- Research
- Open access
- Published:
Two-center clinical validation and quantitative assessment of respiratory triggered retrospectively cardiac gated balanced-SSFP cine cardiovascular magnetic resonance imaging in adults
Journal of Cardiovascular Magnetic Resonance volume 20, Article number: 44 (2018)
Abstract
Background
Breath-hold (BH) requirement remains the limiting factor on the spatio-temporal resolution and coverage of the cine balanced steady-state free precession (bSSFP) cardiovascular magnetic resonance (CMR) imaging. In this prospective two-center clinical trial, we validated the performance of a respiratory triggered (RT) bSSFP cine sequence for evaluation of biventricular function.
Methods
Our study included 23 asymptomatic healthy subjects and 60 consecutive patients from Institute A (n = 39) and Institute B (n = 21) referred for a clinically indicated CMR study. We implemented a RT sequence with a respiratory synchronized drive to steady state (SS) of bSSFP signal, before the commencement of image data acquisition with prospective cardiac arrhythmia rejection and retrospectively cardiac gated reconstruction in real-time. Left (LV) and right (RV) ventricular function and LV mass were evaluated by using RT-bSSFP and conventional BH-bSSFP sequences with one cardiac cycle for SS preparation keeping all the imaging parameters identical. The performance of the sequences was evaluated by using quantitative and semi-quantitative metrics.
Results
Global LV and RV functional parameters and LV mass obtained from the RT-bSSFP and BH-bSSFP sequences were in good agreement. Quantitative metrics designed to capture fluctuation in SS signal intensity showed no significant difference between sequences. In addition, blood-to-myocardial contrast was nearly identical between sequences. The combined clinical score for image quality was excellent or good for 100% of cases with the BH-bSSFP and 83% of cases with the RT-bSSFP sequence. The de facto image acquisition time for RT-bSSFP was statistically significantly longer than that for conventional BH-bSSFP (7.9 ± 3.4 min vs. 5.1 ± 2.6 min).
Conclusions
Cine RT-bSSFP is an alternative for evaluating global biventricular function with contrast and spatio-temporal resolutions that are similar to those attained by using the BH-bSSFP sequence, albeit with a modest time penalty and a small reduction in image quality.
Background
In patients with cardiovascular disease, ventricular volume, mass, and ejection fraction (EF) are powerful predictors of prognosis [1,2,3,4,5]. In several studies of individuals with normal and abnormal ventricles, cardiovascular magnetic resonance (CMR) has been used to evaluate biventricular volume, mass, and EF with high accuracy and low inter- and intra-observer variability [6,7,8,9]. For the accurate measurement of ventricular functional indices in CMR, it is necessary to have: a) sufficient and consistent blood-to-myocardium contrast throughout the cardiac cycle to reliably define endocardial contour, b) adequate spatio-temporal resolution to accurately capture end-diastole and end-systole, and c) immunity to cardiorespiratory motion and pulsatile blood flow. In routine clinical CMR practice, cine balanced steady-state free precession (bSSFP) is a sequence of choice for evaluation of ventricular function [10] as it offers the highest signal-to-noise ratio (SNR) per unit time of all other CMR imaging sequences along with a T2/T1-weighted image contrast [11]. In order to obtain diagnostic image quality, the bSSFP sequence mandates the shortest possible repetition time (TR) to minimize banding artifacts [11], high flip angle for adequate blood to myocardial contrast [12], and uninterrupted radiofrequency (RF) excitations to maintain the magnetization at steady-state (SS) to elude contrast variability [13] and artifacts [14, 15]. In addition, cine bSSFP acquisition employs retrospective cardiac gating to obtain the entire cardiac cycle along with prospective arrhythmia rejection. In clinical practice breath-holds (BH) are used to mitigate respiratory motion artifacts. Respiratory suspension for sufficient duration is not feasible in sedated patients and patients with impaired breath-holding capacity. Therefore, it would be of clinical interest to obtain cine bSSFP images without the constraint of breath holding while retaining the necessary features described above.
