Validation of 4D flow cardiovascular magnetic resonanceĀ in TIPS stent grafts using a 3D-printed flow phantom

Abstract

Background

Four-dimensional (4D) flow cardiovascular magnetic resonance (CMR) is feasible for portal blood flow evaluation after placement of transjugular intrahepatic portosystemic shunts (TIPS) in patients with liver cirrhosis. However, clinical acceptance of 4D flow CMR in TIPS patients is limited due to the lack of validation studies. The purpose of this study was to validate 4D flow CMR-derived measurements in TIPS stent grafts using a three-dimensional (3D)-printed flow phantom.

Methods

A translucent flow phantom of the portal vasculature was 3D-printed. The phantom consisted of the superior mesenteric vein and the splenic vein draining into the portal vein, the TIPS-tract, and the hepatic vein. A TIPS stent graft (GoreĀ® ViatorrĀ®) was positioned within the TIPS-tract. Superior mesenteric vein and splenic vein served as inlets for blood-mimicking fluid. 4D flow CMR acquisitions were performed at 3T at preset flow rates of 0.8 to 2.8Ā l/min using velocity encoding of both 1.0 and 2.0Ā m/s. Flow rates and velocities were measured at predefined levels in the portal vasculature and within the stent graft. Accuracy of 4D flow CMR was assessed through linear regression with reference measurements obtained by flow sensors and two-dimensionalĀ (2D) phase contrast (PC) CMR. Intra- and interobserver agreement were assessed through Blandā€“Altman analyses.

Results

At a velocity encoding of 2.0Ā m/s, 4D flow CMR-derived flow rates and velocities showed an excellent correlation with preset flow rates and 2D PC CMR-derived flow velocities at all vascular levels and within the stent graft (all rā€‰ā‰„ā€‰0.958, pā€‰ā‰¤ā€‰0.003). At a velocity encoding of 1.0Ā m/s, aliasing artifacts were present within the stent graft at flow ratesā€‰ā‰„ā€‰2.0Ā l/min. 4D flow CMR-derived measurements revealed high intra- and interobserver agreement.

Conclusions

The in vitro accuracy and precision of 4D flow CMR is unaffected by the presence of TIPS stent grafts, suggesting that 4D flow CMR may be used to monitor TIPS patency in patients with liver cirrhosis.

Background

Portal hypertension is a potentially fatal complication of liver cirrhosis [1]. It is characterized by an increased portosystemic pressure gradient resulting in refractory ascites and gastroesophageal varices [2]. Implantation of a transjugular intrahepatic portosystemic shunt (TIPS) is an angiographic intervention for relieving the pressure in the portal system [3,4,5]. The TIPS stent graft diverts portal flow into the systemic circulation, thereby decreasing portal hypertension with subsequent reduction of ascites and risk of variceal bleeding [6, 7]. Possible complications of TIPS placement include excessive portosystemic shunting with the risk of TIPS-induced hepatic encephalopathy or TIPS dysfunction due to in-stent stenosis with recurrent portal hypertension [8]. For these reasons, comprehensive hemodynamic monitoring of the portal system and the TIPS stent graft is needed to monitor stent patency [9, 10].

Flow velocities are commonly assessed by Doppler ultrasonography (US) to monitor TIPS patency [3, 11]. However, Doppler US is limited by operator dependence and potentially reduced image quality due to gas-containing viscus or obesity [12, 13], precluding sensitive prediction of TIPS patency [14]. Therefore, a noninvasive, accurate, and operator-independent technique for comprehensive monitoring of portal blood flow and TIPS patency is desirable.

Volumetric blood flow is an alternative measure that can be used to assess TIPS function [15, 16]. Four-dimensional (4D) flow cardiovascular magnetic resonance (CMR) is a noninvasive technique that facilitates time-resolved, three-dimensional (3D)Ā quantitative assessment of hepatic blood flow in a single acquisition with low interobserver variability and good reproducibility [17,18,19]. Previous pilot studies have shown the feasibility of 4D flow CMR for noninvasive monitoring of hepatic blood flow after TIPS placement in vivo [20,21,22]. Unfortunately, the potential of the TIPS stent graft mesh to induce artifacts and to influence the accuracy of 4D flow CMR-derived measurements has not been assessed due to the lack of appropriate reference standards. Adequate validation of 4D flow CMR-derived blood flow rates and flow velocities within TIPS stent grafts has not been performed. However, such validation is a prerequisite for further prospective studies and clinical acceptance of 4D flow CMR for monitoring of TIPS patency as well as for prediction of hepatic encephalopathy and refractory ascites after TIPS implantation.

Therefore, the purpose of our study was to validate 4D flow CMR-derived flow measurements in TIPS stent grafts using a 3D-printed flow phantom.

Methods

Flow phantom and experimental setup

A flow phantom mimicking the portal vasculature was created with a 3DĀ printer using Clear Resin (Form 3, Formlabs Inc., Somerville, Massachusetts, USA). The phantom consisted of the superior mesenteric vein (SMV, Ƙ10 mm) and the splenic vein (SV, Ƙ10 mm) draining into the extrahepatic portal vein (PV, Ƙ15 mm), the intrahepatic TIPS-tract, and the hepatic vein (Fig.Ā 1A). Vessel diameters were chosen within the range of previously reported values [23,24,25,26].

