Accuracy of compressed sensing for left ventricular mass and volumes
Journal of Cardiovascular Magnetic Resonance volume 16, Article number: P381 (2014)
Background
While SSFP CMR is the gold standard for assessing left ventricular (LV) function, it requires a regular cardiac rhythm and frequent breath-holds and not all patients with cardiovascular disease are able to achieve this. It is known that Compressed Sensing (CS) greatly reduces data acquisition time however its accuracy for LV volume and mass is currently unknown. This study compares ventricular function measurements by CS to those obtained from standard SSFP cines.
Methods
Twenty healthy human subjects (9 male, 40 ± 14 years) underwent LV SSFP imaging on a MAGNETOM Skyra 3T scanner (Siemens, Germany). Three sequences were acquired (i) gold standard fully sampled SSFP (FULL) and two 2D prototype sequences featuring CS reconstruction and regularisation in space and time with acceleration factors (ii) R = 4 (R4) or (iii) R = 9.2 (R9.2). 5-8 short axis slices (thickness 6 mm, slice gap 9 mm) and three long axis slices (4-,3-,2-chamber), FOV = 260-340 mm, were acquired for each sequence. FULL images were acquired over 14 heart-beats with TE = 1.54 ms, α = 51°, 25 frames, matrix 256×256 and iPAT factor 2. R4 images were acquired over 4 beats with TE = 1.29 ms, α = 41°, 21 frames, matrix 192×143 with iPAT. R9.2 images were acquired over 2 beats (one dummy beat for steady state preparation, thereby representing 'realtime' acquisition) with TE = 1.27 ms, α = 42°, 19frames, matrix 192×129. Images were reconstructed on-line using a non-linear iterative CS method with k-t regularisation derived from a SENSE type reconstruction [1]. Ventricular volume and mass were measured by two analysts blinded to image type using CIM Version 7 and averaged.
Results
R4 produced comparable end-diastolic volume (EDV) and ejection fraction (EF) results but there were significant differences in end-systolic volume (ESV) and LV mass (LVM). R9.2 results were comparable for ESV but significantly different for EDV, EF and LVM. While statistically significant, these differences were small and consistent, and similar to other acceleration techniques [2, 3]. To quantify the clinical significance of these results, effect sizes (ES) were calculated, with only EF showing any significance difference for R4 and R9.2.
Conclusions
These in-vivo results suggest that CS may be used to efficiently assess ventricular function in normal subjects. Testing on pathological cases is required to further support this conclusion. Further optimization of the sampling pattern and image reconstruction parameters is also likely to improve image quality and reconstruction times.
References
Beck A: SIIMS. 2009, 2: 183-
Feng : MRM. 2013, 70: 64-10.1002/mrm.24440.
Young : JCMR. 2008, 10: 24-
Author information
Authors and Affiliations
Rights and permissions
Open Access This article is published under license to BioMed Central Ltd. This is an Open Access article is distributed under the terms of the Creative Commons Attribution License ( https://creativecommons.org/licenses/by/2.0 ), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly credited. The Creative Commons Public Domain Dedication waiver ( https://creativecommons.org/publicdomain/zero/1.0/ ) applies to the data made available in this article, unless otherwise stated.
About this article
Cite this article
Lydiard, S., Greiser, A., Schmidt, M. et al. Accuracy of compressed sensing for left ventricular mass and volumes. J Cardiovasc Magn Reson 16 (Suppl 1), P381 (2014). https://doi.org/10.1186/1532-429X-16-S1-P381
Published:
DOI: https://doi.org/10.1186/1532-429X-16-S1-P381
Keywords
- Left Ventricular Mass
- Compressed Sensing
- Left Ventricular Volume
- Axis Slice
- Short Axis Slice