Approaches to combat artifacts introduced by respiratory motion in cine CMR imaging fall in three broad categories: (1) rapid imaging with data undersampling [16,17,18]; (2) respiratory motion synchronized data acquisition using signals from an external respiratory bellows or RF navigator positioned over the lung-diaphragm interface [19,20,21,22,23]; and (3) self-gated motion compensation methods [24,25,26]. Combinations of data undersampling and self-gating are being explored for respiratory motion-state cine imaging [27]. Some of these approaches require contrast administration [18, 24, 28], or prospective cardiac gating [16, 17] and undersampling methods may require considerably longer reconstruction times and/or advanced hardware capabilities [16, 18, 24, 25, 27]. The dual RF navigator gated approach was limited to partial acquisition of the cardiac cycle [19]. In contrast, respiratory synchronized acquisition methods using bellows or diaphragmatic RF navigators permit retrospective cardiac gating and real-time reconstruction of bSSFP cine sequences, which are used in clinical practice without imposing any hardware or reconstruction burden [20,21,22,23]. Furthermore, these respiratory motion synchronized techniques allow the flexibility to be seamlessly combined with undersampling strategies such as compressed sensing that may become clinically available in the future.
We have previously demonstrated respiratory synchronized bSSFP cine acquisitions using both respiratory bellows and diaphragmatic RF navigator, in a healthy cohort study [20]. These implementations require SS to be reached before the acquisition of cardiac gated cine data to ensure uniform signal and contrast throughout the cardiac cycle. The bellows based respiratory triggered (RT) implementation was validated in a pediatric population [22]. Similar approaches using respiratory bellows and diaphragmatic RF navigator for respiratory synchronized bSSFP cine acquisition have also been reported [21, 23]. However, these implementations use the available time between the respiratory trigger and the subsequent R-top for SS preparation. This can cause artifacts because the variable time interval may not be long enough to attain SS in certain respiratory cycles [21]. In this manuscript we describe a two-center clinical validation of a bellows based RT-bSSFP in an adult population with rigorous quantitative assessment of the temporal stability of the myocardial signal as well as blood-to-myocardial contrast.
Methods
Numerical simulation
Numerical simulations were performed in MATLAB (The MathWorks™ Inc., Natick, Massachusetts, USA), to determine the minimum number of RF excitations required for myocardial tissue to reach SS in a bSSFP sequence starting with an α/2 preparation pulse, followed after a time period of TR/2 by successively alternating ±α excitation pulses every TR. Temporal evolution of myocardial magnetization from the unperturbed state to SS with uninterrupted RF excitations, characteristic of the bSSFP sequence, was numerically simulated with the following parameters: magnetic field strength, 1.5 T ± 3 ppm; TR/TE/flip angle, 3 ms/1.5 ms/65°; T1, 950 ms; T2, 50 ms. The steady state magnetization (MSS) was defined as magnetization where the rate of change of magnetization was close to zero (< 0.5%).
Cine bSSFP CMR with respiratory triggering (RT)
A commercially available BH cine bSSFP sequence with retrospective cardiac gating and prospective arrhythmia rejection was modified to implement a free-breathing RT sequence on clinical scanner (Achieva, Philips Healthcare, Best, the Netherlands). A schematic showing the building blocks of this technique, described previously [22], is shown in Fig. 1. In the RT sequence, the RF excitations started immediately after the trigger signal (could be set to an arbitrary delay from inspiration or expiration trigger point) derived from the respiratory bellows unmodified from the commercial implementation on the scanner. The data acquisition corresponding to the initial RF excitations were discarded (dummy excitations), for a minimum user-prescribed time (τ), to drive to MSS. The subsequent arrival of the first cardiac R-top (at time point after the trigger > τ) was used as a cardiac synchronization point to accept data acquisition of the segment of phase-encoding steps in k-space in a multi-phase manner during that cardiac cycle. Upon arrival of the subsequent R-top, RF excitations were terminated, and the commercially available real-time arrhythmia rejection algorithm either accepted or rejected the data. In the case of rejection, the same phase-encoding segment was re-acquired. This process was repeated until the entire k-space was acquired, after which the retrospective cardiac gating reconstruction algorithm performed the required nonlinear stretching of variable R-R intervals to reconstruct cine images.
Two-center clinical validation
The prospective clinical validation studies were performed at Institute A and Institute B in patients referred for a clinical CMR for various indications. Institute A implemented the RT sequence and Institute B, located 2000 km away from Institute A, used this investigational sequence without supervision from Institute A. Institute A recruited 23 asymptomatic subjects (10 men). The mean age of these healthy subjects was 44 years (range, 24–60 years). At Institute A, 39 consecutive patients (21 men) were enrolled in the study. The mean age of these patients was 50 years (range, 19–82 years). At Institute B, 21 consecutive patients (12 men) were enrolled in the study. The mean age of these patients was 47 years (range, 23–86 years). All patients underwent biventricular function evaluation as a part of routine clinical protocol to evaluate the following conditions: left and right ventricular (LV, RV) cardiomyopathies (31), myocardial viability (12), valvular function (9), pericarditis (7), and congenital heart disease (1). This study was approved by the Institutional Ethics Committee of Catholic Health Initiative’s Institute for Research and Innovation under evolution MRI techniques protocol and complied with the Health Insurance Portability and Accountability Act of 1996. All subjects gave written informed consent before being enrolled in the study. All healthy subject and patient imaging was performed at 1.5 T CMR scanner (Achieva, Philips Healthcare, Best, the Netherlands) with a 5-element or 32-element phased-array surface coil for signal reception and vector-cardiographic (VCG) gating, with respiratory bellows placed at mediastinum. Authors employed by the CMR scanner manufacturer were not part of the clinical patient data acquisition and image quality assessment.
