Artefacts in 1.5 Tesla and 3 Tesla cardiovascular magnetic resonance imaging in patients with leadless cardiac pacemakers

  • 1Email author,
  • 1,
  • 1,
  • 2,
  • 1,
  • 1,
  • 2, 3 and
  • 1, 4
Journal of Cardiovascular Magnetic Resonance201820:47

https://doi.org/10.1186/s12968-018-0469-4

  • Received: 26 November 2017
  • Accepted: 6 June 2018
  • Published:

Abstract

Background

There are limited data on patients with leadless cardiac pacemakers (LCP) undergoing magnetic resonance imaging. The aim of this prospective, single-center, observational study was to evaluate artefacts on cardiovascular magnetic resonance (CMR) images in patients with LCP.

Methods

Fifteen patients with Micra™ LCP, implanted at least 6 weeks prior to CMR scan, were enrolled and underwent either 1.5 Tesla or 3 Tesla CMR imaging. Artefacts were categorized into grade 1 (excellent image quality), grade 2 (good), grade 3 (poor) and grade 4 (non-diagnostic) for each myocardial segment. One patient was excluded because of an incomplete CMR investigation due to claustrophobia.

Results

LCP caused an arc-shaped artefact (0.99 ± 0.16 cm2) at the right ventricular (RV) apex. Of 224 analyzed myocardial segments of the left ventricle (LV) 158 (70.5%) were affected by grade 1, 27 (12.1%) by grade 2, 17 (7.6%) by grade 3 and 22 (9.8%) by grade 4 artefacts. The artefact burden of grade 3 and 4 artefacts was significantly higher in the 3 Tesla group (3 Tesla vs 1.5 Tesla: 3.7 ± 1.6 vs 1.9 ± 1.4 myocardial segments per patient, p = 0.03). A high artefact burden was particularly observed in the mid anteroseptal, inferoseptal and apical septal myocardial segments of the LV and in the mid and apical segments of the RV. Quantification of LV function and assessment of valves were feasible in all patients. We did not observe any clinical or device-related adverse events.

Conclusion

CMR imaging in patients with LCP is feasible with excellent to good image quality in the majority of LV segments. The artefact burden is comparable small allowing an accurate evaluation of LV function, cardiac structures and valves. However, artefacts in the mid anteroseptal, inferoseptal and apical septal myocardial segments of the LV due to the LCP may impair or even exclude diagnostic evaluation of these segments. Artefacts on CMR images may be reduced by the use of 1.5 Tesla CMR scanners.

Keywords

  • Artefacts
  • Leadless cardiac pacemakers
  • Cardiovascular magnetic resonance imaging
  • 1.5 Tesla
  • 3 Tesla

Background

Cardiovascular magnetic resonance (CMR) imaging, which has become a versatile, non-invasive imaging tool, allows a comprehensive evaluation of patients with cardiovascular diseases [1]. The different CMR imaging sequences offer the assessment of myocardial function, wall motion abnormalities, viability, coronary perfusion, valves and tissue characterization [1]. Potential hazards for patients with conventional cardiac pacemakers undergoing magnetic resonance imaging (MRI) are radiofrequency-induced heating of lead tips, pacing dysfunction and changes in capture threshold [2]. Several studies have demonstrated safety and feasibility of MRI conditional cardiac pacemakers and implantable cardioverter defibrillators (ICD) [311]. Leadless cardiac pacemaker (LCP) therapy was recently introduced clinically to overcome complications in transvenous pacemaker therapy, such as lead dislogdement and perforation with pericardial effusion, pocket hematoma and device infections [1214]. The Micra™ LCP (Medtronic, Minneapolis, USA), which was investigated in this study, is a MRI conditional cardiac, single chamber pacemaker. The device sizes of 25.9 × 6.7 mm with an integrated lithium silver vanadium oxide, carbon monofluoride battery covered in titanium and is fixed with self-expanding nitinol tines in the right ventricle (RV) [14].

Pacemakers cause metallic susceptibility artefacts due to distortion of the magnetic field [15]. There are limited data about patients with LCP undergoing CMR imaging [16]. To the best of our knowledge, there are no prospective studies in the literature about artefacts on CMR imaging in patients with implanted LCP. It is unknown, whether CMR imaging provides best image quality or less artefacts using 1.5 or 3 Tesla CMR scanners in LCP patients.

Methods

Fifteen patients with an LCP (Micra™, Medtronic, Minneapolis, USA) implanted at least 6 weeks prior to CMR scan were enrolled in this prospective, single-center, observational study. Patients with other ferromagnetic implanted devices which may interact with the CMR scanner were excluded. The study participants were randomized in a 1:1 ratio into two groups: The study participants underwent CMR imaging in either a 1.5 Tesla CMR scanner (Magnetom Avanto Fit, Siemens Healthineers, Erlangen, Germany) or a 3 Tesla CMR scanner (Magnetom Skyra, Siemens Healthineers) with a maximum gradient field of 45mT/m and a slew rate of 200 T/m/s. The Micra™ LCP is a MRI conditional single-chamber cardiac pacemaker and the device sizes 25.9 × 6.7 mm. LCP were interrogated before and immediately after the CMR scan and were programmed to an asynchronous, MRI conditional pacing mode (VOO, 80 bpm) for the CMR scan. During the CMR scan patients were monitored by continuous electrocardiogram (ECG) and pulse oximetry. Blood pressure measurements were performed before and after CMR scans. An intercom was available for patient communication in the CMR scanners.

