have previously shown that post-contrast T1 mapping could detect both acute and chronic MI in a cohort of 24 patients but not all patients underwent coronary angiography prior to CMR and only 21% showed evidence of reperfusion. Bauner et al. [10] have subsequently shown that post-contrast T1 could accurately detect chronic MI in 26 patients. Synthetic LGE has also been shown to accurately quantify chronic MI size in a small cohort of patients [12], but in this study, only one short-axis LV image was analyzed. In our study, we performed full LV coverage and we have shown that post-contrast T1 maps can accurately quantify acute MI size (using the semi-automated 2-SD threshold) when compared to the reference standard in a cohort of reperfused STEMI patients. Using the SD technique for quantification can be variable depending on where the ROI is drawn. To minimize the error associated with this, we have used the automated ROI delineation option with minimal manual adjustment when required. Furthermore, we have also shown that the additional sets of IR images that can be easily obtained from the post-contrast MOLLI T1 maps can be used to obtain synthetic LGE images (both MagIR and PSIR) and can accurately quantify acute MI size. PSIR LGE is T1 weighted with retrospective nulling of the remote myocardium thus being insensitive to the inversion time used for acquisition [24]. With the PSIR technique, the tissue with the longest TI nulls first, irrespective of the TI used (post gadolinium, the remote myocardium has the longest TI) and this sequence has previously been shown to improve MI detection more accurately by LGE compared to conventional MagIR LGE images [25]. Therefore, the synthetic PSIR output could complement post-contrast T1 maps for MI size quantification without the need for conventional LGE images, which requires the user to manually adjust the TI. Post-contrast T1 maps have previously been used as the truth standard for LGE imaging in the setting of cardiac amyloidosis, where nulling of the myocardium is challenging [26]. Therefore, post-contrast T1 mapping with the synthetic LGE output may be useful to identify the pattern of scarring in other pathologies with focal fibrosis such as hypertrophic cardiomyopathy, sarcoidosis and myocarditis and warrants further investigation.
A comprehensive CMR study in a reperfused acute STEMI patient requiring data on the edema-based AAR, intramyocardial hemorrhage (IMH), MVO, MI size and ECV, would take on average, an hour to perform. Based on the results from this study, T2-mapping and conventional LGE could be omitted - this would substantially shorten the scanning time by 15–20 min (see Fig. 6a and b) without compromising data acquisition. This approach would significantly improve patient workflow, make the CMR scan more tolerable to acute STEMI patients, and may help minimize patient dropout in future clinical studies. In our center this could be achieved by omitting T1 or T2 mapping and conventional LGE and performing the short axis cine MR images post-contrast instead (that could also provide another method for the AAR) [27] as illustrated in Fig. 6. Native T1-mapping (for AAR) and post-contrast T1 mapping (for LGE) approach has added value in that the ECV can be derived (Fig. 6d), and this may also be accelerated if the ECV maps are generated inline [28]. Alternatively, If only data on edema-based AAR and MI are required, the T1-mapping could be replaced by T2-mapping (Fig. 6c). The abbreviated protocol could be further shortened in patients who cannot tolerate long scans by omitting the T2* maps (for IMH) and EGE images (for early MVO) and could bring down the scan time to around 30 min. We recently showed that the presence of a hypo-intense core on the T1 maps or T2 maps can detect IMH well and with good sensitivity (85–88%) and specificity (85%) and therefore could be used to detect IMH in situations when T2* mapping cannot be performed or is not available [17].
The appearance of the synthetic MagIR and PSIR is noisy in regions such as air or lungs where the signal to noise ratio is low and the T1 estimates are noisy. This appearance of increased noise in these non-tissue regions is similar to the appearance of PSIR LGE images that results due to correction of surface coil intensities. Surface coil correction is also known as normalization and is generally accompanied by noise amplification in non-tissue regions. Various schemes are often employed to improve the appearance by softening the normalization in these regions. However, these methods may result in some degree of error in tissue regions. This issue, albeit mostly cosmetic, was not addressed in this article and represents a potential impediment to acceptance by clinicians.