Cardiovascular magnetic resonance imaging protocol
In all 83 study participants, scout images of the thoracic cavity were obtained along the three orthogonal planes by using a non-VCG-gated bSSFP technique. With these single-phase scout scans, a series of VCG-gated cine images were acquired during suspended respiration (BH technique) in the following order: a two-chamber view, a four-chamber view, and a series of 12 to 14 contiguous short-axis slices covering the entire LV from the apex to the base (the level of the mitral valve annulus). Identical imaging parameters and short-axis slice prescription were used with the RT acquisition. Patients were completely blinded to the imaging sequences and were not given any special instructions for breathing during RT technique.
The imaging parameters for both BH and RT cine bSSFP techniques were as follows: TR/TE/flip angle = 2.5–2.7 ms/1.25–1.35 ms/65°; acquired voxel size = 1.7–2.0 × 1.6–2.0 × 8 mm3; SENSitivity Encoding (SENSE) acceleration factor = 1.3–1.9; temporal resolution = 40-50 ms. In the case of BH, each cine slice was acquired during suspended respiration (6–8 R-R intervals/slice, 10–16 heartbeats brath hold time for 2 slices). In the case of RT, a user controllable parameter, minimum time (τ) to drive SSFP signal to MSS, was kept fixed at the default value of 450 ms for all the subjects. For the purpose of evaluating the impact of signal evolution during approach to SS on image quality, additional single-shot cardiac-triggered cine bSSFP (Fig. 1) images without preparation for SS were also acquired in 10 volunteers in a mid-ventricular slice with other imaging parameters identical to those used for BH and RT sequences.
Image analysis
All data were anonymized, randomized, and transferred to a commercial post processing workstation (ExtendedWorkSpace, Philips Healthcare).
Quantitative assessment of signal variation
For 23 healthy subjects, regions of interest (ROI) were drawn in the liver, myocardium, and blood pool at mid-ventricular level and were propagated across all cardiac phases in all cine bSSFP sequences. A histogram-based analysis was used to define two quantitative metrics designed to capture the extent of signal intensity (SI) variation across the cardiac cycle, as well as the percent duration of the cardiac cycle (PCC) during which tissue magnetization was close to the theoretically predicted MSS. The signal intensity of ROIs was normalized to the minimum SI across the cardiac cycle. To construct a histogram, normalized signal intensity values were plotted against the PCC spent at that normalized intensity level. Cumulative density function of normalized signal intensity of myocardial ROI over the cardiac cycle was computed for all healthy subjects. The ratio of the normalized signal intensity values at 95 and 5% of cumulative density function was computed as a metric to characterize the extent of signal intensity variation during the cardiac cycle. A normalized signal intensity of 1.25, which corresponded to MSS based on the simulations, was used to compute the PCC spent in SS. The blood-myocardial contrast (BMC) was normalized to myocardial signal intensity. Box plots were used to evaluate the signal intenisty variation (SIV), the PCC spent in SS by the liver and myocardium, and BMC; in these plots, non-overlapping notches indicate that the medians of the two groups differed at the 5% significance level. In addition, paired two-sided Student’s t-statistics for SIV, PCC, and BMC were used to compare among cardiac triggered, BH, and RT sequences.
Quantitative assessment of ventricular measurements
Two independent CMR readers, each with at least a decade of experience in clinical CMR, independently drew the ventricular contours necessary to compute LV and RV volumetric indices (end-diastolic volume, end-systolic volume, and EF) and LV mass. Descriptive statistics of each of these parameters for 60 patients using BH and RT are reported as mean ± standard deviation in absolute values and percentages. Bland-Altman analysis [29] and paired two-sided Student’s t-statistics were used to compare each of these parameters computed using the BH sequence with those computed using the RT sequence in 60 patients. Inter-observer variability for each of these parameters was assessed using Bland-Altman analysis in both BH and RT acquisitions in 60 patients.