Fig. 1
figure 1

Flow phantom of the portal vasculature with transjugular intrahepatic portosystemic shunt (TIPS) stent graft. A Cross sectional rendering of the 3D phantom illustrating the design of the flow phantom mimicking the portal vasculature after TIPS implantation. The superior mesenteric vein (SMV) and the splenic vein (SV) serve as inlets. SMV and SV converge to the portal vein (PV), which narrows towards the TIPS tract. The TIPS tract drains into the hepatic vein (Vein) serving as outlet. White arrows indicate the flow direction. B Photograph of the translucent, 3D-printed phantom with the TIPS stent graft (GoreĀ® ViatorrĀ®) placed within the TIPS tract. The covered part of the TIPS stent graft is located within the TIPS tract (arrowhead). The bare part of the TIPS stent is located within the PV (arrow). The phantom was placed in the isocenter of a 3Ā T CMR system. The inlets of the phantom were connected via flexible tubes over two separate flow regulators and flow sensors to two pumps and the system was filled with blood-mimicking fluid

A ViatorrĀ® stent graft (W.L. Gore & Associates, Newark, Delaware, USA) was positioned within the TIPS-tract and PV. The stent graft had a diameter of 10Ā mm and measured 90Ā mm in length. The TIPS stent graft consisted of an expanded polytetrafluoroethylene covered part (70Ā mm) that was placed in the TIPS-tract and an uncovered part (20Ā mm) that was placed in the PV (Fig.Ā 1B).

SMV and SV served as inlets and were connected via flexible tubes over two separate flow regulators and turbine flow sensors (FCH-midi-POM, B.I.O-TECH e.K., Vilshofen an der Donau, Germany) to two pumps. The system was filled with blood-mimicking fluid, consisting of a glycerol-water mixture with a volume fraction of glycerol of 0.331, resulting in a dynamic viscosity of 3.45āˆ™10ā€“3Ā Paāˆ™s at 20Ā° C, similar to the viscosity of human blood at 37Ā Ā°C [27, 28]. The reservoir around the vessels was filled with water to reduce susceptibility artifacts.

Flow rates were preset to 0.8, 1.2, 1.6, 2.0, 2.4, and 2.8Ā l/min and continuously monitored with the two flow sensors at the SMV and SV inlets.

CMR imaging

CMR imaging was performed on a 3Ā T system (Ingenia, Philips Healthcare, Best, The Netherlands) with a 32-channel body phased array coil. An electrocardiogram (ECG) was simulated at a heart rate of 60 /min to perform ECG-triggered acquisitions for 4D flow CMR and two-dimensional (2D)Ā phase contrast (PC) CMR.

The following cartesian 4D flow CMR acquisition parameters were applied: field of view, 240ā€‰Ć—ā€‰320ā€‰Ć—ā€‰187.5Ā mm3 (FHā€‰Ć—ā€‰APā€‰Ć—ā€‰RL); acquired/reconstructed isotropic spatial resolution, 2.5Ā mm/1.25Ā mm; temporal resolution, 67Ā ms; repetition time/echo time, 3.5Ā ms/2.1Ā ms; flip angle, 4Ā°; parallel imaging, SENSE (acceleration factor 4). Imaging time for each acquisition was 8Ā min 42Ā s. 4DĀ flowĀ CMR acquisitions covered the entire phantom.

The following 2D PC CMR acquisition parameters were applied: field of view, 300ā€‰Ć—ā€‰300Ā mm2 (slice thickness 8Ā mm); acquired/reconstructed isotropic in-plane resolution, 1.5Ā mm/0.75Ā mm; temporal resolution 67Ā ms (15 time frames per cardiac cycle); repetition time/echo time, 4.5Ā ms/2.2Ā ms; flip angle, 10Ā°. Imaging time for each 2D PC CMR acquisition was 23Ā s.

2D PC CMR acquisitions were performed at the following predefined levels: SMV, SV, PV, the bare part of the stent (TIPSb), each third of the covered stent (TIPSpā€‰=ā€‰portal sided third, TIPSmā€‰=ā€‰middle third, TIPSvā€‰=ā€‰venous sided third), and hepatic vein (Fig.Ā 2). 4D flow CMR and 2D PC CMR were performed with velocity encoding of 1.0Ā m/s and 2.0Ā m/s at each preset flow rate.

Fig. 2
figure 2

4D flow CMR of a TIPS stent graft within the flow phantom and within the human portal vasculature. A Velocity-coded 4D flow CMR image overlayed with a T2 weighted image of the flow phantom shows the velocity distribution in the portal circulation and TIPS stent graft, which is indicated by color-coded pathlines. Measurement levels for evaluation of flow rates and flow velocities are indicated: superior mesenteric vein (SMV), splenic vein (SV), portal vein (PV), bare part of the stent (TIPSb), each third of the covered stent (TIPSpā€‰ā€‰portal sided third of the stent, TIPSmā€‰ā€‰middle third of the stent, TIPSvā€‰ā€‰venous sided third of the stent), and vein. B 4D flow CMR imaging-based visualization of hemodynamics by color-coded pathlines in the portal system after TIPS placement in a 65-year-old woman. Blood flow is visualized within the aorta by red pathlines and within the inferior vena cava (IVC) by blue pathlines. Note the similarity of the velocity distribution in the flow phantom and the human portal vasculature after TIPS placement. Both the flow phantom and the patient reveal flow acceleration within the TIPS stent graft (asterisk)

CMR data analysis

4D flow CMR datasets were automatically reconstructed to 15 time frames per cardiac cycle. Background phase offset correction was applied and 3D angiograms were rendered [29, 30].