Limitations of study
The conventional LGE scan (acquired 10–15 min post contrast) and post-contrast-T1 maps (acquired 15–20 min post contrast) were acquired 5 min apart. But despite this, there was no difference in MI size. The TI range for the IR images generated in-line from the T1 maps was quite wide (200 ms to 975 ms and at 25 ms increments). But off-line post-processing was not necessary and synthetic MagIR LGE images (Fig. 2) were interpretable and performed as well as conventional LGE for MI size quantification. Our sample size was small but still larger than previously published studies on the edema-based AAR by T1 and T2 mapping [2] and synthetic LGE [12]. We used the edema as a surrogate marker for the AAR, although controversy exists over whether it exists at all [6] or there is a bimodal or unimodal pattern of edema in the first week of an MI [7, 23]. We used 2 different LGE read-outs (FLASH LGE in 22 patients and MOCO-FB LGE in 6 patients) for MI size quantification but both these sequences have been shown to perform equally well for MI size quantification [29]. For some centers not performing such a comprehensive study for STEMI, their scanning time may not be shortened significantly as they would still need to wait 15 min to perform the post-contrast T1 maps. However, they would have the benefit of having 3 outputs to quantify MI size by simply performing post-contrast T1 mapping instead of conventional LGE without having to manually null the remote myocardium. There was a period of 3 months overlap between recruitment of patients in this study and our previous study on T1/T2 at 3 T [2] (part of a positron emission tomography (PET)/MR study) and there may have been an element of selection bias with those >45 years of age and without a history of diabetes being more likely recruited to the latter studies [2, 30]. We have previously shown that the T1 values in the remote myocardium is higher that normal in the acute phase of a STEMI when compared to controls and this might have affected the n-SD technique where an ROI in the remote myocardium is used as reference. There is a regional variation in T1 [31] and T2 [32] values from base to apex, partly contributed by thinner walls and more partial volume effects towards the apex. However, we did not use one cut-off value as threshold for the whole LV but analysis was done on a per-slice basis (ROIs for the 2SD technique were drawn on each slice and the Otsu technique was applied on each slice) to obtain the extent of hyperenhancement. Only around a third of our cohort had the conventional cardiovascular risk factors and therefore these findings apply to this select cohort of STEMI patients only and further studies are required to assess whether synthetic LGE derived from post-contrast T1 maps could be useful in a wider range of patients with STEMI and other pathologies.
Conclusions
T1-mapping can accurately quantify both the edema-based AAR (using native T1 maps) and acute MI size (using post-contrast T1 maps) in this select group of low-risk STEMI patients, obviating the need to perform T2 maps and LGE imaging, thereby shortening the duration of a comprehensive CMR study without significantly compromising on data acquisition. This approach would make the CMR scan more tolerable to acute STEMI patients.
Abbreviations
- AAR:
-
area-at-risk
- CMR:
-
cardiovascular magnetic resonance
- EGE:
-
early gadolinium enhancement
- FLASH:
-
fast low-angle shot
- ICC:
-
intra-class correlation coefficient
- IMH:
-
intramyocardial hemorrhage
- LGE:
-
late gadolinium enhancement
- LV:
-
left ventricular
- MagIR:
-
magnitude-reconstructed inversion recovery
- MI:
-
myocardial infarct
- MOCO-FB:
-
motion-corrected, free-breathing
- MOLLI:
-
modified Look Locker Inversion recovery
- MVO:
-
microvascular obstruction
- PPCI:
-
primary percutaneous coronary intervention
- PSIR:
-
phase-sensitive inversion recovery
- ROI:
-
region of interest
- SD:
-
standard deviation
- STEMI:
-
ST-elevation myocardial infarction
- TEI:
-
transmural extent of infarct
- TI:
-
inversion time
References
Thiele H, Kappl MJ, Conradi S, Niebauer J, Hambrecht R, Schuler G. Reproducibility of chronic and acute infarct size measurement by delayed enhancement-magnetic resonance imaging. J Am Coll Cardiol. 2006;47:1641–5.