Image quality assessment
For 60 patients, an independent expert with 12 years of experience in CMR who was blinded to the study design reviewed the stack of short-axis images. The image quality between the two techniques was compared by using the clinical scores, which were based on three parameters: BMC, endocardial-edge definition, and motion artifact. Each parameter was graded on a scale from 1 (non-diagnostic) to 5 (excellent) on review of all the images in the short-axis stack. In cases where all but one or two of the slices were of high quality, the clinical score was lowered for the entire stack. Details of the scoring criteria are provided in Table 1. For each of the three scoring criteria considered in the study, the percentage of clinical subjects that received a range of clinical scores was plotted as a bar graph. In addition, a similar bar graph was constructed with the combined score, computed as the equal weight average of the three scores, to underscore the overall performance of the technique on the basis of all three criteria. One-sided Wilcoxon signed rank tests were performed on clinical scores assigned to the BH and RT techniques.
Results
Numerical simulation
Numerical simulations showed that the signal intensity for the myocardium could be 6 to 8 times that of the MSS after the first few RF excitations, and as many as 140 continuous RF excitations (with a flip angle of 65o at a repetition time of 3 ms) were required to reach myocardial MSS. After the rate of change of myocardial magnetization fell below 0.5%, the magnetization corresponded to 125% of the MSS predicted by the analytical expression for the steady state signal (Fig. 2a). For healthy subject 3, liver and myocardial signal intensity curves with cardiac triggering (no dummy excitations) confirmed the requirement of continuous RF excitation for 350 ms and 450 ms (at TR interval of 3 ms), respectively, for the liver and myocardium (Fig. 2b) to attain steady signal intensity levels. Liver and myocardial signal intensity curves obtained with the RT sequence (dummy excitations for a duration of 450 ms) matched closely with those obtained using the BH sequence (dummy excitations for a duration of one R-R interval), confirming the attainment of MSS after 140 continuous RF pulses, as indicated by the numerical simulation (Fig. 2). Representative clinical images from healthy subject 3 are shown for all three techniques in Fig. 3.
Quantitative assessment of signal variation
Using histogram analysis, we showed the mean normalized signal intensity among all 23 healthy subjects for liver and myocardial ROIs for the BH and RT sequences, and among 10 healthy subjects for cardiac triggered sequences (Fig. 4a-b). When compared to the histogram mode of normalized signal intensity of liver tissue, the histogram mode of normalized signal intensity of the myocardium occupied a much smaller fraction of the cardiac cycle for all three techniques. The cumulative density function of myocardial normalized signal intensity indicated that, on average, the BH and RT techniques spent more than 95% of the cardiac cycle below a normalized signal intensity of 2, whereas for cardiac triggering, that value was 4 (Fig. 4d). Notably, the quantitative metrics of SIV, BMC, and PCC spent in SS for the cardiac triggered technique were significantly different from those for the BH and RT techniques (p < 0.001) (Fig. 5). In contrast, the difference in SIV and PCC spent in SS by myocardium was not statistically significant between the BH and RT techniques (Fig. 5a, b). Similarly, BMC was also not statistically different between the BH and RT techniques (Fig. 5c). The differences in SIV and the PCC between the liver and myocardium were statistically significant (p < 0.001) for both the BH and RT techniques because of through-plane motion of the myocardium (Fig. 5a, b).
Quantitative assessment of ventricular measurements
The RT sequence ran successfully in all 60 patients. The mean heart rate was 74 bpm (range 43–116), and the mean respiratory rate was 18 bpm (range 12–20). Total image acquisition time for RT was significantly longer than that for conventional BH (7.9 ± 3.4 min vs. 5.1 ± 2.6 min) (p < 0.001). In two cases, RT scan time was three times longer than that of the BH sequence because of specific breathing patterns. Table 2 provides descriptive statistics of the LV and RV volumetric indices and LV mass computed using BH and RT sequences in the 60 patients. Figure 6 depicts Bland-Altman plots comparing each of these parameters between BH and RT in 60 patients. Inter-observer variability for each of these parameters in both BH and RT acquisitions in 60 patients is reported in Table 3. For all the parameters the bias and limits of agreement between BH and RT sequences are in good agreement with inter-observer variability in both BH and RT sequences.
Image quality assessment
The combined clinical score was excellent (87%) to good (13%) for BH and excellent (38%) to good (45%) to moderate (17%) for RT (Fig. 6 H). In 17% of cases, the scores were equal, whereas the difference in combined clinical score was less than 0.5 in 48% and less than 1 in 82% of cases. In only one patient, edge definition and motion artifact were scored as poor for the RT sequence. The difference between clinical scores for the BH and RT sequences was statistically significant (p < 0.0001). Figure 7 shows representative BH and RT images from patients with a good to excellent clinical score for BH.