One radiologist (CR) with three years of experience in 4D flow CMR assessment placed analysis planes at the above-mentioned levels: SMV, SV, PV, TIPSb, TIPSp, TIPSm, TIPSv, and the hepatic vein (Fig.Ā 2). Flow rates and average through-plane velocities were quantified for each analysis plane. The software GTFlow 3.2.4 (GyroTools LLC, Zurich, Switzerland) was used for analysis of 4D flow CMR-derived datasets.

The same radiologist analyzed all 2D PC CMR datasets and quantified the average through-plane velocity using QFlow Analysis in IntelliSpace Portal 11.1 (Philips Healthcare).

Intra- and interobserver agreement

4D flow CMR-derived flow rates and velocities within the covered stent were quantified three times at a velocity encoding of 2.0Ā m/s. For assessment of intraobserver agreement, two measurements were performed by the first radiologist (CR), with an interval of six months between the first and second measurement. For assessment of interobserver agreement, a third measurement was performed by a second reader (IR) with three years of experience in abdominal imaging.

Statistical analysis

Preset flow rates monitored with flow sensors and 2D PC CMR-derived flow velocities served as standard of reference for 4D flow CMR-derived measurements. Measurements were excluded if velocities exceeded the preset velocity encoding and caused phase velocity aliasing artifacts. 4D flow CMR-derived parameters and standards of reference were compared using Blandā€“Altman analyses and Pearsonā€™s correlation coefficient. A correlation was defined as strong if rā€‰>ā€‰0.8 and as excellent if rā€‰>ā€‰0.9 [31]. Pā€‰<ā€‰0.05 indicated statistical significance. Blandā€“Altman analyses were performed to assess intra- and interobserver agreement of flow measurements within the stent graft. Data analysis was performed by ExcelĀ® (version 2201, MicrosoftĀ®, Redmond, Washington, USA) and SPSS Ā (version 20, Statistical Package for the Social Sciences, International Business Machines, Inc., Armonk, New York, USA).

Results

4D flow CMR was successfully performed at all preset flow rates of 0.8, 1.2, 1.6, 2.0, 2.4, and 2.8Ā l/min (Fig.Ā 3).

Fig. 3
figure 3

4D flow CMR and 2D PC CMR of the flow phantom at different flow rates. Preset flow rates were increased from 0.8 to 2.8Ā l/min (left to right). Comparative flow measurements were performed using velocity encoding both at 1Ā m/s (upper rows) and at 2Ā m/s (lower rows). Velocity-coded 4D flow CMR show velocity distribution in the portal system and TIPS stent graft. Color-coded pathlines demonstrate increasing flow velocities with increase of the preset flow rates. Measurement planes indicate the mid of the TIPS stent graft and correspond to the cross-sectional phase-contrast images shown for both 4D flow CMR (upper right corners) and 2D PC CMR (lower rows). Note that aliasing (arrows) occurred at higher flow rates and velocities when velocity encoding was set at 1Ā m/s. Aliasing was absent for all flow rates when velocity encoding set at 2Ā m/s

4D flow CMR-derived datasets with velocity encoding of 1.0Ā m/s could be analyzed completely at flow ratesā€‰ā‰¤ā€‰1.6Ā l/min. At flow ratesā€‰ā‰„ā€‰2.0Ā l/min, aliasing occurred within the covered part of the stent graft and precluded 4D flow CMR-derived measurements of flow rates and flow velocities within the stent graft (Fig.Ā 4A; Additional file 1). Eight measurements had to be excluded due to aliasing. With 2D PC CMR, aliasing occurred only at the highest flow rate of 2.8Ā l/min (Fig.Ā 3).

Fig. 4
figure 4

Comparison of low and high velocity encoded 4D flow CMR of the TIPS stent graft. 4D flow CMR performed at the maximum flow rate of 2.8Ā l/min resulted in flow acceleration within the TIPS stent graft above 1Ā m/s. A Velocity encoding at 1Ā m/s resulted in aliasing within the TIPS stent graft with consecutive erroneous calculation of pathlines as indicated by pathlines perpendicular to the fluid flow (arrow). BĀ Velocity encoding at 2Ā m/s showed no aliasing within the TIPS stent graft and resulted in correct calculation of velocity-coded pathlines. Note that aliasing with erroneous calculation of pathlines results also in erroneous quantification of flow rates and flow velocities. SMVā€‰ā€‰superior mesenteric vein, SVā€‰ā€‰splenic vein, PVā€‰ā€‰portal vein

At a velocity encoding of 2.0Ā m/s, 4D flow CMR-derived datasets and 2D PC CMR analysis planes could be completely analyzed at all preset flow rates (0.8 to 2.8Ā l/min) (Fig.Ā 4B; Additional file 2; Additional file 3).

Validation of 4D flow CMR-derived flow rates

4D flow CMR-derived flow rates at all vascular levels and within the stent graft showed excellent correlation with preset flow rates (0.8 to 2.8Ā l/min), which served as the standard of reference (Fig.Ā 5).