The CMR protocol was conducted according to the recommendations of the Society for Cardiovascular Magnetic Resonance (SCMR) [17]. We obtained multiple slice transversal balanced steady-state free precession (bSSFP) images for anatomical orientation and bSSFP cine images in the long axis (4- and 2-chamber view of the left ventricle (LV), LV outflow tract view, 2-chamber view of the right ventricle (RV), RV outflow tract view) and multiple short axis of the LV for function evaluation. A fast low angle shot (FLASH) gradient echo (GRE) based sequence was performed of the 4-chamber view of the LV and T1- and T2-weighted Turbo Spin Echo (TSE) sequences were obtained of the 4-chamber view and in the short axis. CMR sequences and parameters for 1.5 and 3 Tesla CMR scans are shown in Table 1. One patient was excluded from the analysis because of an incomplete CMR investigation due to claustrophobia.
Table 1

CMR scan protocols at 3 Tesla and 1.5 Tesla

3 T

TF single-shot localizer

TF single-shot localizer TRA

TF single-shot localizer 2ch

TF single-shot localizer 4ch

TF single-shot localizer SA

TF frequency-scout 4ch

TF cine 4ch

FLASH cine 4ch

TF cine SAG

TF cine 2ch

TF cine 2ch RV

TF cine LVOT

TF cine RVOT

T2 TSE DB FS SA

T2 TSE DB FS 4ch

T1 TSE DB 4ch

T1 TSE DB SA

Time of acquisition [sec]

12

45

1

1

8.6

12

9.5

11

159

9.5

9.5

11

11

9.5

9.5

11

12

Slices/slab

5, 5, 3

30

1

1

9

1

1

1

9

1

1

1

1

1

1

1

1

Slice group

3

1

1

1

1

1

1

1

1

1

1

1

1

1

1

1

1

Total slices

13

30

1

1

9

1

1

1

9

1

1

1

1

1

1

1

1

Distance factor [%]

200

20

20

20

100

20

20

20

67

20

20

20

20

50

50

50

50

Orientation

SAG

COR

TRA

TRA

2ch

4ch

SA

4ch

4ch

4ch

SA

2ch

2ch RV

LVOT

RVOT

SA

4ch

4ch

SA

Phase encoding

A> > P A> > P R> > L

A> > P

A> > P

A> > P

R> > L

A> > P

A> > P

A> > P

R> > L

A> > P

L> > R

A> > P

A> > P

R> > L

A> > P

A> > P

R> > L

FOV read [mm]

430

380

420

420

420

420

380

380

380

400

400

380

380

400

380

380

400

FOV phase [%]

90

90

100

100

100

81

81

80

84

81

81

100

100

78

78

75

75

Slice thickness [mm]

6

5

6

6

6

6

6

6

6

6

6

6

6

8

8

8

8

TR [ms]

237

297

269

260

260

269

41

50

41

41

41

41

41

700

700

700

700

TE [ms]

1.14

1.2

1.18

1.14

1.14

1.21

1.25

2.43

1.25

1.24

1.24

1.25

1.25

71

71

31

31

Concatenations

13

3

1

1

9

1

1

1

9

1

1

1

1

1

1

1

1

Flip angle

40

40

40

40

40

40

35

12

35

35

35

35

35

180

180

150

150

Matrix

230 × 256

256 × 230

512 × 460

256 × 256

256 × 256

256 × 256

256 × 208

256 × 208

256 × 206

216 × 256

208 × 256

208 × 256

256 × 256

256 × 256

400 × 512

512 × 400

256 × 192

192 × 256

Phase resolution [%]

65

80

65

65

65

66

81

80

81

81

81

81

81

80

80

80

80

Fat suppression

none

none

none

none

none

none

none

none

none

none

none

none

none

FS

FS

none

none

Dimension

2D

2D

2D

2D

2D

2D

2D

2D

2D

2D

2D

2D

2D

2D

2D

2D

2D

Bandwidth [Hz/px]

977

1085

849

977

977

781

1085

543

1085

1085

1085

1085

1085

849

849

781

781

Turbo factor

             

21

21

9

9

RF pulse type

fast

fast

fast

fast

fast

fast

fast

fast

fast

fast

fast

fast

fast

fast

fast

fast

fast

Gradient mode

fast

fast

fast

fast

fast

fast

fast

fast

fast

fast

fast

fast

fast

normal

normal

fast

fast

1,5 T

TF localizer multi iPAT

TF TRA

TF 2ch

TF 4ch

TF SA

TF cine 4ch

FLASH cine 4ch

TF cine SA

TF cine 2ch

TF cine 2ch RV

TF cine LVOT

TF cine RVOT

T2 TSE DB FS SA

T1 TSE DB SA

T2 TSE DB FS 4ch

T1 TSE DB 4ch

Time of acquisition [sec]

16

40

2.4

2.4

12

5.5

6.7

44

6.1

6.1

6.1

6.1

24

13

20

12

Slices/slab

5, 5, 3

30

1

1

9

1

1

9

1

1

1

1

1

1

1

1

Slice group

3

1

1

1

1

1

1

1

1

1

1

1

1

1

1

1

Total slices

13

30

1

1

9

1

1

9

1

1

1

1

1

1

1

1

Distance factor [%]