Bulluck H, White SK, Rosmini S, Bhuva A, Treibel TA, Fontana M, Abdel-Gadir A, Herrey A, Manisty C, Wan SM, et al. T1 mapping and T2 mapping at 3T for quantifying the area-at-risk in reperfused STEMI patients. J Cardiovasc Magn Reson. 2015;17:73.
Ugander M, Bagi PS, Oki AJ, Chen B, Hsu LY, Aletras AH, Shah S, Greiser A, Kellman P, Arai AE. Myocardial edema as detected by pre-contrast T1 and T2 CMR delineates area at risk associated with acute myocardial infarction. JACC Cardiovasc Imaging. 2012;5:596–603.
Friedrich MG, Abdel-Aty H, Taylor A, Schulz-Menger J, Messroghli D, Dietz R. The salvaged area at risk in reperfused acute myocardial infarction as visualized by cardiovascular magnetic resonance. J Am Coll Cardiol. 2008;51:1581–7.
Wright J, Adriaenssens T, Dymarkowski S, Desmet W, Bogaert J. Quantification of myocardial area at risk with T2-weighted CMR: comparison with contrast-enhanced CMR and coronary angiography. JACC Cardiovasc Imaging. 2009;2:825–31.
Kim HW, Van Assche L, Jennings RB, Wince WB, Jensen CJ, Rehwald WG, Wendell DC, Bhatti L, Spatz DM, Parker MA, et al. Relationship of T2-weighted MRI myocardial Hyperintensity and the ischemic area-at-risk. Circ Res. 2015;117:254–65.
Fernandez-Jimenez R, Sanchez-Gonzalez J, Aguero J, Garcia-Prieto J, Lopez-Martin GJ, Garcia-Ruiz JM, Molina-Iracheta A, Rossello X, Fernandez-Friera L, Pizarro G, et al. Myocardial edema after ischemia/reperfusion is not stable and follows a bimodal pattern: imaging and histological tissue characterization. J Am Coll Cardiol. 2015;65:315–23.
Giri S, Chung YC, Merchant A, Mihai G, Rajagopalan S, Raman SV, Simonetti OP. T2 quantification for improved detection of myocardial edema. J Cardiovasc Magn Reson. 2009;11:56.
Messroghli DR, Walters K, Plein S, Sparrow P, Friedrich MG, Ridgway JP, Sivananthan MU. Myocardial T1 mapping: application to patients with acute and chronic myocardial infarction. Magn Reson Med. 2007;58:34–40.
Bauner KU, Biffar A, Theisen D, Greiser A, Zech CJ, Nguyen ET, Reiser MF, Wintersperger BJ. Extracellular volume fractions in chronic myocardial infarction. Investig Radiol. 2012;47:538–45.
Kellman P, Hansen MS. T1-mapping in the heart: accuracy and precision. J Cardiovasc Magn Reson. 2014;16:2.
Varga-Szemes A, van der Geest RJ, Spottiswoode BS, Suranyi P, Ruzsics B, De Cecco CN, Muscogiuri G, Cannao PM, Fox MA, Wichmann JL, et al. Myocardial late gadolinium enhancement: accuracy of T1 mapping-based synthetic inversion-recovery imaging. Radiology. 2016;278:374–82.
Bulluck H, Rosmini S, Abdel-Gadir A, White SK, Bhuva AN, Treibel TA, Fontana M, Gonzalez-Lopez E, Reant P, Ramlall M, et al. Automated ECV mapping provides insights into the pathophysiology of LV remodeling post reperfused STEMI. J Am Heart Assoc. 2016;5:e003555.
Bulluck H, Rosmini S, Abdel-Gadir A, Bhuva AN, Treibel TA, Fontana M, Weinmann S, Sirker A, Herrey AS, Manisty C, et al. Impact of microvascular obstruction on semiautomated techniques for quantifying acute and chronic myocardial infarction by cardiovascular magnetic resonance. Open Heart. 2016;3:e000535.