Fig. 5
figure 5

Correlation of 4D flow CMR-derived flow rates and preset flow rates. Preset TIPS flow rates were increased from 0.8 to 2.8Ā l/min and flow measurements were performed using velocity encoding both at 1Ā m/s (black crosses) and at 2Ā m/s (red dots). Flow measurements were performed at indicated levels. 4D flow CMR-derived flow rates were significantly correlated with the preset flow rates at all vascular levels and within the stent graft (all rā€‰ā‰„ā€‰0.985 and all pā€‰ā‰¤ā€‰0.049). At velocity encoding of 1Ā m/s, flow measurements were excluded if velocities exceeded velocity encoding and caused aliasing artifacts. At velocity encoding of 2Ā m/s, aliasing did not occur. SMVā€‰ā€‰superior mesenteric vein, SVā€‰ā€‰splenic vein, PVā€‰ā€‰portal vein, TIPSbā€‰ā€‰bare part of the stent, TIPSpā€‰ā€‰portal sided third of the stent, TIPSmā€‰ā€‰middle third of the stent, TIPSvā€‰ā€‰venous sided third of the stent

At a velocity encoding of 1.0Ā m/s, the Pearson correlation coefficient was rā€‰ā‰„ā€‰0.997 for flow rate measurements in the portal vessels (i.e., SMV, SV, PV) and the hepatic vein (all pā€‰<ā€‰0.001). The difference between the 4D flow CMR-derived flow rates and preset flow rates was 0.003Ā l/min (limits of agreement: āˆ’Ā 0.063ā€“0.070Ā l/min) in these vessels. Within the stent graft, the Pearson correlation coefficient was rā€‰ā‰„ā€‰0.993 at all levels (i.e., TIPSb, pā€‰<ā€‰0.001; TIPSp, pā€‰=ā€‰0.007; TIPSm, pā€‰=ā€‰0.049; and TIPSv, pā€‰=ā€‰0.003). The difference between 4D flow CMR-derived flow rates and preset flow rates was āˆ’Ā 0.051Ā l/min (limits of agreement: āˆ’Ā 0.163ā€“0.061Ā l/min) for in-stent measurements.

At a velocity encoding of 2.0Ā m/s, the Pearson correlation coefficient was rā€‰ā‰„ā€‰0.992 for flow rate measurements in the portal vessels and the hepatic vein (all pā€‰<ā€‰0.001). TheĀ difference between the 4D flow CMR-derived flow rates and preset flow rates was 0.030Ā l/min (limits of agreement: āˆ’Ā 0.068ā€“0.128Ā l/min) in these vessels. Within the stent graft, the Pearson correlation coefficient was rā€‰ā‰„ā€‰0.985 at all levels (all pā€‰<ā€‰0.001). The difference between 4D flow CMR-derived flow rates and preset flow rates was 0.023Ā l/min (limits of agreement: āˆ’Ā 0.171ā€“0.218Ā l/min) for in-stent measurements.

Validation of 4D flow CMR-derived flow velocities

4D flow CMR-derived average flow velocities at all vascular levels and within the stent graft showed excellent correlation with 2D PC CMR-derived flow velocities, which served as the standard of reference (Fig.Ā 6). 2D PC CMR-derived reference measurements ranged from flow velocities of 0.08Ā m/s in the PV and 0.23Ā m/s in the covered part of the stent at the minimal preset flow rate of 0.8Ā l/min to flow velocities of 0.30Ā m/s in the PV and 0.80Ā m/s in the covered part of the stent at the maximal preset flow rate of 2.8Ā l/min.

Fig. 6
figure 6

Correlation of 4D flow CMR- and 2D PC CMR-derived flow velocities. Preset flow rates were increased from 0.8 to 2.8Ā l/min and velocity measurements were performed using velocity encoding both at 1Ā m/s (black crosses) and at 2Ā m/s (red squares). Flow measurements were performed at indicated levels. 4D flow CMR-derived velocities were significantly correlated with 2D PC CMR-derived flow velocities at all vascular levels and within the stent graft (all rā€‰ā‰„ā€‰0.958 and all pā€‰ā‰¤ā€‰0.024). At velocity encoding of 1Ā m/s, velocity measurements were excluded if velocities exceeded velocity encoding and caused aliasing artifacts. At velocity encoding of 2Ā m/s, aliasing did not occur. SMVā€‰ā€‰superior mesenteric vein, SVā€‰ā€‰splenic vein, PVā€‰ā€‰portal vein, TIPSbā€‰ā€‰bare part of the stent, TIPSpā€‰ā€‰portal sided third of the stent, TIPSmā€‰ā€‰middle third of TIPS stent, TIPSvā€‰ā€‰venous sided third of the stent

At a velocity encoding of 1.0Ā m/s, the Pearson correlation coefficient was rā€‰ā‰„ā€‰0.995 for flow velocity measurements in the portal vessels (i.e., SMV, SV, PV) and the hepatic vein (all pā€‰<ā€‰0.001). The difference between the 4D flow CMR-derived flow velocity and the 2D PC CMR-derived flow velocity was 0.012Ā m/s (limits of agreement: āˆ’Ā 0.032ā€“0.057Ā m/s) in these vessels. Within the stent graft, the Pearson correlation coefficient was rā€‰ā‰„ā€‰0.990 at all levels (all pā€‰ā‰¤ā€‰0.024). The difference between 4D flow CMR-derived flow velocity and 2D PC CMR-derived flow velocity was āˆ’Ā 0.010Ā m/s (limits of agreement: āˆ’Ā 0.041ā€“0.022Ā m/s) for in-stent measurements.

At a velocity encoding of 2.0Ā m/s, the Pearson correlation coefficient was rā€‰ā‰„ā€‰0.958 for flow velocity measurements in the portal vessels and the hepatic vein (all pā€‰ā‰¤ā€‰0.003). The difference between the 4D flow CMR-derived flow velocity and the 2D PC CMR-derived flow velocity was 0.010Ā m/s (limits of agreement: āˆ’Ā 0.036ā€“0.057Ā m/s) in these vessels. Within the stent graft, the Pearson correlation coefficient was rā€‰ā‰„ā€‰0.985 at all levels (all pā€‰<ā€‰0.001). TheĀ difference between 4D flow CMR-derived flow velocity and 2D PC CMR-derived flow velocity was 0.008Ā m/s (limits of agreement: āˆ’Ā 0.059ā€“0.075Ā m/s) for in-stent measurements.