200

20

20

20

100

20

20

67

20

20

20

20

20

25

20

25

Orientation

SAG

COR

TRA

TRA

2ch

4ch

SA

4ch

4ch

SA

2ch

2ch RV

LVOT

RVOT

SA

SA

4ch

4ch

Phase encoding

A> > P A> > P R> > L

A> > P

A> > P

A> > P

A> > P

A> > P

R> > L

A> > P

A> > P

A> > P

A> > P

A> > P

A> > P

H> > F

R> > L

R> > L

FOV read [mm]

400

350

400

400

400

380

380

400

400

400

380

380

400

400

380

380

FOV phase [%]

100

89.8

100

100

100

83.3

83.3

85.8

99.2

99.2

99.2

99.2

75

75

75

75

Slice thk [mm]

8

5

8

6

6

6

6

6

6

6

6

6

8

8

8

8

TR [ms]

288

341

285

288

288

40

51

52

41

41

41

41

700

700

700

700

TE [ms]

1.1

1.2

1.1

1.1

1.1

1.

2.

1.2

1.2

1.2

1.2

1.2

81

29

81

29

Concatenations

1

3

1

1

1

1

1

9

1

1

1

1

1

1

1

1

Flip angle

80

70

80

80

80

70

15

70

70

70

70

70

180

180

180

180

Matrix

256 × 256

512 × 460

256 × 256

256 × 256

256 × 256

240 × 200

200 × 240

206 × 240

238 × 240

238 × 240

240 × 238

238 × 240

384 × 512

384 × 512

512 × 384

512 × 384

Phase resolution [%]

66

80

65

65

65

85

85

85

85

85

85

85

80

80

80

80

Fat suppr.

none

none

none

none

none

none

none

none

none

none

none

none

FS

none

FS

none

Dimension

2D

2D

2D

2D

2D

2D

2D

2D

2D

2D

2D

2D

2D

2D

2D

2D

Bandwidth [Hz/px]

1149

977

1149

1149

1149

947

496

947

947

947

947

947

235

305

235

305

Turbo factor

            

23

11

23

11

RF pulse type

fast

fast

fast

fast

fast

fast

fast

fast

fast

fast

fast

fast

fast

fast

fast

fast

Gradient mode

fast

fast

fast

fast

fast

fast

fast

fast

fast

fast

fast

fast

fast

fast

fast

fast

TF TrueFISP, FLASH fast low angle shot, TSE turbo spin echo, DB dark blood, FS fat saturation, iPAT integrated parallel acquisition techniques, TRA transversal, SAG sagittal, COR coronal, 2ch 2 chamber, 4ch 4 chamber, LVOT left ventricle outflow tract, RV right ventricle, RVOT right ventricle outflow tract, SA short axis

All available CMR images were reviewed by four experienced CMR observers (two cardiologists and two radiologists). In case of inter-observer differences in artefact quantification agreement was reached in case discussion. The LV was divided into 16 segments according to the American Heart Association (AHA) 16 myocardial segmentation classification system [18]. The RV was divided into 3 segments (basal, mid and apical).

LCP related artefacts were graduated as suggested by Klein-Wiele et al. [19]:
  • Grade 1: excellent image quality, no artefacts affecting myocardial segments or cardiac structures

  • Grade 2: good image quality with artefact adjacent to the myocardial segments or cardiac structures, delineation of myocardial borders may be limited, no impact on diagnostic value

  • Grade 3: artefact moderately affecting cardiac structures, less than half of the myocardial segment is superimposed by the artefact

  • Grade 4: poor image quality with significant artefact affecting more than half of the myocardial segment, non-diagnostic image

The artefact size was measured for every patient on short axis bSSFP cine images which showed the largest artefact. The artefact burden was calculated as affected myocardial segments by grade 1 (excellent), grade 2 (good), grade 3 (moderately) and grade 4 artefacts (poor image quality) divided by 224 (16 myocardial segments × 14 study participants) for the LV and divided by 42 (3 myocardial segments × 14 study participants) for the RV. The artefact burden ratio per patient was calculated as affected myocardial segments by grade 3 and 4 artefacts on 1.5 Tesla and 3 Tesla CMR images for the LV and RV.

The study design was approved by the local ethics committee and was conducted according to the Declaration of Helsinki. Written informed consent was obtained from all study participants.

Statistical analysis

Categorical parameters are described as absolute number and percentage. Continuous values are presented as means ± standard deviation or means with 95% confidence intervals (95% CI). Differences between groups involving normally distributed data were analyzed by the unpaired t test; those involving not normally distributed data, by the Mann-Whitney U test; and those involving proportions, by the chi-square test. A two-sided p-value < 0.05 was considered statistically significant. All calculations were performed with SPSS statistical software (Version 21, SPSS Inc., Chicago, Illinois, USA).

