Pulmonary blood volume measured by cardiovascular magnetic resonance: influence of pulmonary transit time methods and left atrial volume

Abstract

Background

Increased pulmonary blood volume (PBV) is a measure of congestion and is associated with an increased risk of cardiovascular events. PBV can be quantified using cardiovascular magnetic resonance (CMR) imaging as the product of cardiac output and pulmonary transit time (PTT), the latter measured from the contrast time-intensity curves in the right and left side of the heart from first-pass perfusion (FPP). Several methods of estimating PTT exist, including pulmonary transit beats (PTB), peak-to-peak, and center of gravity (CoG). The aim of this study was to determine the accuracy and precision for these methods of quantifying the PBV, taking the left atrium volume (LAV) into consideration.

Methods

Fifty-eight participants (64 ± 11 years, 24 women) underwent 1.5 T CMR. PTT was quantified from (1) a basal left ventricular short-axis image (FPP), and (2) the reference method with a separate contrast administration using an image intersecting the pulmonary artery (PA) and the LA (CoG(PA-LA)).

Results

Compared to the reference, PBV for (a) PTB(FPP) was 14 ± 17% larger, (b) peak-peak(FPP) was 17 ± 16% larger, and (c) CoG(FPP) was 18 ± 10% larger. Subtraction of the LAV (available for n = 50) decreased overall differences to − 1 ± 19%, 2 ± 18%, and 3 ± 12% for PTB(FPP), peak-peak(FPP), and CoG(FPP), respectively. Lowest interobserver variability was seen for CoG(FPP) (− 2 ± 7%).

Conclusions

CoG(PA-LA) and FPP methods measured the same PBV only when adjusting for the LAV, since FPP inherently quantifies a volume consisting of PBV + LAV. CoG(FPP) had the best precision and lowest interobserver variability among the FPP methods of measuring PBV.

Graphical abstract

Background

Left-sided heart failure leads to congestion through increased filling pressures with distended pulmonary vasculature, which potentially could result in an increased pulmonary blood volume (PBV). Classical invasive measurements of PBV used pulmonary transit time (PTT) and found prolonged PTT in mitral stenosis [1, 2], and correlation of PTT to mean left atrial (LA) pressure [1]. We have previously shown and validated with a flow phantom how PBV can be non-invasively quantified with cardiovascular magnetic resonance (CMR) as the product of cardiac output (ml/s) from phase-contrast flow and PTT (s) from a high temporal resolution first-pass sequence [3]. In that and subsequent studies [4], PTT was measured as the transit time for an intravenous contrast bolus between the main pulmonary artery and the left atrium (PA-LA). In the phantom validation, the center-of-gravity (CoG) approach proved the most precise and reproducible [3]. This method of calculating PBV will in this work be referred to as CoG(PA-LA). The PROVE-HF study, where PBV was measured by CMR from a clinical first-pass perfusion (FPP) sequence, showed that heart failure outpatients who present with an elevated PBV indexed to body surface area (PBVI) had an increased risk of major adverse cardiac events [5]. Also, PBVI has been found to be increased in hypertrophic cardiomyopathy [6] and in adults with congenital heart disease [7]. PTT has also been found to correlate with reduced ejection fraction [8]. These findings may therefore have significant implications in the management of patients with heart failure. Using FPP for quantification of PBV would be a major advantage since FPP is readily available in the clinical setting, while the reference CoG(PA-LA) requires separate imaging and contrast injection. However, the PBV by FPP will inherently be overestimated compared to CoG(PA-LA) since the left atrial volume (LAV) is included in FPP, but not in CoG(PA-LA) (Fig. 1). Furthermore, the use of different methods for measuring PTT using FPP makes inter-study comparison cumbersome. Therefore, the purpose of this study was to determine the accuracy and precision for different methods of quantifying PBV from FPP sequences using the CoG(PA-LA) method as reference standard, taking the LAV into consideration.