Bulluck H, Rosmini S, Abdel-Gadir A, White SK, Bhuva AN, Treibel TA, Fontana M, Ramlall M, Hamarneh A, Sirker A, et al. Residual myocardial iron following Intramyocardial hemorrhage during the convalescent phase of Reperfused ST-segment–elevation myocardial infarction and adverse left ventricular remodeling. Circ Cardiovasc Imaging. 2016;9:e004940.
Bulluck H, Go YY, Crimi G, Ludman AJ, Rosmini S, Abdel-Gadir A, Bhuva AN, Treibel TA, Fontana M, Pica S, et al. Defining left ventricular remodeling following acute ST-segment elevation myocardial infarction using cardiovascular magnetic resonance. J Cardiovasc Magn Reson. 2017;19:26.
Bulluck H, Rosmini S, Abdel-Gadir A, Bhuva AN, Treibel TA, Fontana M, Gonzalez-Lopez E, Ramlall M, Hamarneh A, Sirker A, et al. Diagnostic performance of T1 and T2 mapping to detect Intramyocardial hemorrhage in Reperfused ST-segment elevation myocardial infarction (STEMI) patients. J Magn Reson Imaging 2017. In press.
Schulz-Menger J, Bluemke DA, Bremerich J, Flamm SD, Fogel MA, Friedrich MG, Kim RJ, von Knobelsdorff-Brenkenhoff F, Kramer CM, Pennell DJ, et al. Standardized image interpretation and post processing in cardiovascular magnetic resonance: Society for Cardiovascular Magnetic Resonance (SCMR) board of trustees task force on standardized post processing. J Cardiovasc Magn Reson. 2013;15:35.
Otsu N. A threshold selection method from gray-level histograms. IEEE Trans Syst Man Cyberb. 1979;9:62–6.
McAlindon E, Pufulete M, Lawton C, Angelini GD, Bucciarelli-Ducci C. Quantification of infarct size and myocardium at risk: evaluation of different techniques and its implications. Eur Heart J Cardiovasc Imaging. 2015;16:738–46.
Khan JN, Nazir SA, Horsfield MA, Singh A, Kanagala P, Greenwood JP, Gershlick AH, GP MC. Comparison of semi-automated methods to quantify infarct size and area at risk by cardiovascular magnetic resonance imaging at 1.5T and 3.0T field strengths. BMC Res Notes. 2015;8:52.
Langhans B, Nadjiri J, Jahnichen C, Kastrati A, Martinoff S, Hadamitzky M. Reproducibility of area at risk assessment in acute myocardial infarction by T1- and T2-mapping sequences in cardiac magnetic resonance imaging in comparison to Tc99m-sestamibi SPECT. Int J Cardiovasc Imaging. 2014;30:1357–63.
Carrick D, Haig C, Ahmed N, Rauhalammi S, Clerfond G, Carberry J, Mordi I, McEntegart M, Petrie MC, Eteiba H, et al: Temporal evolution of myocardial hemorrhage and edema in patients after acute ST-segment elevation myocardial infarction: pathophysiological insights and clinical implications. J Am Heart Assoc. 2016, 5:Advance online publication.
Kellman P, Arai AE, McVeigh ER, Aletras AH. Phase-sensitive inversion recovery for detecting myocardial infarction using gadolinium-delayed hyperenhancement. Magn Reson Med. 2002;47:372–83.
Huber AM, Schoenberg SO, Hayes C, Spannagl B, Engelmann MG, Franz WM, Reiser MF. Phase-sensitive inversion-recovery MR imaging in the detection of myocardial infarction. Radiology. 2005;237:854–60.
Fontana M, Pica S, Reant P, Abdel-Gadir A, Treibel TA, Banypersad SM, Maestrini V, Barcella W, Rosmini S, Bulluck H, et al. Prognostic value of late gadolinium enhancement cardiovascular magnetic resonance in cardiac Amyloidosis. Circulation. 2015;132:1570–9.