Results

Baseline characteristics

Fifteen patients were enrolled in this prospective, single-center, observational study. The 1.5 Tesla and 3 Tesla group comprised seven patients each. One study participant was excluded from the study because of an incomplete CMR investigation due to claustrophobia. The remaining 14 patients (female: n = 3) had a mean age of 77.8 ± 14.6 years and all of them had undergone implantation of a LCP at least 6 weeks prior the CMR scan. Indications for pacemaker implantation were bradycardic arrhythmias in permanent atrial fibrillation (n = 11) or third degree atrioventricular block (n = 3). Baseline characteristics of the study population including comorbidities are shown in Table 2.
Table 2

Baseline characteristics including indication for pacemaker implantation and comorbidities

Baseline characteristics (n = 14)

 Sex

3 female (21.4%)

 Age

77.8 ± 14.6 years

Indication for pacemaker implantation

 Bradycardiac arrhythmia in atrial fibrillation

11 (78.6%)

 3rd degree AV-block

3 (21.4%)

Comorbidities

 Atrial fibrillation

12 (85.7%)

 Coronary artery disease

6 (42.9%)

 Hypertension

8 (57.1%)

 Diabetes

3 (21.4%)

 Chronic kidney disease

6 (42.9%)

 Peripheral artery disease

2 (14.3%)

 History of stroke

2 (14.3%)

Image quality and artefacts

The LCP caused an arc-shaped artefact at the site of implantation of the RV apex (Fig. 1). Of 224 analyzed myocardial segments of the LV, 158 (70.5%) were affected by grade 1, 27 (12.1%) by grade 2, 17 (7.6%) by grade 3 and 22 (9.8%) by grade 4 artefacts. Of 42 analyzed segments of the RV, 20 (47.6%) were affected by grade 1, 6 (14.3%) by grade 2, 5 (11.9%) by grade 3 and 11 (26.2%) by grade 4 artefacts. Representative bSSFP cine images of the 4-chamber view with artefact grading obtained with 1.5 Tesla and 3 Tesla are shown in Fig. 1. The artefact area, which was quantified on short axis bSSFP cine images, was slightly higher but not significantly different in the 3 Tesla group compared to the 1.5 Tesla group (both groups: 0.99 ± 0.16 cm2, 3 Tesla vs 1.5 Tesla: 1.02 ± 0.19 cm2 vs 0.95 ± 0.14 cm2, p = 0.41). The artefact burden ratio per patient of affected myocardial segments by grade 3 and 4 artefacts on the short axis bSSFP cine images of the LV was significantly higher in the 3 Tesla group (3 Tesla vs 1.5 Tesla: 3.7 ± 1.6 vs 1.9 ± 1.4 myocardial segments per patient, p = 0.03). The detailed analysis of myocardial segments affected by grade 3 and grade 4 artefacts revealed a high artefact burden particularly in the mid anteroseptal, inferoseptal and apical septal myocardial segments (AHA myocardial segments 8, 9, 14) with an artefact burden of 50, 50 and 85.7% in these regions, respectively. These artefacts were more pronounced in patients undergoing CMR on 3 Tesla CMR scanners (3 Tesla vs 1.5 Tesla: 57.1% vs 42.9%. 57.1% vs 42.9, 100% vs 71.4%). There were no artefacts in the basal inferoseptal, inferior, basal to apical anterolateral and inferolateral myocardial segments (AHA myocardial segments 3, 4, 5, 6, 11, 12 and 16). Details of artefact distribution in the short axis are demonstrated in Fig. 2 and Table 3.
Fig. 1
Fig. 1

Balanced steady free precession (bSSFP) cine images of the 4 chamber view with grade 1, grade 2, grade 3 and grade 4 artefacts of the apical septal myocardial segment obtained with 1.5 Tesla and 3 Tesla. No patient in the 3 Tesla group showed a grade 1 artefact of the apical septal myocardial segment. * = artefact caused by the leadless cardiac pacemaker (LCP)

Fig. 2
Fig. 2

Artefact burden depicting the percentage of grade 3 and 4 artefacts in each myocardial wall segment in the short axis on bSSFP cine imaging sequences. Mid anteroseptal, inferoseptal and apical left ventricular (LV) septal segments (AHA segments 8, 9 and 14) and mid and apical right ventricular (RV) segments were affected particularly by artefacts of the LCP. The artefact burden was significantly higher in the 3 Tesla group (3 Tesla vs 1.5 Tesla: 3.7 ± 1.6 vs 1.9 ± 1.4 myocardial segments per patient, p = 0.03)

Table 3

Artefact burden of myocardial segments of the LV and RV according to the AHA myocardial segmentation system in the short axis (n = 14)

All [%]