Fig. 1
figure 1

Cardiovascular magnetic resonance (CMR) slices and positioning of the regions of interest. A Slice position for pulmonary transit time pulmonary artery-left atrium [CoG(PA-LA)]; B corresponding CMR image and the positioning of the regions of interest (ROIs) in the pulmonary artery (PA) in blue and the left atrium (LA) in red. C slice position for first-pass perfusion (FPP); D corresponding CMR image and the positioning of ROIs in the right ventricular outflow tract (RVOT) in blue and basal left ventricle (LV) in red. PV pulmonary veins

Methods

The regional ethical committee in Lund, Sweden, approved the study (Dnr 2017/829), which complies with the Declaration of Helsinki. All participants provided written informed consent after receiving written and oral information. Fifty-eight participants without severe valvular lesions were included, whereof 21 were healthy controls and 37 were patients referred for a clinical cardiovascular magnetic resonance (CMR). All participants underwent CMR imaging at 1.5 T (MAGNETOM Aera, Siemens Healthineers, Erlangen, Germany). Patients had different protocols depending on referral. 55/58 (95%) patients underwent stress perfusion imaging with adenosine, and the remaining 3 patients did not undergo stress perfusion. All scanning protocols included balanced steady-state free precession (bSSFP) images in the left ventricular (LV) 2-, 3-, 4-chamber and short-axis views, through-plane flow measurement in the pulmonary artery, FPP imaging at rest, and pulmonary transit time between pulmonary artery to left atrium [CoG(PA-LA)]. Retrospective electrocardiographic (ECG) gating was used.

All image analysis was done using the freely available software Segment (version 2.0 R6246, Medviso, Lund, Sweden) [9].

Cardiac output

Cardiac output (CO) was measured by CMR flow measurement in a cross section of the pulmonary trunk using a through-plane phase-contrast velocity-encoded gradient echo sequence with retrospective ECG triggering during free breathing and the number of signal averages was 1, as previously described [10]. Flow measurement was also validated using a pulsatile flow phantom resulting in − 1 ± 3% and 3 ± 1% (bias ± SD) in transversal and double-oblique slices respectively, after applying linear background correction. Typical scan parameters were: slice thickness 5 mm; frames per cardiac cycle 35; flip angle 20°; pixel size 1.5 × 1.5 mm; velocity encoding 200 cm/s; TR/TE 9.8/2.7 ms.

Left atrial volume

Cine images acquired in the short-axis plane covering the LA (no slice gap) were obtained in 50 participants. In the remaining 8 participants, no short-axis cine images of the atria were available for quantification of atrial volume. The short-axis images were delineated at atrial end systole and end diastole, and the average value was then calculated to assess the mean LAV (LAVmean) [11]. Images were acquired in end-expiratory breath-hold using bSSFP cine sequences with retrospective ECG triggering. Typical scan parameters were: slice thickness 8 mm; acquired spatial resolution 2.0 × 2.0 mm; flip angle 70°; TR/TE 41/1.1 ms.

Pulmonary transit time and pulmonary blood volume

The PTT was quantified at rest with two different sequences at least 2 min apart: (1) With CoG(PA-LA) as the reference method, and (2) with FPP used for diagnosing myocardial ischemia using a prototype sequence [12]. For both methods, 0.5 mmol/ml gadoteric acid (Dotarem, Gothia Medical, Billdal, Sweden; or Clariscan, GE Health Care, Danderyd, Sweden) was used as contrast agent and injected at 4 ml/s followed by 20 ml saline flush in the same arm.

CoG(PA-LA)

During the injection of a 2 ml contrast agent bolus, imaging was performed using a saturation recovery single-shot bSSFP with minimum saturation time sequence in an image plane intersecting the main pulmonary artery (MPA) and LA. Typical scan parameters were: slice thickness 20 mm; spatial resolution 2.7 × 2.7 mm; temporal resolution 100 ms; flip angle 50°; TR/TE 88/1.2 ms [3].

First Pass Perfusion

During the injection of a 0.05 mmol/kg body weight contrast agent bolus, a time-resolved FPP image was acquired as a part of a dual-sequence, quantitative myocardial perfusion mapping scan [12]. This image was acquired in a short-axis plane intersecting the basal LV and the right ventricular outflow tract (RVOT). Typical scan parameters were: slice thickness 8 mm; number of slices 1 (the most basal of 3); spacing between slices 21 mm; spatial resolution 1.9 × 1.9 mm; temporal resolution 1 acquisition per heart beat; flip angle 50 degrees; TR/TE 142/1.0 ms.

PBV was quantified as the product of CO from flow imaging, and PTT using the information from the bolus injection in four different ways:

  1. (1)

    Center of Gravity (CoG) between the Pulmonary Artery and Left Atrium [CoG(PA-LA)] (s) × CO (ml/s) (as previously published by Ugander et al. [3]). In short, circular fixed regions-of-interest (ROIs) were manually positioned in the MPA and LA (Fig. 1B). The PTT was defined as the time between the respective curves’ center of gravity (CoG) (Fig. 2A). PBV was calculated as CoG(PA-LA) (s) × CO (ml/s).

