Fig. 5From: Validation of 4D flow cardiovascular magnetic resonanceĀ in TIPS stent grafts using a 3D-printed flow phantomCorrelation of 4D flow CMR-derived flow rates and preset flow rates. Preset TIPS flow rates were increased from 0.8 to 2.8Ā l/min and flow measurements were performed using velocity encoding both at 1Ā m/s (black crosses) and at 2Ā m/s (red dots). Flow measurements were performed at indicated levels. 4D flow CMR-derived flow rates were significantly correlated with the preset flow rates at all vascular levels and within the stent graft (all rāā„ā0.985 and all pāā¤ā0.049). At velocity encoding of 1Ā m/s, flow measurements were excluded if velocities exceeded velocity encoding and caused aliasing artifacts. At velocity encoding of 2Ā m/s, aliasing did not occur. SMVāāsuperior mesenteric vein, SVāāsplenic vein, PVāāportal vein, TIPSbāābare part of the stent, TIPSpāāportal sided third of the stent, TIPSmāāmiddle third of the stent, TIPSvāāvenous sided third of the stentBack to article page