Basal

Mid

Apical

RV

Artefact

1

2

3

4

5

6

7

8

9

10

11

12

13

14

15

16

basal

mid

apical

 grade 1

92.9

78.6

85.7

100

100

100

64.3

21.4

21.4

78.6

100

100

35.7

7.1

42.9

100

100

42.9

0

 grade 2

0

14.3

14.3

0

0

0

28.6

28.6

28.6

14.3

0

0

42.9

7.1

14.3

0

0

35.7

7.1

 grade 3

7.1

0

0

0

0

0

7.1

7.1

14.3

7.1

0

0

21.4

14.3

42.9

0

0

14.3

21.4

 grade 4

0

7.1

0

0

0

0

0

42.9

35.7

0

0

0

0

71.4

0

0

0

7.1

71.4

3 T

1

2

3

4

5

6

7

8

9

10

11

12

13

14

15

16

basal

mid

apical

 grade 1

85.7

71.4

85.7

100

100

100

57.1

0

14.3

57.1

100

100

0

0

14.3

100

100

42.9

0

 grade 2

0

14.3

14.3

0

0

0

28.6

42.9

28.6

28.6

0

0

71.4

0

14.3

0

0

28.6

0

 grade 3

14.3

0

0

0

0

0

14.3

14.3

14.3

14.3

0

0

28.6

0

71.4

0

0

28.6

0

 grade 4

0

14.3

0

0

0

0

0

42.9

42.9

0

0

0

0

100

0

0

0

0

100

1.5 T

1

2

3

4

5

6

7

8

9

10

11

12

13

14

15

16

basal

mid

apical

 grade 1

100

85.7

85.7

100

100

100

71.4

42.9

71.4

100

100

100

71.4

14.3

71.4

100

100

42.9

0

 grade 2

0

14.3

14.3

0

0

0

28.6

14.3

0

0

0

0

14.3

14.3

14.3

0

0

42.9

14.3

 grade 3

0

0

0

0

0

0

0

0

28.6

0

0

0

14.3

28.6

14.3

0

0

0

42.9

 grade 4

0

0

0

0

0

0

0

42.9

0

0

0

0

0

42.9

0

0

0

14.3

42.9

Quantification of LV function and volumetry were feasible in 14 patients (100%) with a mean LV ejection fraction of 49 ± 7.4%. We experienced more problems of ECG-triggering in patients undergoing 3 Tesla CMR scans (3 Tesla vs 1.5 Tesla: 3 vs 0 patients). In one 3 Tesla patient, pulse-triggering was necessary for image acquisition. Analysis of the RV was evenly impaired in both cohorts due to artefacts of the LCP severely compromising the image quality of particularly the mid and apical free wall of the RV (grade 3 and 4 artefacts RV basal: 0%, mid: 21.4%, apical: 92.9%). The artefact burden ratio of affected myocardial segments by grade 3 and 4 artefacts on short axis bSSFP cine images of the RV revealed 1.1 ± 0.4 segments per patient. Therefore, an exact quantification of RV function was not possible in both groups and the RV function was only assessed visually (normal: n = 7, mildly impaired: n = 6, moderately impaired: n = 0, severely impaired: n = 1).

The evaluation of T1 and T2 weighted images showed better image quality with smaller arc-shaped artefacts compared to bSSFP cine imaging and less grade 3 and 4 artefacts in the mid anteroseptal, inferoseptal and septal apical myocardial segments (Figs. 3 and 4). Furthermore, when comparing 4-chamber view bSSFP cine images and GRE based FLASH cine images, there was only a slightly better image quality on FLASH sequences in the mid and apical septal myocardial segments of the LV (Fig. 3).
Fig. 3
Fig. 3

Artefact burden of grade 3 and 4 artefacts of the 4 chamber view with different CMR sequences showed better image quality on T1 and T2 weighted images and only slightly less artefacts on gradient recalled echo (GRE) based fast low angle shot (FLASH) images compared to bSSFP images

Fig. 4
Fig. 4

Artefacts due to the LCP in one patient with different CMR sequences obtained with 3 Tesla in short axis orientation showing slightly more artefacts on bSSFP cine images compared to GRE cine, T1, T2 with fat saturation, T1 and T2 mapping, perfusion and LGE images. * = artefact caused by the LCP, + = artefact caused by implantable loop recorder

Aortic, mitral and tricuspid valves could be assessed in all patients (grade 3 and 4 artefacts: 0, 0, 0%, respectively). We did not observe any hemodynamically relevant tricuspid valve insufficiency in our study cohort of patients with LCP.

Safety and device integrity

There were no clinical or device-related serious adverse events during CMR scans. The CMR scans, both 1.5 and 3 Tesla, did not lead to malfunction of the implanted LCP devices. We observed no significant change of pacing thresholds (before CMR scan: 0.59 ± 0.15 V/0.24 ms, after: 0.61 ± 0.17 V/0.24 ms, p = 0.08) and a marginal, but statistically significant increase of sensing amplitude (before CMR scan: 14.9 ± 4.7 mV, after: 15.8 ± 4.5 mV, p = 0.02) and impedances (before CMR scan: 527 ± 100 Ω, after: 541 ± 110 Ω, p = 0.02). Battery voltage remained stable (before CMR scan: > 10 years, after: > 10 years).

Discussion

This prospective, single-center, observational study, demonstrates that CMR imaging in patients with LCP implanted at least 6 weeks prior to the CMR scan is feasible. Overall image quality was excellent or good in the majority of CMR images (myocardial segments of the LV affected by grade 1: 70.5%, grade 2: 12.1%, grade 3: 7.6%, grade 4: 9.8%). Artefacts occurred particularly in the mid anteroseptal, inferoseptal and septal apical myocardial segments (AHA myocardial segments 8, 9 and 14) and in the mid and apical segments of the RV. 3 Tesla CMR imaging led to a significantly higher artefact burden ratio per patient compared to 1.5 Tesla CMR imaging. Assessment of LV function and aortic, mitral and tricuspid valve, as well as tissue characterization by T1- and T2- weighted imaging was feasible with both modalities.