Discussion
The results from the two-center clinical study show that RT-bSSFP yielded diagnostic image quality cine images in all 60 patients encompassing a wide range of heart rates (43–116 bpm), respiratory rates (12–20 bpm), and ventricular indices (LV end-diastolic volume 84–397 mL, LV end-systolic volume 35–325 mL, LVEF 18–71%) with spatial, temporal, and contrast resolutions that were comparable to BH-bSSFP. To the best of our knowledge, this is the first study that has performed clinical validation of the free breathing cine bSSFP imaging in a large cohort of adult patients. The bias and limits of agreement between RT and BH sequences for LV and RV volumetric indices and LV mass were comparable to inter-observer variability in both RT and BH sequences. An interesting observation was that while the LV volumes were slightly overestimated (mean bias of 2–3 mL) with BH sequence compared to RT sequence, RV volumes were slightly underestimated (mean bias of 2–3 mL). Although such small differences in ventricular volumes may not be clinically significant, it may be worth investigating if the intra-thoracic pressure differences between respiratory suspended state and free breathing could be a contributing factor.
There are several key findings from the study that are worth noting. Firstly, conventional respiratory triggering methods for circumventing the breath-holding in bSSFP imaging that do not account for signal intensity modulation during the approach to SS can introduce significant artifacts as demonstrated in this study. Interestingly, even in BH bSSFP acquisitions, through-plane motion of the base toward the apex of the ventricle during systole introduces quantifiable temporal signal heterogeneity in the myocardium. Compared to other previously published respiratory triggered bSSFP cine CMR implementations [21, 23], RT sequence implementation [20, 22] mandates attainment of Mss prior to cardiac gated data acquisition. Metrics to quantify the temporal stability of the signal demonstrated that the RT sequence attained SS in synchrony with the physiologic cycle that is comparable to BH bSSFP. Importantly, the proposed acquisition technique and image quality evaluation metric have the potential to be used more broadly in other clinical applications. Also, it is worth noting that RT-bSSFP cine sequence is inherently compatible with prospective arrhythmia rejection and retrospective cardiac gating that are typically used with BH techniques, allowing the acquisition of a complete cardiac cycle. Furthermore, the cardiorespiratory synchronization approach can be seamlessly combined with other techniques such as non-cartesian sampling, sparse sampling, and self-navigation.
Secondly, bSSFP cine acquisition is feasible with both respiratory triggering using bellows and respiratory triggering with slice tracking using diaphragmatic RF navigators [20]. One advantage of using diaphragmatic RF navigators is the ability to track slice location in real time from the navigator information. However, the relatively long duration (at least one RR interval) between the slice position measurement from RF navigators and the completion of cine data acquisition may diminish this potential benefit. Furthermore, the quality of RF navigators may be compromised in patients with significant iron overload [23]. In contrast to RF navigators, respiratory bellows do not provide a direct measure of the lung-diaphragmatic interface. However, respiratory bellows provide the ability to monitor respiratory motion continuously and independent of CMR data acquisition without incurring additional RF dose.
Thirdly, respiratory triggered cine bSSFP sequences have several strengths. In contrast to continuous multi-NSA sequences which are often used to obtain cine bSSFP images in sedated patients or those with compromised respiratory function [22], the RF duty cycle in these sequences remains at about 50%, mitigating the specific energy deposition concerns and permits data acquisition over a longer scan duration. Freedom from the breath-hold constraint further allows one to attain bSSFP cine images with higher spatial resolution, temporal resolution, or coverage. Therefore, this technique has the potential to broaden the scope of clinical applications for cine bSSFP sequences, such as assessing diastolic function by obtaining a high temporal resolution dataset.
Another advantage of the RT bSSFP cine sequence is that it would reduce the potential for errors in estimation of ventricular chamber volumes due to inconsistent breath-holds. Although CMR is considered the gold standard for the estimation of ventricular chamber volumes, inconsistencies in the level of the diaphragm across breath-holds for each slice can result in under and/or over sampling of the myocardium in the direction of the ventricular long axis. This too may lead to variability in calculated volumes and EF, especially sampling-dependent inclusion or exclusion of the basal slice in end-systole [30]. This problem would be significantly diminished with free-breathing acquisitions [19].