  2. (2)

    Pulmonary transit beats (PTB) [PTB(FPP)], (as previously published by Ricci et al. [5]): ROIs were manually positioned in the most basal of the three FPP image planes acquired. This means that the ROIs were placed in the RVOT and in the basal LV (see Fig. 1D). The number of beats (= PTB) was calculated as the number of images registered between the maximum signal intensity of the ROIs in the RVOT and the LV, respectively. PBV was calculated as the PTB (n) × RVSV (ml) from MPA flow imaging (Fig. 2B).

  3. (3)

    Peak-peak(FPP) For each time-intensity curve, the time stamp at peak signal intensity was extracted as described by Ricci et al. [6] (Fig. 2C). In contrast to the PROVE-HF method, the time difference between the peaks was used instead of the number of PTB. This means that variation in heart rate during data acquisition was taken into consideration. The peak-peak(FPP) was defined as the time between the respective curves’ peak signal intensities. PBV was calculated as peak-peak(FPP) (s) × CO (ml/s).

  4. (4)

    CoG(FPP): For each time-intensity curve, the time point for the CoG was extracted (Fig. 2D). This means that bolus-dilution mechanics were also taken into consideration. The CoG(FPP) was defined as the time between the respective curves’ center of gravity. PBV was calculated as CoG(FPP) (s) × CO (ml/s).

Fig. 2
figure 2

Measurements and pulmonary blood volume (PBV) results from one patient. A Measurement of pulmonary transit time (PTT) from the pulmonary transit time pulmonary artery-left atrium (CoG(PA-LA)). B First-pass perfusion (FPP) method from the PROVE-HF study [4] using pulmonary transit beats (PTB(FPP)). Panel C shows the FPP peak-peak method (peak-peak(FPP)). D FPP center-of-gravity (CoG(FPP)) method. The blue lines indicate contrast intensity measurement in the pulmonary artery (A) or basal right ventricle (B, C and D); red lines indicate measurement in the left atrium (A) or basal left ventricle (B, C and D). CO cardiac output; RVSV right ventricular stroke volume

PBV was indexed to body surface area (BSA), calculated using the Mosteller formula. LA volume was corrected for by subtracting LAVmean from PBV calculated with FPP methods.

Reproducibility

Independent evaluation of PBV by methods 1–4 as described above was performed by two observers with one and 12 years of CMR experience, respectively. All FPP measurements were also performed twice by the same observer.

Statistics

Data are presented as mean ± standard deviation (SD). Correlation between measurements was calculated using Pearson’s correlation coefficient. Differences between healthy controls and patients were tested with Mann–Whitney U-test. Intraclass correlation (ICC) was performed as a two-way random model with absolute agreement. Bland–Altman plots were generated with the 95% limit of agreement. Differences in SD were tested with the F-test. The F-test and the ICC analyses were performed in SPSS (version 27, Statistical Package for the Social Sciences, International Business Machines, Inc., Armonk, New York, USA). All other statistical analyses were performed in GraphPad Prism 8.3.0 for Windows (GraphPad Software, La Jolla, California, USA). A p-value < 0.05 was considered statistically significant. The following sample size calculation was used: minimum acceptable reliability (ICC) (ρ0): 0.6; expected reliability (ICC) (ρ1): 0.8; significance level (α): 0.05; two-tailed power (1 − β): 80%; number of raters/repetitions per subject (k): 2; expected dropout rate: 10%; sample size; n = 49 with 10% dropout, n including dropout = 55 [13].

Results

Table 1 shows participants’ characteristics, including patients with different cardiac diseases and patients with normal exams.

Table 1 Participant characteristics and cardiovascular magnetic resonance anatomic data

Pulmonary blood volume

Results are described in Tables 2, 3, and 4 and Figs. 3 and 4. Highest precision (lowest variability) was seen comparing CoG(FPP) to CoG(PA-LA) (bias 18 ± 10%, Figs. 3C and 4C). Adjusting for LAVmean (n = 50) resulted in differences in the offset for the FPP methods PTB(FPP), peak-peak(FPP) and CoG(FPP) (from 14 ± 17 to − 1 ± 19%, from 17 ± 16 to 2 ± 18%, and from 18 ± 10 to 3 ± 12%, respectively, Fig. 4D–F). Limits of agreement are presented in Table 3. The F-test showed statistically significant differences in the SDs for all FPP methods compared to CoG(PA-LA) except for PTB(FPP)-LAVmean and peak-peak(FPP)-LAVmean.