Intra- and interobserver agreement

Blandā€“Altman analyses of flow rate measurements within the TIPS stent graft revealed high intraobserver agreement (mean difference: 0.9%; limits of agreement: āˆ’Ā 10.7ā€“12.5%) as well as high interobserver agreement (mean difference: 0.2%; limits of agreement: āˆ’Ā 12.5ā€“12.8%) (Fig.Ā 7A, B).

Fig. 7
figure 7

Intra- and interobserver agreement of 4D flow CMR-derived flow rate and velocity measurements. A, B Blandā€“Altman plots of intraobserver agreement (0.9ā€‰Ā±ā€‰5.9%) and interobserver agreement (0.2ā€‰Ā±ā€‰6.4%) for measurements of flow rates within the covered part of the TIPS stent graft (TIPSp, TIPSm, and TIPSv) at velocity encoding of 2Ā m/s. C, D Blandā€“Altman plots of intraobserver agreement (0.4ā€‰Ā±ā€‰1.6%) and interobserver agreement (0.5ā€‰Ā±ā€‰1.5%) for measurements of flow velocities within the covered part of the TIPS stent graft at velocity encoding of 2Ā m/s. Middle solid line indicates mean bias of flow rate and flow velocity measurements. Dashed lines indicate limits of agreement

Blandā€“Altman analyses of flow velocity measurements within the TIPS stent graft revealed high intraobserver agreement (mean difference: 0.4%; limits of agreement: āˆ’Ā 2.8ā€“3.7%) as well as high interobserver agreement (mean difference: 0.5%; limits of agreement: āˆ’Ā 2.5ā€“3.6%) (Fig.Ā 7C, D).

Discussion

In this work we have successfully validated 4D flow CMR-derived flow measurements in TIPS stent grafts using a 3D-printed flow phantom. The use of the 3D-printed phantom facilitated rigorous evaluation of the accuracy of 4D flow CMR-derived flow rates and flow velocities in a model of the portal vasculature and also within a commonly used TIPS stent graft. Moreover, 4D flow CMR-derived in-stent measurements revealed high intra- and interobserver agreement.

Our phantom study was performed at flow rates ranging from 0.8 to 2.8Ā l/min. These in vitro flow rates in the TIPS stent graft were set to cover the typical range of 1.5ā€‰Ā±ā€‰1.0Ā l/min to 1.7ā€‰Ā±ā€‰0.5Ā l/min observed in previous in vivo pilot studies in TIPS patients [20, 21]. Therefore, our 4D flow CMR-based analyses using a 3D-printed flow phantom were performed under conditions that reflect portal flow rates that are commonly observed in patients with liver cirrhosis after TIPS placement.

The preset flow rates (0.8 to 2.8Ā l/min) resulted in average flow velocities ranging from 0.23Ā m/s to 0.80Ā m/s in the TIPS stent graft. These average flow velocities are not directly comparable to previous 4D flow CMR studies, reporting peak flow velocities in the range of 1.2ā€‰Ā±ā€‰0.2 to 1.3ā€‰Ā±ā€‰0.5Ā m/s [20, 21]. However, in the context of our validation study, we focused on measurement of averageĀ flow velocities, since peakĀ velocity measurements are more sensitive to both noise and positioning of region of interest on analysis planes [32]. Further, peakĀ velocities depend on the pulsatility of the flow that could not be modeled with our constant-flow phantom.

Our phantom study revealed high accuracy and high reliability of 4D flow CMR-derived flow rates and flow velocities within the TIPS stent graft as demonstrated by excellent correlation with reference standards (preset flow rates, 2D PC CMR) and high intra- and interobserver agreement. We are aware that in vitro estimations of accuracy and reliability using a 3D-printed phantom are overoptimistic compared to in vivo measurements, but nevertheless allowed for direct and rigorous validation of flow rates and velocities in TIPS stent grafts.

Comparative analyses of velocity encoding (1.0Ā m/s vs 2.0Ā m/s) revealed that peak flow velocities within the TIPS stent graft may exceed the velocity encoding at high flow rates if velocity encoding of 1.0Ā m/s was used. Consequently, aliasing artifacts occurred within the TIPS stent graft, precluding quantitative analyses. Aliasing artifacts were absent at a velocity encoding of 2.0Ā m/s. Thus, we recommend velocity encoding of 2.0Ā m/s for both qualitative and quantitative evaluation of TIPS hemodynamics in vivo, since flow might be further accelerated in patients with reduced lumen due to in-stent TIPS stenosis.

When comparing 4D flow CMR and 2D PC CMR, we observed aliasing at lower flow rates with 4D flow CMR as compared to 2D PC CMR. 2D PC CMR had a larger voxel size (1.5ā€‰Ć—ā€‰1.5ā€‰Ć—ā€‰8ā€‰=ā€‰18 mm3) compared to 4D flow CMR (2.5ā€‰Ć—ā€‰2.5ā€‰Ć—ā€‰2.5 mm3ā€‰=ā€‰15.6 mm3). The larger voxel size of 2D PC CMR leads to more intra-voxel flow averaging and might explain, at least in part, why 2D PC CMR was less prone to artifacts.