Although we made no attempt to include only patients with compromised respiratory function, our clinical results suggest that the reduction in image quality of RT bSSFP was only small compared to the image quality attainable with the BH bSSFP technique. Notably, in our previous study, the image quality obtained with the RT-SSFP technique was better than that obtained with the multiple signal averaging that is routinely used in sedated pediatric patients [22]. In this two-center study in adults, our findings with respect to LV mass and ventricular volumetric results obtained with the RT and BH sequences are in line with the inter-observer variability.
The current implementation of RT-SSFP cine sequence has two key limitations. First, image quality of RT-bSSFP, while good, was inferior to that of the BH-bSSFP sequence. In 10/60 subjects, RT-SSFP image quality scores were moderate (still diagnostic) and we speculate that this reduction in image quality may be partly because of inconsistent respiratory patterns, such as when a patient starts inspiration before data acquisition from the expiratory trigger has completed, or respiratory drift over the course of an examination. The extent of the image quality degradation is dependent on the location of the acquisition segment with respect to center of the k-space. The image quality of the RT sequence could be improved by imposing an additional prospective constraint for rejecting data that were acquired during such respiratory inconsistencies at the cost of increasing scan time. A second limitation of the RT-bSSFP sequence is that it takes on an average 55% more time than the conventional cine BH-bSSFP sequence. For a cardiac cine bSSFP technique that requires n shots for a complete data acquisition, the RT-bSSFP sequence takes the time required for n respiratory cycles, whereas the BH-bSSFP sequence takes the time required for n + 1 heartbeats (1 extra beat to drive to MSS). This is because the RT sequence, in its current implementation, acquires data for a single cardiac cycle after each respiratory trigger. Often, the expiratory phase of the respiratory cycle is longer than one cardiac cycle and may permit data acquisition for several cardiac cycles, which would reduce the acquisition time of the RT sequence. Although this was not considered in this study, data acquisition over consecutive cardiac cycles during expiration with prospective constraint for rejecting the data are an important area of future investigation [31] as respiratory bellows permits continuous and simultaneous tracking of the respiratory motion.
Conclusion
In conclusion, we have shown that the RT sequence is a robust, free-breathing alternative to BH-bSSFP for evaluating ventricular function in patients with impaired breath-holding capacity and/or arrhythmia. The RT sequence yields cine images with contrast and spatio-temporal resolutions that are identical to those attained by using the BH sequence, albeit with a modest time penalty and a small reduction in image quality. We believe that the elimination of the BH constraint on a cine bSSFP sequence has the potential to improve the evaluation of ventricular function in subjects with compromised respiratory capacity with small reduction in image quality. Moreover, this would allow one to acquire high temporal resolution cine images necessary for the assessment of transient phenomena, such as isovolumic relaxation time, as well as the ability to acquire multi-phase, high-resolution anatomic images, such as coronary MR angiography.
Abbreviations
- M:
-
magnetization
- SS:
-
Steady state
- BA:
-
Bland-Altman analysis
- BH:
-
Breath-hold
- BMC:
-
Blood-myocardial contrast
- bSSFP:
-
Balanced steady-state free precession
- CMR:
-
Cardiovascular magnetic resonance
- EDV:
-
End-diastolic volume
- EF:
-
Ejection fraction
- ESV:
-
End-systolic volume
- LV:
-
Left ventricle/left ventricular
- NI:
-
Normalized signal intensity
- PCC:
-
Percent duration of the cardiac cycle
- RF:
-
Radio frequency
- RG:
-
Respiratory gated
- ROI:
-
Region of interest
- RT:
-
Respiratory triggered
- RV:
-
Right ventricle/right ventricular
- SENSE:
-
Snesitivity encoding
- SI:
-
Signal intensity
- SIV:
-
Signal intensity variation
- SNR:
-
Signal-to-noise ratio
- ss:
-
Steady state
- TE:
-
Echo time
- TR:
-
Time
- VCG:
-
Vector electrocardiogram
References
Dodge HT, Baxley WA. Left ventricular volume and mass and their significance in heart disease. Am J Cardiol. 1969;23:528–37.
Levy D, Garrison RJ, Savage DD, Kannel WB, Castelli WP. Prognostic implications of Echocardiographically determined left ventricular mass in the Framingham heart study. N Engl J Med. 1990;322:1561–6.
Anavekar N, Skali H, McMurray J, Swedberg K, Yusus S, Granger C, et al. Influence of ejection fraction on cardiovascular outcomes in a broad Spectrum of heart failure patients. Circulation. 2005;112:3738–44.