Table 2 Pulmonary blood volume (PBV) in absolute numbers and indexed for body surface area (PBVI) for healthy controls and patients
Table 3 Measurements of pulmonary blood volume (PBV) for all participants
Table 4 Statistical analysis
Fig. 3
figure 3

Pulmonary blood volume (PBV). Scatter plots with PBV from pulmonary transit time for pulmonary artery-left atrium (CoG(PA-LA)) as reference compared with A pulmonary transit time for pulmonary transit beats (PTB(FPP)), B pulmonary transit time for peak to peak (peak-peak(FPP)), and C pulmonary transit time for center of gravity (CoG(FPP)). Scatter plots after adjustment for left atrial volumes (LAV) for D PTB(FPP), E peak-peak(FPP), and F CoG(FPP). Dotted lines are the line of identity

Fig. 4
figure 4

Bland–Altman plots of pulmonary blood volume (PBV). Bland–Altman plots for PBV for pulmonary transit time for pulmonary artery-left atrium (CoG(PA-LA)) compared with A pulmonary transit time for pulmonary transit beats (PTB(FPP)), B pulmonary transit time for peak to peak (peak-peak(FPP)), and C pulmonary transit time for center of gravity (CoG(FPP)). Bland–Altman plots after adjustment for mean left atrial volume (LAVmean) for D PTB(FPP), E peak-peak(FPP), and F CoG(FPP). Dotted lines in Bland–Altman plots represent the mean and 95% limits of agreement

Interobserver and intraobserver variability

Results are shown in Tables 3, 4, 56 and 7. Lowest interobserver variability with Bland–Altman was seen for calculation of PBV using the CoG(FPP) method (− 2 ± 7%). PTB(FPP) and peak-peak(FPP) showed similar interobserver variability of − 2 ± 12% and 0 ± 10%, respectively.

Table 5 Intraclass correlation
Table 6 Interobserver intraclass correlation
Table 7 Intraobserver intraclass correlation

Discussion

This study is a head-to-head comparison of three different methods for quantifying the PBV using a FPP sequence, with the validated CoG(PA-LA) as the reference method. It showed that (1) quantification of PBV using FPP images yields about 15–20% higher values compared to the reference method CoG(PA-LA), mainly explained by the inclusion of LAV in the FPP approach since images are acquired in the LV instead of the LA, and (2) the CoG approach is the most robust method with regard to interobserver variability. Thus, using a CoG approach may lead to an increased precision of PBV measurements compared to using the peak-to-peak method or the number of PTB as in PROVE-HF.

Clinical relevance

Congestion in heart failure is an important prognostic factor in acute [14] and chronic heart failure [15, 16]. The degree of congestion is difficult to quantify, but proxy measures have been used to assess congestion by invasive procedures such as right-heart catheterization, chest X-ray, or clinical signs such as rales on lung auscultation or ankle edema. Congestion is defined as elevated LV diastolic pressure [17], and acute forward failure due to decreased contractility or increased afterload can lead to redistribution of fluid into the pulmonary circulation [18]. A measure of PBV could therefore represent an objective and quantitative measure of congestion. The PROVE-HF study showed worse prognosis for heart failure outpatients with an increased PBVI measured by CMR [5]. The clinical relevance of these data pertains to the possibility of quantifying PBV as a measure of congestion without any additional sequences in a routine clinical CMR perfusion examination using FPP for assessment of myocardial perfusion.

Relation to earlier studies

Several different methods have previously been used to calculate the PBV with CMR in patients [4,5,6,7,8] and there is no consensus on which method to use, making inter-study comparison difficult. To our knowledge, however, the only method that has been validated is the CoG(PA-LA) method [3], and this method was therefore used as the reference method in this study.

The present study describes how measurements of PBV with FPP are achieved with the highest accuracy and precision using CoG(PA-LA) as reference. Since the CoG approach used in the reference method has shown a smaller offset compared to peak to peak (4 ± 3% vs 10 ± 2%) in a flow phantom setting [3], adapting the CoG approach to the FPP method could potentially increase the accuracy.