However, we believe that 4D flow CMR has several advantages compared to 2D PC CMR in patients with portal hypertension. 4D flow CMR does not only allow quantification of flow velocities and flow rates but also three-dimensional visualization of complex blood flow patterns in complex anatomical structures such as collaterals and varices [17]. Further, 4D flow CMR allows retrospective evaluation of these flow patterns off-line during post-processing [18].

Our phantom study has important potential scientific and clinical implications. Our in vitro validation may be helpful for future prospective in vivo studies. Further, this work adds confidence in 4D flow CMR-derived measurements needed for clinical acceptance of 4D flow CMR to monitor TIPS patency and to predict hepatic encephalopathy or refractory ascites after TIPS implantation.

Limitations

Our phantom study has several limitations. First, our in vitro setting did not consider artifacts that frequently hamper 4D flow CMR examinations in vivo, such as breathing artifacts and other motion-related artifacts. Also, the signal to noise ratio of in vivo acquisitions may be lower, particularly in patients with TIPS dysfunction and refractory ascites. Regardless, 4D flow CMR-derived measurements in vitro provide important external validation to characterize the hemodynamics of TIPS stent grafts in the setting of portal hypertension.

Second, the 3D printed phantom represents only a simplification of the portal vasculature. All the fluid from the PV drained into the TIPS stent graft, since we did not model the right and left intrahepatic branch of the portal system. Our phantom therefore does not allow to gauge turbulence that may occur in vivo at the portal sided end of the TIPS stent graft, where the blood flow divides into the right PV and the TIPS stent graft.

Finally, we did not model in-stent stenoses of the TIPS. In-stent stenoses can cause flow acceleration requiring velocity encodingā€‰>ā€‰2.0Ā m/s. However, using velocity encodingā€‰>ā€‰2.0Ā m/s in vivo will result in increased noise [32] and potentially in erroneous blood flow quantification in portal vessels with slow flow velocities.

The application of dual-velocity encoded 4D flow CMR for improved assessment of high flow velocities in the TIPS stent graft and slow flow velocities in the portal vasculature might help address this potential issue [33]. Consequently, our next study using the flow phantom will compare the accuracy of a dual-velocity encoded 4D flow CMR to single velocity encoded 4D flow CMR acquisitions.

Conclusions

The in vitro accuracy and precision of 4D flow CMR is unaffected by the presence of TIPS stent grafts, suggesting that 4D flow CMR may be used to monitor TIPS patency in patients with liver cirrhosis. Future prospective in vivo studies are warranted to determine the clinical role of 4D flow CMR, such as prediction of hepatic encephalopathy or refractory ascites after TIPS implantation.

Availability of data and materials

The datasets used and/or analysed during the current study are available from the corresponding author on reasonable request.

Abbreviations

2D:

Two-dimensional

3D:

Three-dimensional

4D:

Four-dimensional

CMR:

Cardiovascular magnetic resonance

ECG:

Electrocardiogram

PC:

Phase contrast

PV:

Portal vein

SMV:

Superior mesenteric vein

SV:

Splenic vein

TIPS:

Transjugular intrahepatic portosystemic shunt

TIPSb:

Bare part of the TIPSĀ stent

TIPSp:

Portal sided third of the TIPSĀ stent

TIPSm:

Middle third of the TIPSĀ stent

TIPSv:

Venous sided third of the TIPSĀ stent

US:

Ultrasonography

References

  1. Gracia-Sancho J, Marrone G, Fernandez-Iglesias A. Hepatic microcirculation and mechanisms of portal hypertension. Nat Rev Gastroenterol Hepatol. 2019;16(4):221ā€“34.

    ArticleĀ  Google ScholarĀ 

  2. Bosch J, Iwakiri Y. The portal hypertension syndrome: etiology, classification, relevance, and animal models. Hepatol Int. 2018;12(Suppl 1):1ā€“10.

    ArticleĀ  Google ScholarĀ 

  3. Rossle M. TIPS: 25 years later. J Hepatol. 2013;59(5):1081ā€“93.

    ArticleĀ  Google ScholarĀ 

  4. Garcia-Pagan JC, Saffo S, Mandorfer M, Garcia-Tsao G. Where does TIPS fit in the management of patients with cirrhosis? JHEP Rep. 2020;2(4): 100122.

    ArticleĀ  Google ScholarĀ 

  5. Strunk H, Marinova M. Transjugular intrahepatic portosystemic shunt (TIPS): pathophysiologic basics, actual indications and results with review of the literature. Rofo. 2018;190(8):701ā€“11.

    ArticleĀ  Google ScholarĀ 

  6. Piecha F, Radunski UK, Ozga AK, Steins D, Drolz A, Horvatits T, et al. Ascites control by TIPS is more successful in patients with a lower paracentesis frequency and is associated with improved survival. JHEP Rep. 2019;1(2):90ā€“8.

    ArticleĀ  Google ScholarĀ 

  7. Garcia-Pagan JC, Caca K, Bureau C, Laleman W, Appenrodt B, Luca A, et al. Early use of TIPS in patients with cirrhosis and variceal bleeding. N Engl J Med. 2010;362(25):2370ā€“9.

    ArticleĀ  CASĀ  Google ScholarĀ 

  8. Bosch J. Small diameter shunts should lead to safe expansion of the use of TIPS. J Hepatol. 2021;74(1):230ā€“4.

    ArticleĀ  CASĀ  Google ScholarĀ 

  9. Thabut D, Moreau R, Lebrec D. Noninvasive assessment of portal hypertension in patients with cirrhosis. Hepatology. 2011;53(2):683ā€“94.