Taniguchi K, Nakano S, Hirose H, Matsuda H, Shirakura R, Sakai K, et al. Preoperative left ventricular function: minimal requirement for successful late results of valve replacement for aortic regurgitation. J Am Coll Cardiol. 1987;10:510–8.
White HD, Norris RM, Brown MA, Brandt PW, Whitlock RM, Wild CJ. Left ventricular end-systolic volume as the major determinant of survival after recovery from myocardial infarction. Circulation. 1987;76:44–51.
Bellenger NG, Davies LC, Francis JM, Coats AJ, Pennell DJ. Reduction in sample size for studies of remodeling in heart failure by the use of cardiovascular magnetic resonance. J Cardiovasc Magn Reson Off J Soc Cardiovasc Magn Reson. 2000;2:271–8.
Bottini PB, Carr AA, Prisant LM, Flickinger FW, Allison JD, Gottdiener JS. Magnetic resonance imaging compared to echocardiography to assess left ventricular mass in the hypertensive patient. Am J Hypertens. 1995;8:221–8.
Grothues F, Smith GC, Moon JCC, Bellenger NG, Collins P, Klein HU, et al. Comparison of interstudy reproducibility of cardiovascular magnetic resonance with two-dimensional echocardiography in normal subjects and in patients with heart failure or left ventricular hypertrophy. Am J Cardiol. 2002;90:29–34.
Pennell DJ, Sechtem UP, Higgins CB, Manning WJ, Pohost GM, Rademakers FE, et al. Clinical indications for cardiovascular magnetic resonance (CMR): consensus panel report. Eur Heart J. 2004;25:1940–65.
Finn JP, Nael K, Deshpande V, Ratib O, Laub G. Cardiac MR imaging: state of the technology. Radiology. 2006;241:338–54.
Scheffler K, Lehnhardt S. Principles and applications of balanced SSFP techniques. Eur Radiol. 2003;13:2409–18.
Srinivasan S, Ennis DB. Optimal flip angle for high contrast balanced SSFP cardiac cine imaging. Magn Reson Med. 2015;73:1095–103.
Scheffler K. On the transient phase of balanced SSFP sequences. Magn Reson Med. 2003;49:781–3.
Hargreaves BA, Vasanawala SS, Pauly JM, Nishimura DG. Characterization and reduction of the transient response in steady-state MR imaging. Magn Reson Med. 2001;46:149–58.
Markl M, Alley MT, Elkins CJ, Pelc NJ. Flow effects in balanced steady state free precession imaging. Magn Reson Med. 2003;50:892–903.
Sudarski S, Henzler T, Haubenreisser H, Dösch C, Zenge MO, Schmidt M, et al. Free-breathing sparse sampling cine MR imaging with iterative reconstruction for the assessment of left ventricular function and mass at 3. 0 T Radiology. 2016;282:74–83.
Vincenti G, Monney P, Chaptinel J, Rutz T, Coppo S, Zenge MO, et al. Compressed sensing single-breath-hold CMR for fast quantification of LV function, volumes, and mass. JACC Cardiovasc Imaging. 2014;7:882–92.
Zhou Z, Han F, Rapacchi S, Nguyen K-L, Brunengraber DZ, G-HJ K, et al. Accelerated ferumoxytol-enhanced 4D multiphase, steady-state imaging with contrast enhancement (MUSIC) cardiovascular MRI: validation in pediatric congenital heart disease. NMR Biomed. 2017;30
Peters DC, Nezafat R, Eggers H, Stehning C, Manning WJ. 2D free-breathing dual navigator-gated cardiac function validated against the 2D breath-hold acquisition. J Magn Reson Imaging. 2008;28:773–7.
Pednekar A, Krishnamurthy R, Arena C, Muthupillai R, Cheong B, Dees D. Respiratory triggered retrospectively cardiac gated cine steady-state free precession (SSFP) imaging. In: 20th proc. Intl. Soc. mag. Reson. Med. Melbourne; 2012. p. 3938.
Henningsson M, Chan R, Goddu B, Goepfert LA, Razavi R, Botnar RM, et al. Contrast-enhanced specific absorption rate-efficient 3D cardiac cine with respiratory-triggered radiofrequency gating. J Magn Reson Imaging. 2013;37:986–92.
Krishnamurthy R, Pednekar A, Atweh LA, Vogelius E, Chu ZD, Zhang W, et al. Clinical validation of free breathing respiratory triggered retrospectively cardiac gated cine balanced steady-state free precession cardiovascular magnetic resonance in sedated children. J Cardiovasc Magn Reson. 2015;17 https://doi.org/10.1186/s12968-014-0101-1.
