Fig. 5 | Journal of Cardiovascular Magnetic Resonance

Fig. 5

From: Validation of 4D flow cardiovascular magnetic resonanceĀ in TIPS stent grafts using a 3D-printed flow phantom

Fig. 5

Correlation of 4D flow CMR-derived flow rates and preset flow rates. Preset TIPS flow rates were increased from 0.8 to 2.8Ā l/min and flow measurements were performed using velocity encoding both at 1Ā m/s (black crosses) and at 2Ā m/s (red dots). Flow measurements were performed at indicated levels. 4D flow CMR-derived flow rates were significantly correlated with the preset flow rates at all vascular levels and within the stent graft (all rā€‰ā‰„ā€‰0.985 and all pā€‰ā‰¤ā€‰0.049). At velocity encoding of 1Ā m/s, flow measurements were excluded if velocities exceeded velocity encoding and caused aliasing artifacts. At velocity encoding of 2Ā m/s, aliasing did not occur. SMVā€‰ā€‰superior mesenteric vein, SVā€‰ā€‰splenic vein, PVā€‰ā€‰portal vein, TIPSbā€‰ā€‰bare part of the stent, TIPSpā€‰ā€‰portal sided third of the stent, TIPSmā€‰ā€‰middle third of the stent, TIPSvā€‰ā€‰venous sided third of the stent