    ArticleĀ  Google ScholarĀ 

  10. Stankovic Z, Blanke P, Markl M. Usefulness of 4D MRI flow imaging to control TIPS function. Am J Gastroenterol. 2012;107(2):327ā€“8.

    ArticleĀ  Google ScholarĀ 

  11. Roldan-Alzate A, Francois CJ, Wieben O, Reeder SB. Emerging applications of abdominal 4D flow MRI. AJR Am J Roentgenol. 2016;207(1):58ā€“66.

    ArticleĀ  Google ScholarĀ 

  12. Sabba C, Merkel C, Zoli M, Ferraioli G, Gaiani S, Sacerdoti D, et al. Interobserver and interequipment variability of echo-Doppler examination of the portal vein: effect of a cooperative training program. Hepatology. 1995;21(2):428ā€“33.

    CASĀ  Google ScholarĀ 

  13. Putz FJ, Verloh N, Erlmeier A, Schelker RC, Schreyer AG, Hautmann MG, et al. Influence of limited examination conditions on contrast-enhanced sonography for characterising liver lesions. Clin Hemorheol Microcirc. 2019;71(2):267ā€“76.

    ArticleĀ  CASĀ  Google ScholarĀ 

  14. Owens CA, Bartolone C, Warner DL, Aizenstein R, Hibblen J, Yaghmai B, et al. The inaccuracy of duplex ultrasonography in predicting patency of transjugular intrahepatic portosystemic shunts. Gastroenterology. 1998;114(5):975ā€“80.

    ArticleĀ  CASĀ  Google ScholarĀ 

  15. Itkin M, Trerotola SO, Stavropoulos SW, Patel A, Mondschein JI, Soulen MC, et al. Portal flow and arterioportal shunting after transjugular intrahepatic portosystemic shunt creation. J Vasc Interv Radiol. 2006;17(1):55ā€“62.

    ArticleĀ  Google ScholarĀ 

  16. Pinter SZ, Rubin JM, Kripfgans OD, Novelli PM, Vargas-Vila M, Hall AL, et al. Volumetric blood flow in transjugular intrahepatic portosystemic shunt revision using 3-dimensional Doppler sonography. J Ultrasound Med. 2015;34(2):257ā€“66.

    ArticleĀ  Google ScholarĀ 

  17. Motosugi U, Roldan-Alzate A, Bannas P, Said A, Kelly S, Zea R, et al. Four-dimensional flow MRI as a marker for risk stratification of gastroesophageal varices in patients with liver cirrhosis. Radiology. 2019;290(1):101ā€“7.

    ArticleĀ  Google ScholarĀ 

  18. Riedel C, Lenz A, Fischer L, Li J, Piecha F, Kluwe J, et al. Abdominal applications of 4D flow MRI. Rofo. 2021;193(4):388ā€“98.

    ArticleĀ  Google ScholarĀ 

  19. Stankovic Z, Jung B, Collins J, Russe MF, Carr J, Euringer W, et al. Reproducibility study of four-dimensional flow MRI of arterial and portal venous liver hemodynamics: influence of spatio-temporal resolution. Magn Reson Med. 2014;72(2):477ā€“84.

    ArticleĀ  Google ScholarĀ 

  20. Bannas P, Roldan-Alzate A, Johnson KM, Woods MA, Ozkan O, Motosugi U, et al. Longitudinal monitoring of hepatic blood flow before and after TIPS by using 4D-flow MR imaging. Radiology. 2016;281(2):574ā€“82.

    ArticleĀ  Google ScholarĀ 

  21. Stankovic Z, Rossle M, Euringer W, Schultheiss M, Salem R, Barker A, et al. Effect of TIPS placement on portal and splanchnic arterial blood flow in 4-dimensional flow MRI. Eur Radiol. 2015;25(9):2634ā€“40.

    ArticleĀ  Google ScholarĀ 

  22. Owen JW, Saad NE, Foster G, Fowler KJ. The feasibility of using volumetric phase-contrast MR imaging (4D flow) to assess for transjugular intrahepatic portosystemic shunt dysfunction. J Vasc Interv Radiol. 2018;29(12):1717ā€“24.

    ArticleĀ  Google ScholarĀ 

  23. Ogawa S, Yamamoto A, Jogo A, Nakano MM, Kageyama K, Sohgawa E, et al. Splenic vein diameter is a risk factor for the portal venous system thrombosis after partial splenic artery embolization. Cardiovasc Intervent Radiol. 2021;44(6):921ā€“30.

    ArticleĀ  Google ScholarĀ 

  24. Nakamura Y, Urashima M, Toyota N, Ono C, Iida M, Fukumoto W, et al. Non-occlusive mesenteric ischemia (NOMI): utility of measuring the diameters of the superior mesenteric artery and superior mesenteric vein at multidetector CT. Jpn J Radiol. 2013;31:737ā€“43.

    ArticleĀ  Google ScholarĀ 

  25. Stamm ER, Meier JM, Pokharel SS, Clark T, Glueck DH, Lind KE, et al. Normal main portal vein diameter measured on CT is larger than the widely referenced upper limit of 13 mm. Abdom Radiol (NY). 2016;41(10):1931ā€“6.