Moghari MH, Komarlu R, Annese D, Geva T, Powell AJ. Free-breathing steady-state free precession cine cardiac magnetic resonance with respiratory navigator gating. Magn Reson Med. 2015;73:1555–61.
Han F, Zhou Z, Han E, Gao Y, Nguyen K-L, Finn JP, et al. Self-gated 4D multiphase, steady-state imaging with contrast enhancement (MUSIC) using rotating cartesian K-space (ROCK): validation in children with congenital heart disease. Magn Reson Med. 2016;
Sayin O, Saybasili H, Zviman MM, Griswold M, Halperin H, Seiberlich N, et al. Real-time free-breathing cardiac imaging with self-calibrated through-time radial GRAPPA. Magn Reson Med. 2017;77:250–64.
Yerly J, Ginami G, Nordio G, Coristine AJ, Coppo S, Monney P, et al. Coronary endothelial function assessment using self-gated cardiac cine MRI and k-t sparse SENSE. Magn Reson Med. 2016;76:1443–54.
Feng L, Axel L, Chandarana H, Block KT, Sodickson DK, Otazo R. XD-GRASP: golden-angle radial MRI with reconstruction of extra motion-state dimensions using compressed sensing. Magn Reson Med. 2016;75:775–88.
Han F, Rapacchi S, Khan S, Ayad I, Salusky I, Gabriel S, et al. Four-dimensional, multiphase, steady-state imaging with contrast enhancement (MUSIC) in the heart: a feasibility study in children. Magn Reson Med. 2015;74:1042–9.
Bland JM, Altman DG. Statistical methods for assessing agreement between two methods of clinical measurement. Lancet. 1986;1:307–10.
Karamitsos TD, Hudsmith LE, Selvanayagam JB, Neubauer S, Francis JM. Operator induced variability in left ventricular measurements with cardiovascular magnetic resonance is improved after training. J Cardiovasc Magn Reson. 2007;9:777–83.
Pednekar A. Clinical Validation of Free Breathing CArdioREspiratory Synchronized (CARESynch) Balanced Steady-State Free Precession (bSSFP) Cine Imaging. In: 20th Annual SCMR Scientific Sessions Abstract Supplement. Washington, DC; 2017.
Acknowledgements
We gratefully acknowledge Janie Swaab, RT, Sharon Berry, RT, Kathy Kohut, RT, Debra Dees, RN, and Melissa Andrews, RN for their help in data collection.
Availability of data and materials
The datasets generated and analyzed during this study are available from the corresponding author on reasonable request.
Author information
Authors and Affiliations
Contributions
AP: performed numerical simulation, performed CMR pulse sequence design and implementation, performed statistical analysis, drafted the manuscript. HW: participated in data collection, helped drafting the manuscript. SF: participated in data collection and data analysis. BC: participated in data collection, data analysis, and drafting the manuscript. RM: participated in CMR pulse sequence design, designed the study, helped in statistical analysis and equal contribution in drafting the manuscript. All authors read and approved the final manuscript.
Corresponding author
Ethics declarations
Ethics approval and consent to participate
This study was approved by the Institutional Ethics Committee of Catholic Health Initiative’s Institute for Research and Innovation under evolution CMR techniques protocol and complied with the Health Insurance Portability and Accountability Act of 1996. All subjects gave written informed consent before being enrolled in the study.
Competing interests
AP: employee of Philips Healthcare until January 2017.
HW: employee of Philips Healthcare.
Authors employed by the CMR scanner manufacturer were not part of the clinical patient data acquisition and image quality assessment.
Publisher’s Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Rights and permissions
Open Access This article is distributed under the terms of the Creative Commons Attribution 4.0 International License (http://creativecommons.org/licenses/by/4.0/), which permits unrestricted use, distribution, and reproduction in any medium, provided you give appropriate credit to the original author(s) and the source, provide a link to the Creative Commons license, and indicate if changes were made. The Creative Commons Public Domain Dedication waiver (http://creativecommons.org/publicdomain/zero/1.0/) applies to the data made available in this article, unless otherwise stated.
About this article
Cite this article
Pednekar, A.S., Wang, H., Flamm, S. et al. Two-center clinical validation and quantitative assessment of respiratory triggered retrospectively cardiac gated balanced-SSFP cine cardiovascular magnetic resonance imaging in adults. J Cardiovasc Magn Reson 20, 44 (2018). https://doi.org/10.1186/s12968-018-0467-6
Received:
Accepted:
Published:
DOI: https://doi.org/10.1186/s12968-018-0467-6