Differences in PBV between FPP and CoG(PA-LA)

Calculation of a volume (V) can be done by multiplying flow (Q) and transit time (t) giving the formula V = Q × t, which is valid for a system with a single input and a single output even with parallel pathways inside [19]. Since the flow (Q) in the MPA can be readily measured by CMR [20], and the anatomical entry and exit boundaries for the flow are determined by the respective region of interest (ROI), the main challenges in calculating PBV are to define the transit time (t) and to choose the entry and exit boundaries.

Using images acquired with FPP, the PROVE-HF approach utilizes the number of heart beats (PTB) between the peaks of the time-intensity curves for each ROI [5,6,7]. PTB is multiplied by the RV stroke volume, yielding PBV using the PTB(FPP) method. Caveats are that arrhythmias or skipped beats due to missed triggering could affect the number of PTB (see Fig. 5) and that RV stroke volume is not the net volume to the pulmonary circulation as it includes regurgitant volumes of the tricuspid and pulmonary valves. Interestingly, the number of PTBs alone has been shown to correlate with several invasive hemodynamic parameters, including pulmonary capillary wedge pressure and LV end-diastolic pressure [8]. The peak-peak(FPP) method [6] measures t as the time difference between the peak values for the time-intensity curves in each ROI, which means heart rate variation is accounted for and it is less prone to miscalculations due to missed triggering. In the present study, the peak-peak(FPP) method performed similarly to the PTB(FPP) when compared to the CoG(PA-LA) method (14 ± 17% vs 17 ± 16%), which is in line with the findings by Ricci et al.[6].

Fig. 5
figure 5

Schematic figure of contrast-intensity measurement using first-pass perfusion imaging with extrasystole and missed triggering. Contrast intensity is shown on the y-axis, and number of acquired images during first-pass perfusion on the x-axis. The blue line denotes the contrast intensity in the right ventricle; red line denotes left ventricle. The green vertical line indicates an extrasystole, and the red vertical line indicates a missed triggering. Note that the number of heart beats is affected by a missed triggering (x-axis), leading to a false number of PTBs and subsequent underestimation of the pulmonary blood volume using the PTT from pulmonary transit beats. Methods using time on the x-axis and center of gravity approaches would be less affected

An alternative to the use of peak-to-peak methods is to calculate the time difference between the centroids of the respective contrast bolus arrivals, which is mathematically more correct, as previously described [19] (see Fig. 2). The CoG is defined as the time point where half of the total signal intensity has reached the ROI. The peak signal intensity can be more affected by placement of the ROI since an ROI can yield two peaks with similar signal intensity but separated in time by 2–3 heart beats (see Fig. 2, panel C for an example). This can make it difficult to know which peak value represents the true peak signal intensity and thus possibly greatly affect t, while t from the CoG method would not be affected. This is in line with the current study that shows higher precision and a lower interobserver variability for CoG(FPP) than the PROVE-HF and peak-to-peak methods.

PBV with CoG(PA-LA) and FPP use different imaging planes to measure the passage of a contrast bolus through the pulmonary circulation (Fig. 1). PBV with CoG(PA-LA) measures the blood volume between the PA-LA (Fig. 1, panels A and B). By comparison, PBV with FPP measures the blood volume between the RV and the LV, and thus includes the LA volume (Fig. 1C and D). While LAV measured with the biplane method was not found to correlate with PTT by Cao et al. [8], the prognostic value of PBV from FPP techniques might be affected by different LA sizes. Several studies have shown the relationship between enlarged LA and, for example, atrial fibrillation [21], increased end-diastolic filling pressures [22], and increased all-cause mortality for patients referred for a clinical CMR [23]. Measurements with CoG(PA-LA) results in what can be called the PBV whereas measurements with FPP methods results in what may be called PBV + LAV. We suggest that PBV assessed with FPP should be referred to as PBV + LAV since it cannot be used interchangeably with PBV assessed with CoG(PA-LA) and it should be clear to the clinician which volume has been quantified.

Both PBV and PBV + LAV are likely to carry prognostic information. Increased PBV and increased LAV are both measures of congestion and increased PBV + LAV is likely to also be a measure of congestion. Increased PBV + LAV could be due to increase PBV or increased LAV, or an increase in both. Notwithstanding, PBV + LAV is the easier method to obtain in clinical routine and more likely to be used on a routine basis. We quantified the LAV using the gold standard cine short-axis images, but LAV can also be quantified from biplane area-length or transversal slices [24] which is more available in the clinical routine.