Several studies have demonstrated the safety and feasibility of MRI conditional, transvenous pacemakers and ICD undergoing MRI [311]. Potential adverse effects of MRI on implanted cardiac devices include: radiofrequency-induced heating of the lead tips, pacing inhibition/dysfunction, asynchronous pacing with the possibility of induction of ventricular tachyarrhythmias, change or loss of programmed data and changes in capture treshold [2]. A closer distance of the scanning area to the pacing system and a higher field strength increases this risk [2]. Therefore, pacemakers and ICD including leads must be interrogated before and after MRI scans and have to be programmed to MRI conditional pacing modes during the scan. It is recommended to monitor patients with cardiac devices with ECG, pulse oximetry and blood pressure measurements during MRI scans [2]. Soejima et al. demonstrated in an ex-vivo study the safety for the Micra™ LCP in a not perfused phantom model with a device heating of less than 0.4 °C at 1.5 Tesla and 0.5 °C at 3 Tesla MRI [20]. Furthermore they report no MRI-related complication in a clinical case study. We did not observe any clinical or device-related serious adverse events in our study cohort. Pacing threshold did not change significantly and battery voltage remained stable before and after CMR scans. However, we observed a marginal, but statistically significant increase of sensing amplitude (before CMR scan: 14.9 ± 4.7 mV, after: 15.8 ± 4.5 mV, p = 0.02) and impedances (before CMR scan: 527 ± 100 Ω, after: 541 ± 110 Ω, p = 0.02). We do not consider these marginal changes to be clinically relevant because we did not observe any malfunction of LCP during or after CMR scans. Further studies are needed to provide follow up data of LCP after MRI.

Pacemakers and other implanted cardiac electronic devices lead to metallic susceptibility artefacts due to distortion of the magnetic field [15]. Artefacts tend to be larger on 3 Tesla CMR scanners which could be confirmed by our study findings with a higher artefact burden of affected myocardial segments of the LV by grade 3 and 4 artefacts in patients undergoing CMR imaging at 3 Tesla (3 Tesla vs 1.5 Tesla: 3.7 ± 1.6 vs 1.9 ± 1.4 AHA myocardial segments, p = 0.03). These susceptibility artefacts were pronounced in particular in the mid anteroseptal, inferoseptal and apical septal AHA myocardial segments 8, 9 and 14 of the LV and were even more pronounced in patients undergoing CMR on 3 Tesla. The RV was mainly affected in the mid and apical myocardial segments. The size of the arc-shaped artefact by the LCP (0.99 ± 0.16 cm2) was not significantly different in both groups (3 Tesla vs 1.5 Tesla: 1.02 ± 0.19 cm2 vs 0.95 ± 0.14 cm2, p = 0.41). The size of the area affected by the artefact was 7-fold larger compared to the size of the LCP device itself. Quantification of LV function and volumetry were feasible in all patients (n = 14). However, the RV function could only be assessed visually, because of the high artefact burden, demonstrating 7 patients with normal, 6 patients with mildly and 1 patient with severely reduced RV function. However, the impaired RV function cannot be interpreted as a reduced RV function due to the LCP because we did not compare RV function before and after implantation of the LCP.

As described above, mid anteroseptal, inferoseptal and apical septal myocardial segments of the LV and mid and apical segments of the RV were affected by artefacts of the LCP in the majority of patients which may impair or even exclude diagnostic evaluation of these segments especially on perfusion and late gadolinum enhancement images. In contrast to our findings, Klein-Wiele et al. revealed in a population of 61 patients with MRI conditional, transvenous pacemakers no relevant artefacts in patients with right-sided devices irrespective of the imaging sequence [19]. There were no pacemaker induced artefacts in left-sided implants in first pass perfusion sequence, flow analysis and T1 weighted imaging. bSSFP cine sequences tend to have more artefact burden than late gadolinum enhancement sequences [19]. As reported by Sasaki et al. right-sided pacemakers and ICD did not cause susceptibility artefacts on CMR images but artefacts of the anterior and apical LV were described with left-sided ICD [21]. To the best of our knowledge, there are no prospective studies investigating artefacts on CMR scans in patients with LCP. We recently reported about a case of a patient with a LCP and an arc-shaped artefact at the RV apex [16]. As described by Klein-Wiele et al., pacing leads artefacts are smaller and do not usually interfere with myocardial structures which may be a potential advantage of transvenous cardiac pacemaker systems undergoing CMR imaging [19].

LCP are RV single-chamber pacemakers which are implanted by using a femoral percutaneous approach [12]. The next step of leadless technology will be dual-chamber pacing to treat patients with AV block [13]. Aurrichio et al. reported about feasibility, safety and short-term outcome of leadless ultrasound-based endocardial LV resychronization in patients with heart failure [22]. A study by Tjong FV et al. reported about leadless pacing combined with subcutaneous defibrillation therapy [23]. Therefore, the clinical use of these leadless cardiac devices will increase in the next years. Keller J et al. reported artefacts by the can of subcutaneous ICD affecting the LV [24]. This is consistent with our expierence in this patient undergoing CMR imaging that diagnostic evaluation of the LV is severely impaired by artefacts of the subcutaneous ICD.

CMR is an important non-invasive imaging tool to assess patients with cardiovascular diseases by evaluating myocardial function, wall motion abnormalities, viability, coronary perfusion, valves and tissue characterization [1]. Therefore, we were able to demonstrate that CMR is feasible in patients with LCP. Furthermore, the overall image quality was excellent to good and the artefact burden due to the LCP was comparable small and allowed a comprehensive evaluation of the cardiac structures and function, wall motion abnormalities and tissue characterization. However, mid anteroseptal, inferoseptal and apical septal myocardial segments of the LV were affected by artefacts due to the LCP which may impair or even exclude diagnostic evaluation of these segments, especially on perfusion and LGE images (Fig. 4). The artefact burden on CMR images could be significantly reduced by the use of 1.5 Tesla MRI scanners.

