Correlation of scar localization between cardiac magnetic resonance imaging and electro-anatomic mapping at electrophysiology study in patients with cardiac rhythm management devices

Background

Cardiac magnetic resonance imaging (CMR) can accurately delineate myocardial scar for substrate guided ventricular tachycardia (VT) ablation. Susceptibility artifact from cardiac rhythm management device (CRMD) generator and leads can significantly affect image quality limiting the evaluation of myocardial scar and foci of VT. We conducted a retrospective study to correlate scars from CMR to the voltage mapping obtained at electrophysiology study (EPS).

Methods

We studied 27 patients with CRMD who underwent CMR and EPS at Loyola University Medical Center between 11/2012 to 9/2014. Baseline demographics and clinical parameters are outlined (Table 1). A total of 432 left ventricular (LV) segments were analyzed for wall motion and late gadolinium enhancement using the AHA 17 segment model excluding the apical cap. The presence of right ventricular (RV) scar and wall motion abnormalities was also recorded. Data from the EPS included the type of access (endocardial and/or epicardial), the presence and location of scar, and the VT foci.

Table 1 Baseline clinical parameters

Results

Of the 432 LV segments, 127 (29.3%) were affected by artifact precluding assessment for myocardial scar. The mid anterior (81.4%) and basal inferior (3.7%) segments were the most and least affected by artifact respectively. CMR scar parameters were compared with EPS scar mapping (Table 2). CMR was able to identify LV scar, particularly endocardial scar, in most cases despite the presence of device-related artifact. In 17/22 patients (77.3%) with CMR identified LV scar, the LV endocardial scar location on CMR was comparable to the voltage map at EPS. In only one patient was LV endocardial scar missed on CMR in a non-evaluable segment due to artifact, although that segment did have a wall motion abnormality. Exclusion of LV endocardial scar by CMR is associated with a low diagnostic yield of endocardial mapping at EPS. In 18/22 (81.8%) patients with inducible VT, the focus of VT arose from an area delineated as scar on CMR. In the remaining 4 patients, only one patient had inducible VT unrelated to MRI scar. Of the ten RV scars identified at EPS, only 4 were detected by DE-MRI.

Table 2 Comparison of MRI and EPS parameters

Conclusions

CMR can be a valuable tool in localizing LV scar prior to EPS for ablation of VT. However, 29.3% of LV segments are not evaluable due to artifact from device generator (anterior wall in 65.4%). Regional wall motion abnormalities in non-evaluable left and right ventricular segments can guide identification of scar at EPS.

Funding

None.

Author information

Authors and Affiliations

Authors

Rights and permissions

This article is published under license to BioMed Central Ltd. This is an Open Access article distributed under the terms of the Creative Commons Attribution License (http://creativecommons.org/licenses/by/4.0), which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Fernando, R.R., Ali, N.F., Bakhos, L. et al. Correlation of scar localization between cardiac magnetic resonance imaging and electro-anatomic mapping at electrophysiology study in patients with cardiac rhythm management devices. J Cardiovasc Magn Reson 17 (Suppl 1), P244 (2015). https://doi.org/10.1186/1532-429X-17-S1-P244

Download citation

  • Published:

  • DOI: https://doi.org/10.1186/1532-429X-17-S1-P244

Keywords