Indexing of PBV

In this study, we indexed the PBV values to the BSA. While BSA is easy to calculate and has been used in several studies investigating PBV [5,6,7], it is not necessarily the best anthropometric variable to index against. The BSA does not necessarily reflect the pulmonary vascular system, especially when extrapolating the Mosteller formula in extreme cases of obesity, cachexia, or bodybuilding. A more physiologically appealing entity to consider would be the volume of the lungs. This, however, requires extra sequences and extra measurements for indexing. Also, this approach would have caveats such as patients with emphysema who have larger lungs with large air pockets (bullae) and relatively less lung parenchyma. Thus, there appears to be no perfect metric to index the pulmonary circulation, and BSA or body height appear to be the most practical way for indexation in clinical routine.

Limitations

Our study has several limitations. One limitation is that CoG(FPP) uses the same calculation method as CoG(PA-LA) and therefore has an inherent advantage over the other methods. The validation of the CoG(PA-LA) method, however, showed that CoG was the best method using a phantom [3], and therefore is a justifiable reference method. The measurements should ideally be evaluated in light of a true standard reference measurement of PBV. Such a measurement, however, does not exist and we believe that the CoG(PA-LA) method is the closest to a reference method. In FPP, the dynamics of signal intensity in the blood pool during the contrast bolus injection is not linear in relation to the changes in gadolinium concentration. However, previous data from phantom validation have shown excellent agreement between PBV from direct measurements (timer and beaker) and calculation from CoG(PA-LA) [3]. A large contrast bolus, meaning a high gadolinium concentration, could increase the non-linearity of the signal intensity. A high enough concentration could also lead to signal intensity saturation which would affect the absolute peaks. The same effects could happen with a very fast injection of contrast. A too slow injection speed would yield flatter but more extended curves, where recirculation of contrast could affect the curves. The first contrast bolus injection for CoG(PA-LA) would probably affect the absolute peak values for the FPP signal intensity curves. The relative timing, however, of the peak values would not be affected by the pre-bolus.

Conclusions

The most clinically useful method to assess PBV is with FPP since these data are the same as those acquired during a myocardial perfusion scan and thus comes “for free” without additional contrast or image acquisition. Quantification with CoG(FPP) calculations had lowest inter-observer variability, and best precision with similar accuracy compared to the other FPP methods when taking the LAV into account. Importantly, FPP methods include the LAV in the quantification of PBV and thus provide a summed evaluation of PBV + LAV, a notable distinction since a dilated LA alone can indicate congestion.

Availability of data and materials

Data will be provided upon request.

Abbreviations

BSA:

Body surface area

bSSFP:

Balanced steady-state free precession

CMR:

Cardiovascular magnetic resonance

CO:

Cardiac output

CoG:

Center of gravity

ECG:

Electrocardiogram

EDV:

End-diastolic volume

EF:

Ejection fraction

ESV:

End-systolic volume

FPP:

First-pass perfusion

ICC:

Intraclass correlation

LA:

Left atrium/left atrial

LAV:

Left atrial volume

LAVmean :

Mean left atrial volume

LV:

Left ventricle/left ventricular

MPA:

Main pulmonary artery

PA:

Pulmonary artery

PBV:

Pulmonary blood volume

PBVI:

Pulmonary blood volume indexed to body surface area

PTB:

Pulmonary transit beats

PTT:

Pulmonary transit time

PV:

Pulmonary vein

Q:

Flow

ROI:

Region of interest

RV:

Right ventricle/right ventricular

RVOT:

Right ventricular outflow tract

RVSV:

Right ventricular stroke volume

SD:

Standard deviation

t:

Transit time

V:

Volume

References

  1. McGaff CJ, Roveti GC, Glassman E, Milnor WR. Pulmonary blood volume in rheumatic heart disease and its alteration by isoproterenol. Circulation. 1963;27:77–84.

    Article  CAS  Google Scholar 

  2. Roy SB, Bhardwaj P, Bhatia ML. Pulmonary blood volume in mitral stenosis. Br Med J. 1965;2(5476):1466–9.

    Article  CAS  Google Scholar 

  3. Ugander M, Kanski M, Engblom H, Gotberg M, Olivecrona GK, Erlinge D, et al. Pulmonary blood volume variation decreases after myocardial infarction in pigs: a quantitative and noninvasive MR imaging measure of heart failure. Radiology. 2010;256(2):415–23.