Limitations of this study

Besides the limited number of patients this study has further limitations. This study enrolled only patients with Micra™ LCP and the described artefacts are specific for this LCP. Further studies are needed to evaluate the artefact burden of LCP from other manufactures. However, according to our data the size of the artefact by the LCP did not differ widely with only a small standard deviation on CMR images (0.99± 0.16 cm2). Therefore, we conclude that the localization of the artefact depends mainly on the site of implantation of the LCP which is usually at the apex of the RV or at the apical or mid right interventricular septum. The impaired RV function in 7 patients of our study population cannot be interpreted as a reduced RV function due to the LCP because we did not compare RV function before and after implantation of the LCP. Furthermore, we were not able to report on late gadolinum enhancement sequences as these sequences were not part of our CMR scan protocol. As reported by Klein-Wiele, late gadolinum enhancement sequences showed a lower artefact burden compared to bSSFP cine images [19].

Conclusion

This prospective, single-center, observational study demonstrates that CMR imaging in patients with LCP implanted at least six prior to the CMR scan is feasible. Overall image quality was excellent to good in the majority of CMR images. Assessment of LV function and aortic, mitral and tricuspid valve as well as tissue characterization with T1 and T2 weighted imaging were feasible with 1.5 Tesla and 3 Tesla. However, mid anteroseptal, inferoseptal and apical septal myocardial segments of the LV were affected by artefacts of the LCP in the majority of patients which may impair or even exclude diagnostic evaluation of these segments. Artefact burden on CMR images may be reduced by the use of 1.5 Tesla scanners.

Abbreviations

AHA: 

American Heart Association

bSSFP: 

Balanced steady-state free precession

CMR: 

Cardiovascular magnetic resonance

ECG: 

Electrocardiogram

FLASH: 

Fast low angle shot

GRE: 

Gradient echo

ICD: 

Implantable cardioverter defibrillators

LCP: 

Leadless cardiac pacemakers

LV: 

Left ventricle/left ventricular

MRI: 

Magnetic resonance imaging

RV: 

Right ventricle/right ventricular

TSE: 

Turbo spin echo

Declarations

Acknowledgements

The authors want to thank Alexander Kypta MD for supporting this study and, Thomas Lukas MD, Walter Aufreiter and the dedicated staff of the MRI unit of the Kepler University Hospital for image acquisition.

Availability of data and materials

The datasets used and/or analyzed during the current study are available from the corresponding author on reasonable request.

Authors’ contributions

DK, CR, JK, HB and CS conceived the study design. DK, CR, PS and FF obtained CMR images and analyzed CMR scans. JK, HB and JK interrogated LCP before and after CMR scans. DK, CR, JK and HB analyzed the data. DK drafted the manuscript. As authors, we attest to each of our substantial contributions to the manuscript and critical revision. All authors read and approved the final manuscript.

Ethics approval and consent to participate

The study design was approved by the local ethics committee (Ethikkommission Land Oberösterreich, Kepler University Hospital Linz, B-123-16) and was conducted according to the Declaration of Helsinki. Written informed consent was obtained from all study participants.

Competing interests

CS received speakers’ honoraria from Medtronic. DK, CR, JK, PS, JK, HB and FF declare no competing interests.

Publisher’s Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Open AccessThis article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.

Authors’ Affiliations

(1)
Department of Cardiology, Kepler University Hospital Linz, Med Campus III, Krankenhausstraße 9, 4021 Linz, Austria
(2)
Department of Radiology, Kepler University Hospital Linz, Linz, Austria
(3)
Medical Faculty of the Friedrich Alexander University of Erlangen-Nürnberg, Erlangen, Germany
(4)
Department of Internal Medicine II, Paracelsus Medical University Salzburg, Salzburg, Austria