Nordlund D, Kanski M, Jablonowski R, Koul S, Erlinge D, Carlsson M, Engblom H, Aletras AH, Arheden H. Experimental validation of contrast-enhanced SSFP cine CMR for quantification of myocardium at risk in acute myocardial infarction. J Cardiovasc Magn Reson. 2017;19:12.
Treibel TA, Fontana M, Maestrini V, Castelletti S, Rosmini S, Simpson J, Nasis A, Bhuva AN, Bulluck H, Abdel-Gadir A, et al. Automatic measurement of the myocardial Interstitium: synthetic extracellular volume quantification without Hematocrit sampling. JACC Cardiovasc Imaging. 2016;9:54–63.
Piehler KM, Wong TC, Puntil KS, Zareba KM, Lin K, Harris DM, Deible CR, Lacomis JM, Czeyda-Pommersheim F, Cook SC, et al. Free-breathing, motion-corrected late gadolinium enhancement is robust and extends risk stratification to vulnerable patients. Circ Cardiovasc Imaging. 2013;6:423–32.
Bulluck H, White SK, Frohlich GM, Casson SG, O'Meara C, Newton A, Nicholas J, Weale P, Wan SM, Sirker A, et al. Quantifying the area at risk in Reperfused ST-segment-elevation myocardial infarction patients using hybrid cardiac positron emission tomography-magnetic resonance imaging. Circ Cardiovasc Imaging. 2016;9:e003900.
Reiter U, Reiter G, Dorr K, Greiser A, Maderthaner R, Fuchsjager M. Normal diastolic and systolic myocardial T1 values at 1.5-T MR imaging: correlations and blood normalization. Radiology. 2014;271:365–72.
Bonner F, Janzarik N, Jacoby C, Spieker M, Schnackenburg B, Range F, Butzbach B, Haberkorn S, Westenfeld R, Neizel-Wittke M, et al. Myocardial T2 mapping reveals age- and sex-related differences in volunteers. J Cardiovasc Magn Reson. 2015;17:9.
Acknowledgements
We express our gratitude to the staff and patients at the UCLH Heart Hospital and Peter Weale for providing us with the Work In Progress investigational sequences under a research collaboration agreement with Siemens Healthcare.
Funding
This work was supported by the British Heart Foundation (FS/10/039/28270), the Rosetrees Trust, and the National Institute for Health Research University College London Hospitals Biomedical Research Centre.
Availability of data and materials
Further details on the data used to reach the conclusions in this article are available upon request.
Author information
Authors and Affiliations
Contributions
Made substantial contributions to conception and design, or acquisition of data, or analysis and interpretation of data: HB, MHH, MF, DSK. Been involved in drafting the manuscript or revising it critically for important intellectual content: AS, ASH, CM, DMY, PK, JCM, DJH. Given final approval of the version to be published: HB, MHH, MF, DSK, AS, ASH, CM, DMY, PK, JCM, DJH. All authors agreed to be accountable for all aspects of the work in ensuring that questions related to the accuracy or integrity of any part of the work are appropriately investigated and resolved.
Corresponding author
Ethics declarations
Authors’ information
No applicable.
Ethics approval and consent to participate
The UK National Research Ethics Service, London-Harrow Health Research Authority approved this study and all patients provided informed written consent.
Consent for publication
Not applicable.
Competing interests
The authors declare that they have no competing interests.
Rights and permissions
Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.
About this article
Cite this article
Bulluck, H., Hammond-Haley, M., Fontana, M. et al. Quantification of both the area-at-risk and acute myocardial infarct size in ST-segment elevation myocardial infarction using T1-mapping. J Cardiovasc Magn Reson 19, 57 (2017). https://doi.org/10.1186/s12968-017-0370-6
Received:
Accepted:
Published:
DOI: https://doi.org/10.1186/s12968-017-0370-6