T1 mapping of myocardium using inversion recovery cine at 3.0T
Journal of Cardiovascular Magnetic Resonance volume 16, Article number: P379 (2014)
Background
T1 mapping is useful in the diagnosis of myocardial fibrosis [1]. It is commonly performed using MOLLI [2]. However, MOLLI and its variants are sensitive to arrhythmia, tissue T2 values, off-resonance, etc. and usually underestimates T1 [3]. We propose an arrhythmia insensitive myocardial T1 mapping technique at 3T that takes less than 6 seconds.
Methods
The sequence, IR-rttfl, performs realtime turboflash acquisition after an inversion pulse to capture the recovery of inverted magnetization. It is immune to T1 unrelated issues (e.g., field inhomogeneity). Diastolic images (about 12) are selected for T1 calculation using [4]. A trigger delay (about 300-400 ms) is applied before the inversion pulse so that initial magnetization recovery was sampled in the first diastole. Parallel imaging (TGRAPPA rate 3) and asymmetric echo improve temporal resolution. The sequence was implemented on a 3T scanner (TIM Trio, Siemens). Its accuracy was tested in phantoms doped with gadolinium with known T1 values measured by spin echo. The technique was then evaluated in ten healthy volunteers (IRB approved with informed consent). T1 values from IR-rttfl and MOLLI were compared in four volunteers. Gadolinium contrast study was also performed on one healthy volunteer and one infarct patient. Imaging parameters: TR/TE = 2.3 ms/1.1 ms, flip angle = 5o, base matrix = 192, temporal resolution = 80-100 ms (subject dependent). The scan acquired 60 measurements in a breathhold (time < 6 s).
Results
Figure 1 shows IR-rttfl and a typical T1 map from the patient before and after contrast. In phantoms, IR-rttfl is more accurate than MOLLI (Table 1). In healthy volunteers, the myocardial T1s were 1254 ms ± 81 ms. In four of them, the T1 values from IR-rttfl and MOLLI were 1247 ms ± 83 ms and 1157 ms ± 68 ms respectively (deviation consistent with [3]). Partition coefficient in the healthy volunteer for IR-rttfl and MOLLI was 0.42 and 0.39 respectively. The T1 of the infarct 10 min after contrast injection was 278 ms and 352 ms for IR-rttfl and MOLLI respectively.
Conclusions
The arrhythmia insensitive IR-rttfl performs myocardial T1 mapping in a short breathhold. It is more accurate than MOLLI in phantoms. The results are comparable to MOLLI in healthy volunteers. Results from the initial contrast studies are promising.
Funding
Guangdong innovation team, Sichun province collaboration project.
References
Dass S et al: Circ Cardiovasc Imaging. 2012, 5: 726-10.1161/CIRCIMAGING.112.976738.
Messroghli DR et al: MRM. 2004, 52: 141-10.1002/mrm.20110.
Chow K et al: MRM. 2013
Deichmann R, Haase A: JMR. 1992, 96: 608-
Author information
Authors and Affiliations
Rights and permissions
Open Access This article is published under license to BioMed Central Ltd. This is an Open Access article is distributed under the terms of the Creative Commons Attribution License ( https://creativecommons.org/licenses/by/2.0 ), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly credited. The Creative Commons Public Domain Dedication waiver ( https://creativecommons.org/publicdomain/zero/1.0/ ) applies to the data made available in this article, unless otherwise stated.
About this article
Cite this article
Zhu, Y., Gao, Y. & Chung, YC. T1 mapping of myocardium using inversion recovery cine at 3.0T. J Cardiovasc Magn Reson 16 (Suppl 1), P379 (2014). https://doi.org/10.1186/1532-429X-16-S1-P379
Published:
DOI: https://doi.org/10.1186/1532-429X-16-S1-P379
Keywords
- Healthy Volunteer
- Temporal Resolution
- Myocardial Fibrosis
- Contrast Study
- Inversion Pulse