    Article  Google Scholar 

  4. Kanski M, Arheden H, Wuttge DM, Bozovic G, Hesselstrand R, Ugander M. Pulmonary blood volume indexed to lung volume is reduced in newly diagnosed systemic sclerosis compared to normals–a prospective clinical cardiovascular magnetic resonance study addressing pulmonary vascular changes. J Cardiovasc Magn Reson. 2013;15:86.

    Article  Google Scholar 

  5. Ricci F, Barison A, Todiere G, Mantini C, Cotroneo AR, Emdin M, et al. Prognostic value of pulmonary blood volume by first-pass contrast-enhanced CMR in heart failure outpatients: the PROVE-HF study. Eur Heart J Cardiovasc Imaging. 2017;19:896–904.

    Article  Google Scholar 

  6. Ricci F, Aung N, Thomson R, Boubertakh R, Camaioni C, Doimo S, et al. Pulmonary blood volume index as a quantitative biomarker of haemodynamic congestion in hypertrophic cardiomyopathy. Eur Heart J Cardiovasc Imaging. 2019;20(12):1368–76.

    Article  Google Scholar 

  7. Ait Ali L, Aquaro GD, Peritore G, Ricci F, De Marchi D, Emdin M, et al. Cardiac magnetic resonance evaluation of pulmonary transit time and blood volume in adult congenital heart disease. J Magn Reson Imaging. 2019;50(3):779–86.

    Article  Google Scholar 

  8. Cao JJ, Li L, McLaughlin J, Passick M. Prolonged central circulation transit time in patients with HFpEF and HFrEF by magnetic resonance imaging. Eur Heart J Cardiovasc Imaging. 2018;19(3):339–46.

    Article  Google Scholar 

  9. Heiberg E, Sjogren J, Ugander M, Carlsson M, Engblom H, Arheden H. Design and validation of segment–freely available software for cardiovascular image analysis. BMC Med Imaging. 2010;10:1.

    Article  Google Scholar 

  10. Carlsson M, Andersson R, Bloch KM, Steding-Ehrenborg K, Mosen H, Stahlberg F, et al. Cardiac output and cardiac index measured with cardiovascular magnetic resonance in healthy subjects, elite athletes and patients with congestive heart failure. J Cardiovasc Magn Reson. 2012;14:51.

    Article  Google Scholar 

  11. Maceira AM, Cosín-Sales J, Roughton M, Prasad SK, Pennell DJ. Reference left atrial dimensions and volumes by steady state free precession cardiovascular magnetic resonance. J Cardiovasc Magn Reson. 2010;12(1):65.

    Article  Google Scholar 

  12. Kellman P, Hansen MS, Nielles-Vallespin S, Nickander J, Themudo R, Ugander M, et al. Myocardial perfusion cardiovascular magnetic resonance: optimized dual sequence and reconstruction for quantification. J Cardiovasc Magn Reson. 2017;19(1):43.

    Article  Google Scholar 

  13. Sample Size Calculator. https://wnarifin.github.io/ssc/ssicc.html. Accessed 02 June 2021.

  14. Rubio-Gracia J, Demissei BG, Ter Maaten JM, Cleland JG, O’Connor CM, Metra M, et al. Prevalence, predictors and clinical outcome of residual congestion in acute decompensated heart failure. Int J Cardiol. 2018;258:185–91.

    Article  Google Scholar 

  15. Lucas C, Johnson W, Hamilton MA, Fonarow GC, Woo MA, Flavell CM, et al. Freedom from congestion predicts good survival despite previous class IV symptoms of heart failure. Am Heart J. 2000;140(6):840–7.

    Article  CAS  Google Scholar 

  16. Melenovsky V, Andersen MJ, Andress K, Reddy YN, Borlaug BA. Lung congestion in chronic heart failure: haemodynamic, clinical, and prognostic implications. Eur J Heart Fail. 2015;17(11):1161–71.

    Article  CAS  Google Scholar 

  17. Gheorghiade M, Follath F, Ponikowski P, Barsuk JH, Blair JE, Cleland JG, et al. Assessing and grading congestion in acute heart failure: a scientific statement from the acute heart failure committee of the heart failure association of the European Society of Cardiology and endorsed by the European Society of Intensive Care Medicine. Eur J Heart Fail. 2010;12(5):423–33.

    Article  Google Scholar 

  18. Cotter G, Felker GM, Adams KF, Milo-Cotter O, O’Connor CM. The pathophysiology of acute heart failure–is it all about fluid accumulation? Am Heart J. 2008;155(1):9–18.

