In vivo validation of an ultra-high field, high temporal resolution myocardial tagging technique for assessment of diastolic function in mice
Journal of Cardiovascular Magnetic Resonance volume 15, Article number: P129 (2013)
Background
Heart failure with preserved ejection fraction accounts for approximately half of all heart failure cases and is associated with similar morbidity and mortality. Although abnormalities of diastolic function are felt to play an important role, no specific treatments have been identified for this common condition, primarily because of poor understanding of its pathophysiology. Despite development of several murine models of this disease, accurate non-invasive assessment of diastolic function has been challenging. Echocardiographic measurements have been limited by small heart sizes, rapid ventricular rates and high inter-observer variability. The aim of this study was to assess the ability of ultra-high field, high temporal resolution CMR tagging to assess diastolic function in mice, compared to the gold-standard technique of invasive pressure-volume loop analysis.
Methods
High fat diet-induced diabetic mice (n=7) and age-matched control mice (n=4) underwent tissue tagged CMR using a 14.1-Tesla, 89 mm vertical bore magnet with a 600 MHz birdcage resonator. Myocardial tagged images were obtained at the mid-ventricular short axis from end-systole to end-diastole using a spatial modulation of magnetization (SPAMM) sequence (Figure 1A and B). Tagged images were processed using a Matlab-based program to calculate maximal circumferential strain (Ecc) rate (Figure 1C, D and E). Concurrent invasive pressure-volume loop analysis was performed within 24 hours using 1.4 Fr pressure-volume catheters. The end-diastolic pressure volume relationship (EDPVR) was calculated by acquiring pressure-volume loops following acute compression of the inferior vena cava.
Results
LV ejection fraction was similar in both groups (65.8 ± 1.8 % vs. 65.3 ± 1.6 %, p=0.86). However, Ecc rate was significantly reduced in diabetic mice compared with controls (4.24 ± 0.28 1/s vs. 5.94 ± 0.19 1/s, p<0.01), indicating significant relaxation impairment during the rapid filling phase in diabetic mice (Figure 2A). Pressure-volume loop analysis showed that diabetic mice had a significantly steeper EDPVR compared with controls (0.38 ± 0.04 vs. 0.21 ± 0.03, p<0.05), confirming diastolic dysfunction in the diabetic mice (Figure 2B). There was a significant correlation between Ecc rate and EDPVR (r2 = 0.65, p<0.01, Figure 2C).
Conclusions
This study provides preliminary validation of ultra-high field, high temporal resolution CMR tagging for non-invasive assessment of diastolic function in mice compared to the gold standard of invasive pressure-volume loop analysis. This technique has potential for assessing the effects of putative therapeutic compounds on mice models of diastolic dysfunction.
Funding
This work was supported by NIH R01 and T32.
Author information
Authors and Affiliations
Rights and permissions
This article is published under license to BioMed Central Ltd. This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/2.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
About this article
Cite this article
Chung, J., Liu, H., Jeong, EM. et al. In vivo validation of an ultra-high field, high temporal resolution myocardial tagging technique for assessment of diastolic function in mice. J Cardiovasc Magn Reson 15 (Suppl 1), P129 (2013). https://doi.org/10.1186/1532-429X-15-S1-P129
Published:
DOI: https://doi.org/10.1186/1532-429X-15-S1-P129