    ArticleĀ  Google ScholarĀ 

  26. Ozaki K, Matsui O, Kobayashi S, Sanada J, Koda W, Minami T, et al. Selective atrophy of the middle hepatic venous drainage area in hepatitis C-related cirrhotic liver: morphometric study by using multidetector CT. Radiology. 2010;257(3):705ā€“14.

    ArticleĀ  Google ScholarĀ 

  27. Cheng NS. Formula for the viscosity of a glycerol-water mixture. Ind Eng Chem Res. 2008;47(9):3285ā€“8.

    ArticleĀ  CASĀ  Google ScholarĀ 

  28. Cho YI, Kensey KR. Effects of the non-Newtonian viscosity of blood on flows in a diseased arterial vessel. Part 1: steady flows. Biorheology. 1991;28(3ā€“4):241ā€“62.

    ArticleĀ  CASĀ  Google ScholarĀ 

  29. Lenz A, Petersen J, Riedel C, Weinrich JM, Kooijman H, Schoennagel BP, et al. 4D Flow cardiovascular magnetic resonance for monitoring of aortic valve repair in bicuspid aortic valve disease. J Cardiovasc Magn Reson. 2020;22(1):29.

    ArticleĀ  Google ScholarĀ 

  30. Petersen J, Lenz A, Adam G, Reichenspurner H, Bannas P, Girdauskas E. Changes in transvalvular flow patterns after aortic valve repair: comparison of symmetric versus asymmetric aortic valve geometry. Eur J Cardiothorac Surg. 2021;59(5):1087ā€“94.

    ArticleĀ  Google ScholarĀ 

  31. Bannas P, Kramer H, Hernando D, Agni R, Cunningham AM, Mandal R, et al. Quantitative magnetic resonance imaging of hepatic steatosis: validation in ex vivo human livers. Hepatology. 2015;62(5):1444ā€“55.

    ArticleĀ  CASĀ  Google ScholarĀ 

  32. Lotz J, Meier C, Leppert A, Galanski M. Cardiovascular flow measurement with phase-contrast MR imaging: basic facts and implementation. Radiographics. 2002;22(3):651ā€“71.

    ArticleĀ  Google ScholarĀ 

  33. Nett EJ, Johnson KM, Frydrychowicz A, Del Rio AM, Schrauben E, Francois CJ, et al. Four-dimensional phase contrast MRI with accelerated dual velocity encoding. J Magn Reson Imaging. 2012;35(6):1462ā€“71.

    ArticleĀ  Google ScholarĀ 

Download references

Acknowledgements

The authors thank Shuo Zhang for his technical support.

Funding

Open Access funding enabled and organized by Projekt DEAL. We acknowledge financial support from the Open Access Publication Fund of UKE - UniversitƤtsklinikum Hamburg-Eppendorf- and DFG ā€“ German Research Foundation.Ā CR and MH were supported by ā€˜Forschungszentrum Medizintechnik Hamburgā€™ (Grant No. 04fmthh2019). SBR is supported by the National Institutes of Health (R01 DK125783).

Author information

Authors and Affiliations

Authors

Contributions

CR: conception and design of the study, methods development, data acquisition, analysis, and interpretation, drafting of manuscript and revision. IR: data acquisition and analysis, manuscript revision. AL: methods development, interpretation of data, manuscript revision. BPS, MH, FP, SBR: interpretation of data, manuscript revision. GA: conception of the study, interpretation of data, manuscript revision. PB: conception and design of the study, methods development, data acquisition, analysis, and interpretation, drafting of manuscript and revision. All authors read and approved the final manuscript.

Corresponding author

Correspondence to Christoph Riedel.

Ethics declarations

Ethics approval and consent to participate

An Institutional Review Board approval was not necessary for this phantom study.

Consent for publication

Not applicable.

Competing interests

The authors declare that they have no competing interests.

Additional information

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Supplementary Information

Additional file 1: Movie S1. Velocity-coded 4D flow CMR overlayed with a T2 weighted image of the flow phantom at a flow rate of 2.8Ā l/min using a velocity encoding of 1Ā m/s. Flow velocities within the covered part of the stent graft exceeded 1Ā m/s and resulted in aliasing artifacts with erroneous calculation of pathlines, precluding 4D flow CMR-derived measurements of flow rate and flow velocity.

Additional file 2: Movie S2. Velocity-coded 4D flow CMR overlayed with a T2 weighted image of the flow phantom at a flow rate of 2.8Ā l/min using a velocity encoding of 2Ā m/s. Flow velocities within the covered part of the stent graft did not exceed 2Ā m/s. Thus, the absence of aliasing allowed adequate 4D flow CMR-derived measurements of flow rate and flow velocity. Note the ability of 4D flow CMR for time-resolved visualization of complex blood flows such as mixing of the blood from the superior mesenteric vein and the splenic venin within the confluence venosum.

Additional file 3: Movie S3. Side-by-side comparison of 4D flow CMR-derived phase-contrast images within the portal vein and the TIPS stent using a velocity encoding of 1Ā m/s and 2Ā m/s at a flow rate of 2.8Ā l/min. At a velocity encoding of 1Ā m/s, aliasing occurred within the covered part of the TIPS stent graft while aliasing was absent within the portal vein. At a velocity encoding of 2Ā m/s, aliasing neither occurred within the portal vasculature nor the TIPS stent graft.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Riedel, C., Ristow, I., Lenz, A. et al. Validation of 4D flow cardiovascular magnetic resonanceĀ in TIPS stent grafts using a 3D-printed flow phantom. J Cardiovasc Magn Reson 25, 9 (2023). https://doi.org/10.1186/s12968-023-00920-5