    Article  Google Scholar 

  19. Grodins FS. Basic concepts in the determination of vascular volumes by indicator-dilution methods. Circ Res. 1962;10:429–46.

    Article  CAS  Google Scholar 

  20. Carlsson M, Cain P, Holmqvist C, Stahlberg F, Lundback S, Arheden H. Total heart volume variation throughout the cardiac cycle in humans. Am J Physiol Heart Circ Physiol. 2004;287(1):H243–50.

    Article  CAS  Google Scholar 

  21. Tsang TS, Barnes ME, Bailey KR, Leibson CL, Montgomery SC, Takemoto Y, et al. Left atrial volume: important risk marker of incident atrial fibrillation in 1655 older men and women. Mayo Clin Proc. 2001;76(5):467–75.

    Article  CAS  Google Scholar 

  22. Tsang TS, Barnes ME, Gersh BJ, Bailey KR, Seward JB. Left atrial volume as a morphophysiologic expression of left ventricular diastolic dysfunction and relation to cardiovascular risk burden. Am J Cardiol. 2002;90(12):1284–9.

    Article  Google Scholar 

  23. Khan MA, Yang EY, Zhan Y, Judd RM, Chan W, Nabi F, et al. Association of left atrial volume index and all-cause mortality in patients referred for routine cardiovascular magnetic resonance: a multicenter study. J Cardiovasc Magn Reson. 2019;21(1):4.

    Article  Google Scholar 

  24. Wandelt LK, Kowallick JT, Schuster A, Wachter R, Stumpfig T, Unterberg-Buchwald C, et al. Quantification of left atrial volume and phasic function using cardiovascular magnetic resonance imaging-comparison of biplane area-length method and Simpson’s method. Int J Cardiovasc Imaging. 2017;33(11):1761–9.

    Article  Google Scholar 

Download references

Acknowledgements

We thank Ann-Helen Arvidsson and Christel Carlander for help with data collection. We also thank Sebastian Bidhult for developing the protocol for PTT acquisition, Dr. Kelvin Chow, Siemens Healthineers, Erlangen, Germany for providing the SASHA sequence from which the protocol for the PTT acquisition was made possible in-house, Dr. Peter Kellman, National Instutites of Health, Bethesda, Maryland, USA for providing the FPP sequence, and Johannes Töger for performing the 2D flow validation.

Funding

Open access funding provided by Lund University. This study was supported in part by the Swedish Research Council; the Swedish Heart–Lung Foundation; Lund University Faculty of Medicine; Skåne University Hospital, Lund, Sweden; Region Skåne (all funding HA).

Author information

Authors and Affiliations

Authors

Contributions

AN, MK, MC and HA designed the study. AN and MK acquired the data and performed data analysis. All authors contributed significantly to writing the manuscript. All authors read and approved the final manuscript.

Corresponding author

Correspondence to Håkan Arheden.

Ethics declarations

Ethics approval and consent to participate

The regional ethical committee in Lund, Sweden approved the study (Dnr 2017/829), which complies with the Declaration of Helsinki. All participants provided written informed consent after receiving written and oral information.

Consent for publication

Not applicable.

Competing interests

The authors declare that they have no competing interests.

Additional information

Publisher's Note

Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Rights and permissions

Open Access This article is licensed under a Creative Commons Attribution 4.0 International License, which permits use, sharing, adaptation, distribution and reproduction in any medium or format, as long as you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons licence, and indicate if changes were made. The images or other third party material in this article are included in the article's Creative Commons licence, unless indicated otherwise in a credit line to the material. If material is not included in the article's Creative Commons licence and your intended use is not permitted by statutory regulation or exceeds the permitted use, you will need to obtain permission directly from the copyright holder. To view a copy of this licence, visit http://creativecommons.org/licenses/by/4.0/. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated in a credit line to the data.

Reprints and permissions

About this article

Check for updates. Verify currency and authenticity via CrossMark

Cite this article

Nelsson, A., Kanski, M., Engblom, H. et al. Pulmonary blood volume measured by cardiovascular magnetic resonance: influence of pulmonary transit time methods and left atrial volume. J Cardiovasc Magn Reson 23, 123 (2021). https://doi.org/10.1186/s12968-021-00809-1

Download citation

  • Received:

  • Accepted:

  • Published:

  • DOI: https://doi.org/10.1186/s12968-021-00809-1

Keywords