References

  1. Penell DJ. Cardiovascular magnetic resonance. Circulation. 2010;121(5):692–705.View ArticleGoogle Scholar
  2. Brignole M, Auricchio A, Baron-Esquivias G, Bordachar P, Boriani G, Breithardt OA, et al. 2013 ESC guidelines on cardiac pacing and cardiac resynchronization therapy. Europace. 2013;15(8):1070–118.View ArticlePubMedGoogle Scholar
  3. Kalb B, Indik JH, Ott P, Martin DR. MRI of patients with implanted cardiac devices. J Magn Reson Imaging. 2017; https://doi.org/10.1002/jmri.25824.
  4. Russo RJ, Costa HS, Silva PD, Anderson JL, Arshad A, Biederman RW, et al. Assessing the risks associated with MRI in patients with a pacemaker or defibrillator. N Engl J Med. 2017;376(8):755–64.View ArticlePubMedGoogle Scholar
  5. Strom JB, Whelan JB, Shen C, Zheng SQ, Mortele KJ, Kramer DB. Safety and utility of magnetic resonance imaging in patients with cardiac implantable electronic devices. Heart Rhythm. 2017;14(8):1138–44.View ArticlePubMedPubMed CentralGoogle Scholar
  6. Shah AD, Patel AU, Knezevic A, Hoskins MH, Hirsh DS, Merchant FM, et al. Clinical performance of magnetic resonance imaging conditional and nonconditional cardiac implantable electronic devices. Pacing Clin Electrophysiol. 2017;40(5):467–75.View ArticlePubMedGoogle Scholar
  7. Bertelsen L, Petersen HH, Philbert BT, Svendsen JH, Thomsen C, Vejlstrup N. Safety of magnetic resonance scanning without monitoring of patients with pacemakers. Europace. 2017;19(5):818–23.PubMedGoogle Scholar
  8. Camacho JC, Moreno CC, Shah AD, Mittal PK, Mengistu A, Lloyd MS, et al. Safety and quality of 1.5-T MRI in patients with conventional and MRI-conditional cardiac implantable electronic devices after implementation of a standardized protocol. AJR Am J Roentgenol. 2016;207(3):599–604.View ArticlePubMedGoogle Scholar
  9. Horwood L, Attili A, Luba F, Ibrahim EH, Parmar H, Stojanovska J, et al. Magnetic resonance imaging in patients with cardiac implanted electronic devices: focus on contraindications to magnetic resonance imaging protocols. Europace. 2017;19(5):812–7.PubMedGoogle Scholar
  10. Lowe MD, Plummer CJ, Mainsty CH, Linker NJ, British Heart Rhythm Society. Safe use of MRI in people with cardiac implantable electronic devices. Heart. 2015;101(24):1950–3.View ArticlePubMedGoogle Scholar
  11. Kypta A, Blessberger H, Hoenig S, Saleh K, Lambert T, Kammler J, et al. Clinical safety of an MRI conditional implantable cardioverter defibrillator system: a prospective monocenter ICD-magnetic resonance imaging feasibility study (MIMI). J Magn Reson Imaging. 2016;43(3):574–84.View ArticlePubMedGoogle Scholar
  12. Tjong FV, Reddy VY. Permanent leadless cardiac pacemaker therapy: a comprehensive review. Circulation. 2017;135(15):1458–70.View ArticlePubMedGoogle Scholar
  13. Kiehl EL, Cantillon DJ. Leadless cardiac pacing: what primary care providers and non-EP cardiologists should know. Cleve Clin J Med. 2016;83(11 Suppl 2):S24–34.View ArticlePubMedGoogle Scholar
  14. Sperzel J, Burri H, Gras D, Tjong FV, Knops RE, Hindricks G, et al. State of the art of leadless pacing. Europace. 2015;17(10):1508–13.View ArticlePubMedPubMed CentralGoogle Scholar
  15. Ferreira PF, Gatehouse PD, Mohiaddin RH, Firmin DN. Cardiovascular magnetic resonance artefacts. J Cardiovasc Magn Reson. 2013;15:41.View ArticlePubMedPubMed CentralGoogle Scholar
  16. Kypta A, Blessberger H, Kiblboeck D, Steinwender C. Three tesla cardiac magnetic resonance imaging in a patient with a leadless cardiac pacemaker system. Eur Heart J. 2017;38(34):2628.View ArticlePubMedPubMed CentralGoogle Scholar
  17. Kramer CM, Barkhausen J, Flamm SD, Kim RJ, Nagel E. Society of Cardiovascular Magnetic Resonance Board of trustees task force on standardized protocols. Standardized cardiovascular magnetic resonance (CMR) protocols 2013 update. J Cardiovasc Magn Reson. 2013;15:91.View ArticlePubMedPubMed CentralGoogle Scholar
  18. Cerqueira MD, Weissman NJ, Dilsizian V, Jacobs AK, Kaul S, Laskey WK, et al. Standardized myocardial segmentation and nomenclature for tomographic imaging of the heart. Circulation. 2002;105:539–42.View ArticlePubMedGoogle Scholar
  19. Klein-Wiele O, Garmer M, Busch M, Mateiescu S, Urbien R, Barbone G, et al. Cardiovascular magnetic resonance in patients with magnetic resonance conditional pacemaker systems at 1.5T: influence of pacemaker related artefacts on image quality including first pass perfusion, aortic and mitral valve assessment, flow measurement, short tau inversion recovery and T1-weighted imaging. Int J Cardiovasc Imaging. 2017;33(3):383–94.View ArticlePubMedGoogle Scholar
  20. Soejima K, Edmonson J, Ellingson ML, Herberg B, Wiklund C, Zhao J. Safety evaluation of a leadless pacemaker for magnetic resonance imaging use. Heart Rhythm. 2016;13(10):2056–63.View ArticlePubMedGoogle Scholar
  21. Sasaki T, Hansford R, Zviman MM, Kolandaivelu A, Bluemke DA, Berger RD, et al. Quantitative assessment of artifacts on cardiac magnetic resonance imaging of patients with pacemakers and implantable cardioverter-defibrillators. Circ Cardiovasc Imaging. 2011;4:662–70.View ArticlePubMedPubMed CentralGoogle Scholar
  22. Auricchio A, Delnoy PP, Butter C, Brachmann J, Van Erven L, Spitzer S, et al. Feasibility, safety, and short-term outcome of leadless ultrasound-based endocardial left ventricular resynchronization in heart failure patients: results of the wireless stimulation endocardially for CRT (WiSE-CRT) study. Europace. 2014;16(5):681–8.View ArticlePubMedGoogle Scholar
  23. Tjong FV, Brouwer TF, Smeding L, Kooiman KM, de Groot JR, Ligon D, et al. Combined leadless pacemaker and subcutaneous implantable defibrillator therapy: feasibility, safety, and performance. Europace. 2016;18(11):1740–7.View ArticlePubMedGoogle Scholar
  24. Keller J, Neuzil P, Vymazal J, Janotka M, Brada J, Zacek R, et al. Magnetic resonance imaging in patients with subcutaneous implantable cardioverter-defibrillator. Europace. 2015;17(5):761–6.View ArticlePubMedPubMed CentralGoogle Scholar